Cargando…
A new proximal femoral nail antirotation design: Is it effective in preventing varus collapse and cut-out?
OBJECTIVES: This study aims to compare the mechanical features of the existing proximal femoral nail antirotation (PFNA) system and the new PFNA system that we designed using three-dimensional (3D) finite element analysis. MATERIALS AND METHODS: This experimental study was conducted between 2019 and...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Bayçınar Medical Publishing
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7607958/ https://www.ncbi.nlm.nih.gov/pubmed/32962571 http://dx.doi.org/10.5606/ehc.2020.75113 |
_version_ | 1783604745238216704 |
---|---|
author | Öner, Kerim Durusoy, Serhat Özer, Alaettin |
author_facet | Öner, Kerim Durusoy, Serhat Özer, Alaettin |
author_sort | Öner, Kerim |
collection | PubMed |
description | OBJECTIVES: This study aims to compare the mechanical features of the existing proximal femoral nail antirotation (PFNA) system and the new PFNA system that we designed using three-dimensional (3D) finite element analysis. MATERIALS AND METHODS: This experimental study was conducted between 2019 and 2020. We constructed two femur models with Arbeitsgemeinschaft für Osteosynthesefragen (AO) type A1 fractures using 3D computed tomography scans. The new and standard PFNA designs were inserted into the femur models and subsequently transferred to the program. We investigated the distribution of stress on the tip of the lag screw, the calcar region, lag screw-nail junction, and the additional screw inserted through the greater trochanter (only present in the new PFNA design) using 3D finite element analysis. RESULTS: When the von Mises stress distributions in our models were examined, the maximum stress at the lag screw-nail junction was 18 mpa in the new design PFNA, while it was 20 mpa in the classic PFNA model. The maximum stress at the junction of the additional screw that had greater trochanter inlet with the nail was found as 42.5 mpa. The maximum stress on the calcar region was found to be 10 mpa at the new design PFNA, while it was 13 mpa with 30% increase in the classic PFNA. The stress on the tip of the lag screw was found to be 49 mpa in the classic PFNA design, while in the new design PFNA it was found as 28 mpa with a decrease of more than 40%. CONCLUSION: As per our findings, the new PFNA design leads to reduced stress on the lag screw-nail junction, the calcar region, and the tip of the lag screw. |
format | Online Article Text |
id | pubmed-7607958 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Bayçınar Medical Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-76079582020-11-04 A new proximal femoral nail antirotation design: Is it effective in preventing varus collapse and cut-out? Öner, Kerim Durusoy, Serhat Özer, Alaettin Jt Dis Relat Surg Original Article OBJECTIVES: This study aims to compare the mechanical features of the existing proximal femoral nail antirotation (PFNA) system and the new PFNA system that we designed using three-dimensional (3D) finite element analysis. MATERIALS AND METHODS: This experimental study was conducted between 2019 and 2020. We constructed two femur models with Arbeitsgemeinschaft für Osteosynthesefragen (AO) type A1 fractures using 3D computed tomography scans. The new and standard PFNA designs were inserted into the femur models and subsequently transferred to the program. We investigated the distribution of stress on the tip of the lag screw, the calcar region, lag screw-nail junction, and the additional screw inserted through the greater trochanter (only present in the new PFNA design) using 3D finite element analysis. RESULTS: When the von Mises stress distributions in our models were examined, the maximum stress at the lag screw-nail junction was 18 mpa in the new design PFNA, while it was 20 mpa in the classic PFNA model. The maximum stress at the junction of the additional screw that had greater trochanter inlet with the nail was found as 42.5 mpa. The maximum stress on the calcar region was found to be 10 mpa at the new design PFNA, while it was 13 mpa with 30% increase in the classic PFNA. The stress on the tip of the lag screw was found to be 49 mpa in the classic PFNA design, while in the new design PFNA it was found as 28 mpa with a decrease of more than 40%. CONCLUSION: As per our findings, the new PFNA design leads to reduced stress on the lag screw-nail junction, the calcar region, and the tip of the lag screw. Bayçınar Medical Publishing 2020-06-24 /pmc/articles/PMC7607958/ /pubmed/32962571 http://dx.doi.org/10.5606/ehc.2020.75113 Text en Copyright © 2020, Turkish Joint Diseases Foundation http://creativecommons.org/licenses/by-nc/4.0/ This is an open access article under the terms of the Creative Commons Attribution-NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes. |
spellingShingle | Original Article Öner, Kerim Durusoy, Serhat Özer, Alaettin A new proximal femoral nail antirotation design: Is it effective in preventing varus collapse and cut-out? |
title | A new proximal femoral nail antirotation design: Is it effective in preventing varus collapse and cut-out? |
title_full | A new proximal femoral nail antirotation design: Is it effective in preventing varus collapse and cut-out? |
title_fullStr | A new proximal femoral nail antirotation design: Is it effective in preventing varus collapse and cut-out? |
title_full_unstemmed | A new proximal femoral nail antirotation design: Is it effective in preventing varus collapse and cut-out? |
title_short | A new proximal femoral nail antirotation design: Is it effective in preventing varus collapse and cut-out? |
title_sort | new proximal femoral nail antirotation design: is it effective in preventing varus collapse and cut-out? |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7607958/ https://www.ncbi.nlm.nih.gov/pubmed/32962571 http://dx.doi.org/10.5606/ehc.2020.75113 |
work_keys_str_mv | AT onerkerim anewproximalfemoralnailantirotationdesignisiteffectiveinpreventingvaruscollapseandcutout AT durusoyserhat anewproximalfemoralnailantirotationdesignisiteffectiveinpreventingvaruscollapseandcutout AT ozeralaettin anewproximalfemoralnailantirotationdesignisiteffectiveinpreventingvaruscollapseandcutout AT onerkerim newproximalfemoralnailantirotationdesignisiteffectiveinpreventingvaruscollapseandcutout AT durusoyserhat newproximalfemoralnailantirotationdesignisiteffectiveinpreventingvaruscollapseandcutout AT ozeralaettin newproximalfemoralnailantirotationdesignisiteffectiveinpreventingvaruscollapseandcutout |