Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging

BACKGROUND: Small-animal single-photon emission computed tomography (SPECT) systems with multi-pinhole collimation and large stationary detectors have advantages compared to systems with moving small detectors. These systems benefit from less labour-intensive maintenance and quality control as fewer...

Descripción completa

Detalles Bibliográficos
Autores principales: Hoffmann, Jan V., Janssen, Jan P., Kanno, Takayuki, Shibutani, Takayuki, Onoguchi, Masahisa, Lapa, Constantin, Grunz, Jan-Peter, Buck, Andreas K., Higuchi, Takahiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7606439/
https://www.ncbi.nlm.nih.gov/pubmed/33140263
http://dx.doi.org/10.1186/s40658-020-00335-6
_version_ 1783604482058223616
author Hoffmann, Jan V.
Janssen, Jan P.
Kanno, Takayuki
Shibutani, Takayuki
Onoguchi, Masahisa
Lapa, Constantin
Grunz, Jan-Peter
Buck, Andreas K.
Higuchi, Takahiro
author_facet Hoffmann, Jan V.
Janssen, Jan P.
Kanno, Takayuki
Shibutani, Takayuki
Onoguchi, Masahisa
Lapa, Constantin
Grunz, Jan-Peter
Buck, Andreas K.
Higuchi, Takahiro
author_sort Hoffmann, Jan V.
collection PubMed
description BACKGROUND: Small-animal single-photon emission computed tomography (SPECT) systems with multi-pinhole collimation and large stationary detectors have advantages compared to systems with moving small detectors. These systems benefit from less labour-intensive maintenance and quality control as fewer prone parts are moving, higher accuracy for focused scans and maintaining high resolution with increased sensitivity due to focused pinholes on the field of view. This study aims to investigate the performance of a novel ultra-high-resolution scanner with two-detector configuration (U-SPECT5-E) and to compare its image quality to a conventional micro-SPECT system with three stationary detectors (U-SPECT(+)). METHODS: The new U-SPECT5-E with two stationary detectors was used for acquiring data with (99m)Tc-filled point source, hot-rod and uniformity phantoms to analyse sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR). Three dedicated multi-pinhole mouse collimators with 75 pinholes each and 0.25-, 0.60- and 1.00-mm pinholes for extra ultra-high resolution (XUHR-M), general-purpose (GP-M) and ultra-high sensitivity (UHS-M) imaging were examined. For CNR analysis, four different activity ranges representing low- and high-count settings were investigated for all three collimators. The experiments for the performance assessment were repeated with the same GP-M collimator in the three-detector U-SPECT(+) for comparison. RESULTS: Peak sensitivity was 237 cps/MBq (XUHR-M), 847 cps/MBq (GP-M), 2054 cps/MBq (UHS-M) for U-SPECT5-E and 1710 cps/MBq (GP-M) for U-SPECT(+). In the visually analysed sections of the reconstructed mini Derenzo phantoms, rods as small as 0.35 mm (XUHR-M), 0.50 mm (GP-M) for the two-detector as well as the three-detector SPECT and 0.75 mm (UHS-M) were resolved. Uniformity for maximum resolution recorded 40.7% (XUHR-M), 29.1% (GP-M, U-SPECT5-E), 16.3% (GP-M, U-SPECT(+)) and 23.0% (UHS-M), respectively. UHS-M reached highest CNR values for low-count images; for rods smaller than 0.45 mm, acceptable CNR was only achieved by XUHR-M. GP-M was superior for imaging rods sized from 0.60 to 1.50 mm for intermediate activity concentrations. U-SPECT5-E and U-SPECT(+) both provided comparable CNR. CONCLUSIONS: While uniformity and sensitivity are negatively affected by the absence of a third detector, the investigated U-SPECT5-E system with two stationary detectors delivers excellent spatial resolution and CNR comparable to the performance of an established three-detector-setup.
format Online
Article
Text
id pubmed-7606439
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-76064392020-11-04 Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging Hoffmann, Jan V. Janssen, Jan P. Kanno, Takayuki Shibutani, Takayuki Onoguchi, Masahisa Lapa, Constantin Grunz, Jan-Peter Buck, Andreas K. Higuchi, Takahiro EJNMMI Phys Original Research BACKGROUND: Small-animal single-photon emission computed tomography (SPECT) systems with multi-pinhole collimation and large stationary detectors have advantages compared to systems with moving small detectors. These systems benefit from less labour-intensive maintenance and quality control as fewer prone parts are moving, higher accuracy for focused scans and maintaining high resolution with increased sensitivity due to focused pinholes on the field of view. This study aims to investigate the performance of a novel ultra-high-resolution scanner with two-detector configuration (U-SPECT5-E) and to compare its image quality to a conventional micro-SPECT system with three stationary detectors (U-SPECT(+)). METHODS: The new U-SPECT5-E with two stationary detectors was used for acquiring data with (99m)Tc-filled point source, hot-rod and uniformity phantoms to analyse sensitivity, spatial resolution, uniformity and contrast-to-noise ratio (CNR). Three dedicated multi-pinhole mouse collimators with 75 pinholes each and 0.25-, 0.60- and 1.00-mm pinholes for extra ultra-high resolution (XUHR-M), general-purpose (GP-M) and ultra-high sensitivity (UHS-M) imaging were examined. For CNR analysis, four different activity ranges representing low- and high-count settings were investigated for all three collimators. The experiments for the performance assessment were repeated with the same GP-M collimator in the three-detector U-SPECT(+) for comparison. RESULTS: Peak sensitivity was 237 cps/MBq (XUHR-M), 847 cps/MBq (GP-M), 2054 cps/MBq (UHS-M) for U-SPECT5-E and 1710 cps/MBq (GP-M) for U-SPECT(+). In the visually analysed sections of the reconstructed mini Derenzo phantoms, rods as small as 0.35 mm (XUHR-M), 0.50 mm (GP-M) for the two-detector as well as the three-detector SPECT and 0.75 mm (UHS-M) were resolved. Uniformity for maximum resolution recorded 40.7% (XUHR-M), 29.1% (GP-M, U-SPECT5-E), 16.3% (GP-M, U-SPECT(+)) and 23.0% (UHS-M), respectively. UHS-M reached highest CNR values for low-count images; for rods smaller than 0.45 mm, acceptable CNR was only achieved by XUHR-M. GP-M was superior for imaging rods sized from 0.60 to 1.50 mm for intermediate activity concentrations. U-SPECT5-E and U-SPECT(+) both provided comparable CNR. CONCLUSIONS: While uniformity and sensitivity are negatively affected by the absence of a third detector, the investigated U-SPECT5-E system with two stationary detectors delivers excellent spatial resolution and CNR comparable to the performance of an established three-detector-setup. Springer International Publishing 2020-11-02 /pmc/articles/PMC7606439/ /pubmed/33140263 http://dx.doi.org/10.1186/s40658-020-00335-6 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Original Research
Hoffmann, Jan V.
Janssen, Jan P.
Kanno, Takayuki
Shibutani, Takayuki
Onoguchi, Masahisa
Lapa, Constantin
Grunz, Jan-Peter
Buck, Andreas K.
Higuchi, Takahiro
Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging
title Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging
title_full Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging
title_fullStr Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging
title_full_unstemmed Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging
title_short Performance evaluation of fifth-generation ultra-high-resolution SPECT system with two stationary detectors and multi-pinhole imaging
title_sort performance evaluation of fifth-generation ultra-high-resolution spect system with two stationary detectors and multi-pinhole imaging
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7606439/
https://www.ncbi.nlm.nih.gov/pubmed/33140263
http://dx.doi.org/10.1186/s40658-020-00335-6
work_keys_str_mv AT hoffmannjanv performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging
AT janssenjanp performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging
AT kannotakayuki performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging
AT shibutanitakayuki performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging
AT onoguchimasahisa performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging
AT lapaconstantin performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging
AT grunzjanpeter performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging
AT buckandreask performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging
AT higuchitakahiro performanceevaluationoffifthgenerationultrahighresolutionspectsystemwithtwostationarydetectorsandmultipinholeimaging