Analytical Efficacy of a Gas Mixer and Stabilizer for Laser Ablation ICP Mass Spectrometry

[Image: see text] The analytical efficacy of five gas mixers and five stabilizers on signal stabilization and washout time obtained for laser ablation inductively coupled plasma mass spectrometry (LA-ICPMS) was evaluated in the present study. In the case of gas mixers examined, a total of 95 pattern...

Descripción completa

Detalles Bibliográficos
Autores principales: Kon, Yoshiaki, Yokoyama, Takaomi D., Ohata, Masaki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2020
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7643187/
https://www.ncbi.nlm.nih.gov/pubmed/33163789
http://dx.doi.org/10.1021/acsomega.0c03658
_version_ 1783606229355986944
author Kon, Yoshiaki
Yokoyama, Takaomi D.
Ohata, Masaki
author_facet Kon, Yoshiaki
Yokoyama, Takaomi D.
Ohata, Masaki
author_sort Kon, Yoshiaki
collection PubMed
description [Image: see text] The analytical efficacy of five gas mixers and five stabilizers on signal stabilization and washout time obtained for laser ablation inductively coupled plasma mass spectrometry (LA-ICPMS) was evaluated in the present study. In the case of gas mixers examined, a total of 95 patterns of their attitudes as well as different directions of the gas flows were examined, and it was found that the signal variation and the washout time were strongly dependent on these factors. Even in a simple Y-shaped fitting (Y-mixer), signal stability and washout time had large variations with respect to its different attitudes as well as gas flow directions. The shortest washout time for each gas mixer was almost the same ranging from 1.0 to 1.2 s. The signal variations observed were 11–15% of relative standard deviation (RSD) under optimum conditions for each gas mixer. The optimum condition of a Y-mixer for LA-ICPMS represented 11% RSD and 1.0 s for signal variation and washout time, respectively. In the case of stabilizers examined, almost all stabilizers improved signal variations from 11 to 3.0–9.3%, but washout times became longer than those of the only Y-mixer from 1.0 to 1.2–8.9 s. The important thing is that the signal stability and the washout time are trade-off correlations for gas mixers and stabilizers. A suitable gas mixer or a stabilizer on the trade-off line can be selected with respect to different applications. It was also observed that variations of both signal stability and washout time correlated with the volume of stabilizers despite their different inner structures; that is, a stable signal and longer washout time seemed to be observed when the volume of the chamber became larger. This suggested that the signal stabilization obtained by stabilizers was ascribable to not only elimination of larger particles from laser ablation but also particle mixing effect, which compensated signal variation.
format Online
Article
Text
id pubmed-7643187
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher American Chemical Society
record_format MEDLINE/PubMed
spelling pubmed-76431872020-11-06 Analytical Efficacy of a Gas Mixer and Stabilizer for Laser Ablation ICP Mass Spectrometry Kon, Yoshiaki Yokoyama, Takaomi D. Ohata, Masaki ACS Omega [Image: see text] The analytical efficacy of five gas mixers and five stabilizers on signal stabilization and washout time obtained for laser ablation inductively coupled plasma mass spectrometry (LA-ICPMS) was evaluated in the present study. In the case of gas mixers examined, a total of 95 patterns of their attitudes as well as different directions of the gas flows were examined, and it was found that the signal variation and the washout time were strongly dependent on these factors. Even in a simple Y-shaped fitting (Y-mixer), signal stability and washout time had large variations with respect to its different attitudes as well as gas flow directions. The shortest washout time for each gas mixer was almost the same ranging from 1.0 to 1.2 s. The signal variations observed were 11–15% of relative standard deviation (RSD) under optimum conditions for each gas mixer. The optimum condition of a Y-mixer for LA-ICPMS represented 11% RSD and 1.0 s for signal variation and washout time, respectively. In the case of stabilizers examined, almost all stabilizers improved signal variations from 11 to 3.0–9.3%, but washout times became longer than those of the only Y-mixer from 1.0 to 1.2–8.9 s. The important thing is that the signal stability and the washout time are trade-off correlations for gas mixers and stabilizers. A suitable gas mixer or a stabilizer on the trade-off line can be selected with respect to different applications. It was also observed that variations of both signal stability and washout time correlated with the volume of stabilizers despite their different inner structures; that is, a stable signal and longer washout time seemed to be observed when the volume of the chamber became larger. This suggested that the signal stabilization obtained by stabilizers was ascribable to not only elimination of larger particles from laser ablation but also particle mixing effect, which compensated signal variation. American Chemical Society 2020-10-23 /pmc/articles/PMC7643187/ /pubmed/33163789 http://dx.doi.org/10.1021/acsomega.0c03658 Text en © 2020 The Authors. Published by American Chemical Society This is an open access article published under a Creative Commons Attribution (CC-BY) License (http://pubs.acs.org/page/policy/authorchoice_ccby_termsofuse.html) , which permits unrestricted use, distribution and reproduction in any medium, provided the author and source are cited.
spellingShingle Kon, Yoshiaki
Yokoyama, Takaomi D.
Ohata, Masaki
Analytical Efficacy of a Gas Mixer and Stabilizer for Laser Ablation ICP Mass Spectrometry
title Analytical Efficacy of a Gas Mixer and Stabilizer for Laser Ablation ICP Mass Spectrometry
title_full Analytical Efficacy of a Gas Mixer and Stabilizer for Laser Ablation ICP Mass Spectrometry
title_fullStr Analytical Efficacy of a Gas Mixer and Stabilizer for Laser Ablation ICP Mass Spectrometry
title_full_unstemmed Analytical Efficacy of a Gas Mixer and Stabilizer for Laser Ablation ICP Mass Spectrometry
title_short Analytical Efficacy of a Gas Mixer and Stabilizer for Laser Ablation ICP Mass Spectrometry
title_sort analytical efficacy of a gas mixer and stabilizer for laser ablation icp mass spectrometry
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7643187/
https://www.ncbi.nlm.nih.gov/pubmed/33163789
http://dx.doi.org/10.1021/acsomega.0c03658
work_keys_str_mv AT konyoshiaki analyticalefficacyofagasmixerandstabilizerforlaserablationicpmassspectrometry
AT yokoyamatakaomid analyticalefficacyofagasmixerandstabilizerforlaserablationicpmassspectrometry
AT ohatamasaki analyticalefficacyofagasmixerandstabilizerforlaserablationicpmassspectrometry