Cargando…

In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days

This in vitro study evaluates the performance of a disposable vibrating-mesh nebulizer when used for 28 days. A lung model was used to simulate the breathing pattern of an adult with chronic obstructive pulmonary disease. The vibrating-mesh nebulizer was used for three treatments/day over 28 days wi...

Descripción completa

Detalles Bibliográficos
Autores principales: Lin, Hui-Ling, Chen, Chi-Shuo, Fink, James B., Lee, Guo-Hao, Huang, Chun-Wei, Chen, Jui-Chi, Chiang, Zi Yi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7602390/
https://www.ncbi.nlm.nih.gov/pubmed/33076232
http://dx.doi.org/10.3390/pharmaceutics12100971
_version_ 1783603667054624768
author Lin, Hui-Ling
Chen, Chi-Shuo
Fink, James B.
Lee, Guo-Hao
Huang, Chun-Wei
Chen, Jui-Chi
Chiang, Zi Yi
author_facet Lin, Hui-Ling
Chen, Chi-Shuo
Fink, James B.
Lee, Guo-Hao
Huang, Chun-Wei
Chen, Jui-Chi
Chiang, Zi Yi
author_sort Lin, Hui-Ling
collection PubMed
description This in vitro study evaluates the performance of a disposable vibrating-mesh nebulizer when used for 28 days. A lung model was used to simulate the breathing pattern of an adult with chronic obstructive pulmonary disease. The vibrating-mesh nebulizer was used for three treatments/day over 28 days without cleaning after each test. Results showed that the inhaled drug dose was similar during four weeks of use (p = 0.157), with 16.73 ± 4.46% at baseline and 15.29 ± 2.45%, 16.21 ± 2.21%, 17.56 ± 1.98%, and 17.13 ± 1.81%, after the first, second, third, and fourth weeks, respectively. The particle size distribution, residual drug volume, and nebulization time remained similar across four weeks of use (p = 0.110, p = 0.763, and p = 0.573, respectively). Mesh was inspected using optical microscopy and showed that approximately 50% of mesh pores were obscured after 84 runs, and light penetration through the aperture plate was significantly reduced after the 21st use (p < 0.001) with no correlation to nebulizer performance. We conclude that the vibrating-mesh nebulizer delivered doses of salbutamol solution effectively over four weeks without cleaning after each use even though the patency and clarity of the aperture plate were reduced by the first week of use.
format Online
Article
Text
id pubmed-7602390
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-76023902020-11-01 In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days Lin, Hui-Ling Chen, Chi-Shuo Fink, James B. Lee, Guo-Hao Huang, Chun-Wei Chen, Jui-Chi Chiang, Zi Yi Pharmaceutics Article This in vitro study evaluates the performance of a disposable vibrating-mesh nebulizer when used for 28 days. A lung model was used to simulate the breathing pattern of an adult with chronic obstructive pulmonary disease. The vibrating-mesh nebulizer was used for three treatments/day over 28 days without cleaning after each test. Results showed that the inhaled drug dose was similar during four weeks of use (p = 0.157), with 16.73 ± 4.46% at baseline and 15.29 ± 2.45%, 16.21 ± 2.21%, 17.56 ± 1.98%, and 17.13 ± 1.81%, after the first, second, third, and fourth weeks, respectively. The particle size distribution, residual drug volume, and nebulization time remained similar across four weeks of use (p = 0.110, p = 0.763, and p = 0.573, respectively). Mesh was inspected using optical microscopy and showed that approximately 50% of mesh pores were obscured after 84 runs, and light penetration through the aperture plate was significantly reduced after the 21st use (p < 0.001) with no correlation to nebulizer performance. We conclude that the vibrating-mesh nebulizer delivered doses of salbutamol solution effectively over four weeks without cleaning after each use even though the patency and clarity of the aperture plate were reduced by the first week of use. MDPI 2020-10-15 /pmc/articles/PMC7602390/ /pubmed/33076232 http://dx.doi.org/10.3390/pharmaceutics12100971 Text en © 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Lin, Hui-Ling
Chen, Chi-Shuo
Fink, James B.
Lee, Guo-Hao
Huang, Chun-Wei
Chen, Jui-Chi
Chiang, Zi Yi
In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days
title In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days
title_full In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days
title_fullStr In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days
title_full_unstemmed In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days
title_short In Vitro Evaluation of a Vibrating-Mesh Nebulizer Repeatedly Use over 28 Days
title_sort in vitro evaluation of a vibrating-mesh nebulizer repeatedly use over 28 days
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7602390/
https://www.ncbi.nlm.nih.gov/pubmed/33076232
http://dx.doi.org/10.3390/pharmaceutics12100971
work_keys_str_mv AT linhuiling invitroevaluationofavibratingmeshnebulizerrepeatedlyuseover28days
AT chenchishuo invitroevaluationofavibratingmeshnebulizerrepeatedlyuseover28days
AT finkjamesb invitroevaluationofavibratingmeshnebulizerrepeatedlyuseover28days
AT leeguohao invitroevaluationofavibratingmeshnebulizerrepeatedlyuseover28days
AT huangchunwei invitroevaluationofavibratingmeshnebulizerrepeatedlyuseover28days
AT chenjuichi invitroevaluationofavibratingmeshnebulizerrepeatedlyuseover28days
AT chiangziyi invitroevaluationofavibratingmeshnebulizerrepeatedlyuseover28days