Agreement Between Spatiotemporal Gait Parameters Measured by a Markerless Motion Capture System and Two Reference Systems—a Treadmill-Based Photoelectric Cell and High-Speed Video Analyses: Comparative Study
BACKGROUND: Markerless systems to capture body motion require no markers to be attached to the body, thereby improving clinical feasibility and testing time. However, the lack of markers might affect the accuracy of measurements. OBJECTIVE: This study aimed to determine the absolute reliability and...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
JMIR Publications
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7647810/ https://www.ncbi.nlm.nih.gov/pubmed/33095181 http://dx.doi.org/10.2196/19498 |
_version_ | 1783606986003185664 |
---|---|
author | García-Pinillos, Felipe Jaén-Carrillo, Diego Soto Hermoso, Victor Latorre Román, Pedro Delgado, Pedro Martinez, Cristian Carton, Antonio Roche Seruendo, Luis |
author_facet | García-Pinillos, Felipe Jaén-Carrillo, Diego Soto Hermoso, Victor Latorre Román, Pedro Delgado, Pedro Martinez, Cristian Carton, Antonio Roche Seruendo, Luis |
author_sort | García-Pinillos, Felipe |
collection | PubMed |
description | BACKGROUND: Markerless systems to capture body motion require no markers to be attached to the body, thereby improving clinical feasibility and testing time. However, the lack of markers might affect the accuracy of measurements. OBJECTIVE: This study aimed to determine the absolute reliability and concurrent validity of the Kinect system with MotionMetrix software for spatiotemporal variables during running at a comfortable velocity, by comparing data between the combination system and two widely used systems—OptoGait and high-speed video analysis at 1000 Hz. METHODS: In total, 25 runners followed a running protocol on a treadmill at a speed of 12 km/h. The Kinect+MotionMetrix combination measured spatiotemporal parameters during running (ie, contact time, flight time, step frequency, and step length), which were compared to those obtained from two reference systems. RESULTS: Regardless of the system, flight time had the highest coefficients of variation (OptoGait: 16.4%; video analysis: 17.3%; Kinect+MotionMetrix: 23.2%). The rest of the coefficients of variation reported were lower than 8.1%. Correlation analysis showed very high correlations (r>0.8; P<.001) and almost perfect associations (intraclass correlation coefficient>0.81) between systems for all the spatiotemporal parameters except contact time, which had lower values. Bland-Altman plots revealed smaller systematic biases and random errors for step frequency and step length and larger systematic biases and random errors for temporal parameters with the Kinect+MotionMetrix system as compared to OptoGait (difference: contact time +3.0%, flight time −7.9%) and high-speed video analysis at 1000 Hz (difference: contact time +4.2%, flight time −11.3%). Accordingly, heteroscedasticity was found between systems for temporal parameters (r(2)>0.1). CONCLUSIONS: The results indicate that the Kinect+MotionMetrix combination slightly overestimates contact time and strongly underestimates flight time as compared to the OptoGait system and high-speed video analysis at 1000 Hz. However, it is a valid tool for measuring step frequency and step length when compared to reference systems. Future studies should determine the reliability of this system for determining temporal parameters. |
format | Online Article Text |
id | pubmed-7647810 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | JMIR Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-76478102020-11-17 Agreement Between Spatiotemporal Gait Parameters Measured by a Markerless Motion Capture System and Two Reference Systems—a Treadmill-Based Photoelectric Cell and High-Speed Video Analyses: Comparative Study García-Pinillos, Felipe Jaén-Carrillo, Diego Soto Hermoso, Victor Latorre Román, Pedro Delgado, Pedro Martinez, Cristian Carton, Antonio Roche Seruendo, Luis JMIR Mhealth Uhealth Original Paper BACKGROUND: Markerless systems to capture body motion require no markers to be attached to the body, thereby improving clinical feasibility and testing time. However, the lack of markers might affect the accuracy of measurements. OBJECTIVE: This study aimed to determine the absolute reliability and concurrent validity of the Kinect system with MotionMetrix software for spatiotemporal variables during running at a comfortable velocity, by comparing data between the combination system and two widely used systems—OptoGait and high-speed video analysis at 1000 Hz. METHODS: In total, 25 runners followed a running protocol on a treadmill at a speed of 12 km/h. The Kinect+MotionMetrix combination measured spatiotemporal parameters during running (ie, contact time, flight time, step frequency, and step length), which were compared to those obtained from two reference systems. RESULTS: Regardless of the system, flight time had the highest coefficients of variation (OptoGait: 16.4%; video analysis: 17.3%; Kinect+MotionMetrix: 23.2%). The rest of the coefficients of variation reported were lower than 8.1%. Correlation analysis showed very high correlations (r>0.8; P<.001) and almost perfect associations (intraclass correlation coefficient>0.81) between systems for all the spatiotemporal parameters except contact time, which had lower values. Bland-Altman plots revealed smaller systematic biases and random errors for step frequency and step length and larger systematic biases and random errors for temporal parameters with the Kinect+MotionMetrix system as compared to OptoGait (difference: contact time +3.0%, flight time −7.9%) and high-speed video analysis at 1000 Hz (difference: contact time +4.2%, flight time −11.3%). Accordingly, heteroscedasticity was found between systems for temporal parameters (r(2)>0.1). CONCLUSIONS: The results indicate that the Kinect+MotionMetrix combination slightly overestimates contact time and strongly underestimates flight time as compared to the OptoGait system and high-speed video analysis at 1000 Hz. However, it is a valid tool for measuring step frequency and step length when compared to reference systems. Future studies should determine the reliability of this system for determining temporal parameters. JMIR Publications 2020-10-23 /pmc/articles/PMC7647810/ /pubmed/33095181 http://dx.doi.org/10.2196/19498 Text en ©Felipe García-Pinillos, Diego Jaén-Carrillo, Victor Soto Hermoso, Pedro Latorre Román, Pedro Delgado, Cristian Martinez, Antonio Carton, Luis Roche Seruendo. Originally published in JMIR mHealth and uHealth (http://mhealth.jmir.org), 23.10.2020. https://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mHealth and uHealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included. |
spellingShingle | Original Paper García-Pinillos, Felipe Jaén-Carrillo, Diego Soto Hermoso, Victor Latorre Román, Pedro Delgado, Pedro Martinez, Cristian Carton, Antonio Roche Seruendo, Luis Agreement Between Spatiotemporal Gait Parameters Measured by a Markerless Motion Capture System and Two Reference Systems—a Treadmill-Based Photoelectric Cell and High-Speed Video Analyses: Comparative Study |
title | Agreement Between Spatiotemporal Gait Parameters Measured by a Markerless Motion Capture System and Two Reference Systems—a Treadmill-Based Photoelectric Cell and High-Speed Video Analyses: Comparative Study |
title_full | Agreement Between Spatiotemporal Gait Parameters Measured by a Markerless Motion Capture System and Two Reference Systems—a Treadmill-Based Photoelectric Cell and High-Speed Video Analyses: Comparative Study |
title_fullStr | Agreement Between Spatiotemporal Gait Parameters Measured by a Markerless Motion Capture System and Two Reference Systems—a Treadmill-Based Photoelectric Cell and High-Speed Video Analyses: Comparative Study |
title_full_unstemmed | Agreement Between Spatiotemporal Gait Parameters Measured by a Markerless Motion Capture System and Two Reference Systems—a Treadmill-Based Photoelectric Cell and High-Speed Video Analyses: Comparative Study |
title_short | Agreement Between Spatiotemporal Gait Parameters Measured by a Markerless Motion Capture System and Two Reference Systems—a Treadmill-Based Photoelectric Cell and High-Speed Video Analyses: Comparative Study |
title_sort | agreement between spatiotemporal gait parameters measured by a markerless motion capture system and two reference systems—a treadmill-based photoelectric cell and high-speed video analyses: comparative study |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7647810/ https://www.ncbi.nlm.nih.gov/pubmed/33095181 http://dx.doi.org/10.2196/19498 |
work_keys_str_mv | AT garciapinillosfelipe agreementbetweenspatiotemporalgaitparametersmeasuredbyamarkerlessmotioncapturesystemandtworeferencesystemsatreadmillbasedphotoelectriccellandhighspeedvideoanalysescomparativestudy AT jaencarrillodiego agreementbetweenspatiotemporalgaitparametersmeasuredbyamarkerlessmotioncapturesystemandtworeferencesystemsatreadmillbasedphotoelectriccellandhighspeedvideoanalysescomparativestudy AT sotohermosovictor agreementbetweenspatiotemporalgaitparametersmeasuredbyamarkerlessmotioncapturesystemandtworeferencesystemsatreadmillbasedphotoelectriccellandhighspeedvideoanalysescomparativestudy AT latorreromanpedro agreementbetweenspatiotemporalgaitparametersmeasuredbyamarkerlessmotioncapturesystemandtworeferencesystemsatreadmillbasedphotoelectriccellandhighspeedvideoanalysescomparativestudy AT delgadopedro agreementbetweenspatiotemporalgaitparametersmeasuredbyamarkerlessmotioncapturesystemandtworeferencesystemsatreadmillbasedphotoelectriccellandhighspeedvideoanalysescomparativestudy AT martinezcristian agreementbetweenspatiotemporalgaitparametersmeasuredbyamarkerlessmotioncapturesystemandtworeferencesystemsatreadmillbasedphotoelectriccellandhighspeedvideoanalysescomparativestudy AT cartonantonio agreementbetweenspatiotemporalgaitparametersmeasuredbyamarkerlessmotioncapturesystemandtworeferencesystemsatreadmillbasedphotoelectriccellandhighspeedvideoanalysescomparativestudy AT rocheseruendoluis agreementbetweenspatiotemporalgaitparametersmeasuredbyamarkerlessmotioncapturesystemandtworeferencesystemsatreadmillbasedphotoelectriccellandhighspeedvideoanalysescomparativestudy |