Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens
INTRODUCTION: Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier B.V.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7648882/ https://www.ncbi.nlm.nih.gov/pubmed/33221622 http://dx.doi.org/10.1016/j.jcv.2020.104686 |
_version_ | 1783607202063319040 |
---|---|
author | Stohr, Joep J.J.M. Wennekes, Meike van der Ent, Martijn Diederen, Bram M.W. Kluytmans-van den Bergh, Marjolein F.Q. Bergmans, Anneke M.C. Kluytmans, Jan A.J.W. Pas, Suzan D. |
author_facet | Stohr, Joep J.J.M. Wennekes, Meike van der Ent, Martijn Diederen, Bram M.W. Kluytmans-van den Bergh, Marjolein F.Q. Bergmans, Anneke M.C. Kluytmans, Jan A.J.W. Pas, Suzan D. |
author_sort | Stohr, Joep J.J.M. |
collection | PubMed |
description | INTRODUCTION: Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a lab-developed assay (LDA) and the cobas impact of freeze-thawing combined with lysis buffer. METHODS: Both retrospective and prospectively selected deep respiratory samples and oro- and nasopharyngeal samples in either E-swab® or GLY- were tested using the SARS-CoV-2 assay on the cobas®6800 System and compared to a lab developed assay. Additonally, SARS-CoV-2 RNA stability was assessed after one freeze-thaw cycle with or without lysis buffer. RESULTS: In total, 221 (58.3 %) oro- and nasopharyngeal swabs, 131 (34.6 %) deep respiratory specimens, and n = 25 (6.6 %) swabs of unknown origin were included to study clinical performance. Only 4 samples gave discrepant results, all being positive in the LDA and not the cobas®6800 system. For stability testing, 66 samples without and 110 with lysis buffer were included. No clinically significant difference was found in test results after one freeze-thaw cycle and addition of lysis buffer. CONCLUSION: Based on our findings, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results as the LDA in oro-/nasopharyngeal swabs and deep respiratory specimens. Moreover, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results before and after a freeze-thaw cycle, with better preservation of low viral loads in lysis buffer. |
format | Online Article Text |
id | pubmed-7648882 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Elsevier B.V. |
record_format | MEDLINE/PubMed |
spelling | pubmed-76488822020-11-09 Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens Stohr, Joep J.J.M. Wennekes, Meike van der Ent, Martijn Diederen, Bram M.W. Kluytmans-van den Bergh, Marjolein F.Q. Bergmans, Anneke M.C. Kluytmans, Jan A.J.W. Pas, Suzan D. J Clin Virol Article INTRODUCTION: Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a lab-developed assay (LDA) and the cobas impact of freeze-thawing combined with lysis buffer. METHODS: Both retrospective and prospectively selected deep respiratory samples and oro- and nasopharyngeal samples in either E-swab® or GLY- were tested using the SARS-CoV-2 assay on the cobas®6800 System and compared to a lab developed assay. Additonally, SARS-CoV-2 RNA stability was assessed after one freeze-thaw cycle with or without lysis buffer. RESULTS: In total, 221 (58.3 %) oro- and nasopharyngeal swabs, 131 (34.6 %) deep respiratory specimens, and n = 25 (6.6 %) swabs of unknown origin were included to study clinical performance. Only 4 samples gave discrepant results, all being positive in the LDA and not the cobas®6800 system. For stability testing, 66 samples without and 110 with lysis buffer were included. No clinically significant difference was found in test results after one freeze-thaw cycle and addition of lysis buffer. CONCLUSION: Based on our findings, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results as the LDA in oro-/nasopharyngeal swabs and deep respiratory specimens. Moreover, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results before and after a freeze-thaw cycle, with better preservation of low viral loads in lysis buffer. Elsevier B.V. 2020-12 2020-11-08 /pmc/articles/PMC7648882/ /pubmed/33221622 http://dx.doi.org/10.1016/j.jcv.2020.104686 Text en © 2020 Elsevier B.V. All rights reserved. Since January 2020 Elsevier has created a COVID-19 resource centre with free information in English and Mandarin on the novel coronavirus COVID-19. The COVID-19 resource centre is hosted on Elsevier Connect, the company's public news and information website. Elsevier hereby grants permission to make all its COVID-19-related research that is available on the COVID-19 resource centre - including this research content - immediately available in PubMed Central and other publicly funded repositories, such as the WHO COVID database with rights for unrestricted research re-use and analyses in any form or by any means with acknowledgement of the original source. These permissions are granted for free by Elsevier for as long as the COVID-19 resource centre remains active. |
spellingShingle | Article Stohr, Joep J.J.M. Wennekes, Meike van der Ent, Martijn Diederen, Bram M.W. Kluytmans-van den Bergh, Marjolein F.Q. Bergmans, Anneke M.C. Kluytmans, Jan A.J.W. Pas, Suzan D. Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens |
title | Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens |
title_full | Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens |
title_fullStr | Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens |
title_full_unstemmed | Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens |
title_short | Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens |
title_sort | clinical performance and sample freeze-thaw stability of the cobas®6800 sars-cov-2 assay for the detection of sars-cov-2 in oro-/nasopharyngeal swabs and lower respiratory specimens |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7648882/ https://www.ncbi.nlm.nih.gov/pubmed/33221622 http://dx.doi.org/10.1016/j.jcv.2020.104686 |
work_keys_str_mv | AT stohrjoepjjm clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens AT wennekesmeike clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens AT vanderentmartijn clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens AT diederenbrammw clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens AT kluytmansvandenberghmarjoleinfq clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens AT bergmansannekemc clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens AT kluytmansjanajw clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens AT passuzand clinicalperformanceandsamplefreezethawstabilityofthecobas6800sarscov2assayforthedetectionofsarscov2inoronasopharyngealswabsandlowerrespiratoryspecimens |