Cargando…

Protection of the Femoral Trunnion During Acetabular Revision Hip Surgery—A Novel Technique

Revision of a failed acetabular component in the presence of an intact femoral prosthesis comprises approximately 20% of revision total hip replacements performed by the senior author. During this procedure, the femoral trunnion is at risk of damage because of scratching from retractors, reamers, or...

Descripción completa

Detalles Bibliográficos
Autores principales: Kawalkar, Abhijit C., Vertzyas, Nick, Neil, Michael J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7649106/
https://www.ncbi.nlm.nih.gov/pubmed/33204787
http://dx.doi.org/10.1016/j.artd.2020.09.015
Descripción
Sumario:Revision of a failed acetabular component in the presence of an intact femoral prosthesis comprises approximately 20% of revision total hip replacements performed by the senior author. During this procedure, the femoral trunnion is at risk of damage because of scratching from retractors, reamers, or other instruments. Scratching can result in mechanically assisted crevice corrosion or adverse local-tissue reaction or trunnionosis leading to implant failure. We have used a variety of techniques for protection of the femoral taper over the last 30 years. We describe our current method of taper protection, using a cut segment of #5 silicon laryngeal airway, which is a simple and effective method.