Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device
Ambulatory blood pressure (BP) monitoring is recommended to improve the management of hypertension. Here, we investigated the accuracy of BP estimated using a wearable cuff-less device, InBodyWATCH, compared with BP measured using a manual sphygmomanometer. Thirty-five adults were enrolled (age 57.1...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7642418/ https://www.ncbi.nlm.nih.gov/pubmed/33149118 http://dx.doi.org/10.1038/s41598-020-75892-y |
_version_ | 1783606080326074368 |
---|---|
author | Moon, Joon Ho Kang, Myung-Kyun Choi, Chang-Eun Min, Jeonghee Lee, Hae-Young Lim, Soo |
author_facet | Moon, Joon Ho Kang, Myung-Kyun Choi, Chang-Eun Min, Jeonghee Lee, Hae-Young Lim, Soo |
author_sort | Moon, Joon Ho |
collection | PubMed |
description | Ambulatory blood pressure (BP) monitoring is recommended to improve the management of hypertension. Here, we investigated the accuracy of BP estimated using a wearable cuff-less device, InBodyWATCH, compared with BP measured using a manual sphygmomanometer. Thirty-five adults were enrolled (age 57.1 ± 17.9 years). The BP was estimated using InBodyWATCH with an individualized estimation based on a neural network model. Three paired sets of BPs from the two devices were compared using correlation analysis and Bland–Altman plots (n = 105 paired BP readings). The correlations for both systolic and diastolic BP (SBP and DBP) between the two devices were high (r = 0.964 and 0.939, both P < 0.001). The mean difference was 2.2 ± 6.1 mmHg for SBP and −0.2 ± 4.2 mmHg for DBP; these were not significant (P = 0.472 for SBP and P = 0.880 for DBP). The proportions of estimated SBP/DBP obtained from the InBodyWATCH within ± 5 mmHg of manual SBP/DBP were 71.4%/83.8%; within ± 10 mmHg they were 86.7%/98.1%; and within ± 15 mmHg they were 97.1%/99.0%. The estimated BP from this wearable cuff-less device correlated highly with the manual BP and showed good accuracy, suggesting its potential to be used in ambulatory BP monitoring. |
format | Online Article Text |
id | pubmed-7642418 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Nature Publishing Group UK |
record_format | MEDLINE/PubMed |
spelling | pubmed-76424182020-11-06 Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device Moon, Joon Ho Kang, Myung-Kyun Choi, Chang-Eun Min, Jeonghee Lee, Hae-Young Lim, Soo Sci Rep Article Ambulatory blood pressure (BP) monitoring is recommended to improve the management of hypertension. Here, we investigated the accuracy of BP estimated using a wearable cuff-less device, InBodyWATCH, compared with BP measured using a manual sphygmomanometer. Thirty-five adults were enrolled (age 57.1 ± 17.9 years). The BP was estimated using InBodyWATCH with an individualized estimation based on a neural network model. Three paired sets of BPs from the two devices were compared using correlation analysis and Bland–Altman plots (n = 105 paired BP readings). The correlations for both systolic and diastolic BP (SBP and DBP) between the two devices were high (r = 0.964 and 0.939, both P < 0.001). The mean difference was 2.2 ± 6.1 mmHg for SBP and −0.2 ± 4.2 mmHg for DBP; these were not significant (P = 0.472 for SBP and P = 0.880 for DBP). The proportions of estimated SBP/DBP obtained from the InBodyWATCH within ± 5 mmHg of manual SBP/DBP were 71.4%/83.8%; within ± 10 mmHg they were 86.7%/98.1%; and within ± 15 mmHg they were 97.1%/99.0%. The estimated BP from this wearable cuff-less device correlated highly with the manual BP and showed good accuracy, suggesting its potential to be used in ambulatory BP monitoring. Nature Publishing Group UK 2020-11-04 /pmc/articles/PMC7642418/ /pubmed/33149118 http://dx.doi.org/10.1038/s41598-020-75892-y Text en © The Author(s) 2020 Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. |
spellingShingle | Article Moon, Joon Ho Kang, Myung-Kyun Choi, Chang-Eun Min, Jeonghee Lee, Hae-Young Lim, Soo Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device |
title | Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device |
title_full | Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device |
title_fullStr | Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device |
title_full_unstemmed | Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device |
title_short | Validation of a wearable cuff-less wristwatch-type blood pressure monitoring device |
title_sort | validation of a wearable cuff-less wristwatch-type blood pressure monitoring device |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7642418/ https://www.ncbi.nlm.nih.gov/pubmed/33149118 http://dx.doi.org/10.1038/s41598-020-75892-y |
work_keys_str_mv | AT moonjoonho validationofawearablecufflesswristwatchtypebloodpressuremonitoringdevice AT kangmyungkyun validationofawearablecufflesswristwatchtypebloodpressuremonitoringdevice AT choichangeun validationofawearablecufflesswristwatchtypebloodpressuremonitoringdevice AT minjeonghee validationofawearablecufflesswristwatchtypebloodpressuremonitoringdevice AT leehaeyoung validationofawearablecufflesswristwatchtypebloodpressuremonitoringdevice AT limsoo validationofawearablecufflesswristwatchtypebloodpressuremonitoringdevice |