Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens

INTRODUCTION: Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a...

Descripción completa

Detalles Bibliográficos
Autores principales: Stohr, Joep J.J.M., Wennekes, Meike, van der Ent, Martijn, Diederen, Bram M.W., Kluytmans-van den Bergh, Marjolein F.Q., Bergmans, Anneke M.C., Kluytmans, Jan A.J.W., Pas, Suzan D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier B.V. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7648882/
https://www.ncbi.nlm.nih.gov/pubmed/33221622
http://dx.doi.org/10.1016/j.jcv.2020.104686
_version_ 1783607202063319040
author Stohr, Joep J.J.M.
Wennekes, Meike
van der Ent, Martijn
Diederen, Bram M.W.
Kluytmans-van den Bergh, Marjolein F.Q.
Bergmans, Anneke M.C.
Kluytmans, Jan A.J.W.
Pas, Suzan D.
author_facet Stohr, Joep J.J.M.
Wennekes, Meike
van der Ent, Martijn
Diederen, Bram M.W.
Kluytmans-van den Bergh, Marjolein F.Q.
Bergmans, Anneke M.C.
Kluytmans, Jan A.J.W.
Pas, Suzan D.
author_sort Stohr, Joep J.J.M.
collection PubMed
description INTRODUCTION: Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a lab-developed assay (LDA) and the cobas impact of freeze-thawing combined with lysis buffer. METHODS: Both retrospective and prospectively selected deep respiratory samples and oro- and nasopharyngeal samples in either E-swab® or GLY- were tested using the SARS-CoV-2 assay on the cobas®6800 System and compared to a lab developed assay. Additonally, SARS-CoV-2 RNA stability was assessed after one freeze-thaw cycle with or without lysis buffer. RESULTS: In total, 221 (58.3 %) oro- and nasopharyngeal swabs, 131 (34.6 %) deep respiratory specimens, and n = 25 (6.6 %) swabs of unknown origin were included to study clinical performance. Only 4 samples gave discrepant results, all being positive in the LDA and not the cobas®6800 system. For stability testing, 66 samples without and 110 with lysis buffer were included. No clinically significant difference was found in test results after one freeze-thaw cycle and addition of lysis buffer. CONCLUSION: Based on our findings, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results as the LDA in oro-/nasopharyngeal swabs and deep respiratory specimens. Moreover, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results before and after a freeze-thaw cycle, with better preservation of low viral loads in lysis buffer.
format Online
Article
Text
id pubmed-7648882
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Elsevier B.V.
record_format MEDLINE/PubMed
spelling pubmed-76488822020-11-09 Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens Stohr, Joep J.J.M. Wennekes, Meike van der Ent, Martijn Diederen, Bram M.W. Kluytmans-van den Bergh, Marjolein F.Q. Bergmans, Anneke M.C. Kluytmans, Jan A.J.W. Pas, Suzan D. J Clin Virol Article INTRODUCTION: Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a lab-developed assay (LDA) and the cobas impact of freeze-thawing combined with lysis buffer. METHODS: Both retrospective and prospectively selected deep respiratory samples and oro- and nasopharyngeal samples in either E-swab® or GLY- were tested using the SARS-CoV-2 assay on the cobas®6800 System and compared to a lab developed assay. Additonally, SARS-CoV-2 RNA stability was assessed after one freeze-thaw cycle with or without lysis buffer. RESULTS: In total, 221 (58.3 %) oro- and nasopharyngeal swabs, 131 (34.6 %) deep respiratory specimens, and n = 25 (6.6 %) swabs of unknown origin were included to study clinical performance. Only 4 samples gave discrepant results, all being positive in the LDA and not the cobas®6800 system. For stability testing, 66 samples without and 110 with lysis buffer were included. No clinically significant difference was found in test results after one freeze-thaw cycle and addition of lysis buffer. CONCLUSION: Based on our findings, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results as the LDA in oro-/nasopharyngeal swabs and deep respiratory specimens. Moreover, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results before and after a freeze-thaw cycle, with better preservation of low viral loads in lysis buffer. Elsevier B.V. 2020-12 2020-11-08 /pmc/articles/PMC7648882/ /pubmed/33221622 http://dx.doi.org/10.1016/j.jcv.2020.104686 Text en © 2020 Elsevier B.V. All rights reserved. Since January 2020 Elsevier has created a COVID-19 resource centre with free information in English and Mandarin on the novel coronavirus COVID-19. The COVID-19 resource centre is hosted on Elsevier Connect, the company's public news and information website. Elsevier hereby grants permission to make all its COVID-19-related research that is available on the COVID-19 resource centre - including this research content - immediately available in PubMed Central and other publicly funded repositories, such as the WHO COVID database with rights for unrestricted research re-use and analyses in any form or by any means with acknowledgement of the original source. These permissions are granted for free by Elsevier for as long as the COVID-19 resource centre remains active.
spellingShingle Article
Stohr, Joep J.J.M.
Wennekes, Meike
van der Ent, Martijn
Diederen, Bram M.W.
Kluytmans-van den Bergh, Marjolein F.Q.
Bergmans, Anneke M.C.
Kluytmans, Jan A.J.W.
Pas, Suzan D.
Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens
title Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens
title_full Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens
title_fullStr Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens
title_full_unstemmed Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens
title_short Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens
title_sort clinical performance and sample freeze-thaw stability of the cobas®6800 sars-cov-2 assay for the detection of sars-cov-2 in oro-/nasopharyngeal swabs and lower respiratory specimens
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7648882/
https://www.ncbi.nlm.nih.gov/pubmed/33221622
http://dx.doi.org/10.1016/j.jcv.2020.104686
work_keys_str_mv AT stohrjoepjjm clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens
AT wennekesmeike clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens
AT vanderentmartijn clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens
AT diederenbrammw clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens
AT kluytmansvandenberghmarjoleinfq clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens
AT bergmansannekemc clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens
AT kluytmansjanajw clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens
AT passuzand clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens