Improving Effectiveness of Phototherapy in an Academic Center: A Quality Improvement Project
Neonatal hyperbilirubinemia is a common cause of delayed discharge and readmissions in our institution. As previously published, the irradiance our phototherapy (PT) units provided was below the irradiance recommended by the AAP for intensive phototherapy (>30 µW/cm(2)/nm). We measured irradiance...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7605035/ https://www.ncbi.nlm.nih.gov/pubmed/33195747 http://dx.doi.org/10.1177/2333794X20969275 |
Sumario: | Neonatal hyperbilirubinemia is a common cause of delayed discharge and readmissions in our institution. As previously published, the irradiance our phototherapy (PT) units provided was below the irradiance recommended by the AAP for intensive phototherapy (>30 µW/cm(2)/nm). We measured irradiance delivered by our PT units (Drager 4000) using a standardized footprint grid. By varying number of blue and white fluorescent PT lights, height of PT unit above the neonate and type of bed used (open bassinet versus isolette), we determined the optimal PT arrangement needed to deliver intensive PT (30 µW/cm(2)/nm). We then developed a standardized, multidisciplinary protocol specifying light arrangement and distance required needed to achieve the desired irradiance level. We were able to show improved irradiance following above changes. Onsite measurement of irradiance provided by local phototherapy units and development of a multidisciplinary, standardized protocol are necessary to assure delivery of recommended levels PT for neonates with hyperbilirubinemia. |
---|