Plane wave versus focused transmissions for contrast enhanced ultrasound imaging: the role of parameter settings and the effects of flow rate on contrast measurements
Contrast enhanced ultrasound (CEUS) and dynamic contrast enhanced ultrasound (DCE-US) can be used to provide information about the vasculature aiding diagnosis and monitoring of a number of pathologies including cancer. In the development of a CEUS imaging system, there are many choices to be made,...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
IOP Publishing
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7655116/ https://www.ncbi.nlm.nih.gov/pubmed/30917360 http://dx.doi.org/10.1088/1361-6560/ab13f2 |
_version_ | 1783608173947518976 |
---|---|
author | Moghimirad, Elahe Bamber, Jeffrey Harris, Emma |
author_facet | Moghimirad, Elahe Bamber, Jeffrey Harris, Emma |
author_sort | Moghimirad, Elahe |
collection | PubMed |
description | Contrast enhanced ultrasound (CEUS) and dynamic contrast enhanced ultrasound (DCE-US) can be used to provide information about the vasculature aiding diagnosis and monitoring of a number of pathologies including cancer. In the development of a CEUS imaging system, there are many choices to be made, such as whether to use plane wave (PW) or focused imaging (FI), and the values for parameters such as transmit frequency, F-number, mechanical index, and number of compounding angles (for PW imaging). CEUS image contrast may also be dependent on subject characteristics, e.g. flow speed and vessel orientation. We evaluated the effect of such choices on vessel contrast for PW and FI in vitro, using 2D ultrasound imaging. CEUS images were obtained using a Vantage(TM) (Verasonics Inc.) and a pulse-inversion (PI) algorithm on a flow phantom. Contrast (C) and contrast reduction (CR) were calculated, where C was the initial ratio of signal in vessel to signal in background and CR was its reduction after 200 frames (acquired in 20 s). Two transducer orientations were used: parallel and perpendicular to the vessel direction. Similar C and CR was achievable for PW and FI by choosing optimal parameter values. PW imaging suffered from high frequency grating lobe artefacts, which may lead to degraded image quality and misinterpretation of data. Flow rate influenced the contrast based on: (1) false contrast increase due to the bubble motion between the PI positive and negative pulses (for both PW and FI), and (2) contrast reduction due to the incoherency caused by bubble motion between the compounding angles (for PW only). The effects were less pronounced for perpendicular transducer orientation compared to a parallel one. Although both effects are undesirable, it may be more straight forward to account for artefacts in FI as it only suffers from the former effect. In conclusion, if higher frame rate imaging is not required (a benefit of PW), FI appears to be a better choice of imaging mode for CEUS, providing greater image quality over PW for similar rates of contrast reduction. |
format | Online Article Text |
id | pubmed-7655116 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | IOP Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-76551162020-11-12 Plane wave versus focused transmissions for contrast enhanced ultrasound imaging: the role of parameter settings and the effects of flow rate on contrast measurements Moghimirad, Elahe Bamber, Jeffrey Harris, Emma Phys Med Biol Paper Contrast enhanced ultrasound (CEUS) and dynamic contrast enhanced ultrasound (DCE-US) can be used to provide information about the vasculature aiding diagnosis and monitoring of a number of pathologies including cancer. In the development of a CEUS imaging system, there are many choices to be made, such as whether to use plane wave (PW) or focused imaging (FI), and the values for parameters such as transmit frequency, F-number, mechanical index, and number of compounding angles (for PW imaging). CEUS image contrast may also be dependent on subject characteristics, e.g. flow speed and vessel orientation. We evaluated the effect of such choices on vessel contrast for PW and FI in vitro, using 2D ultrasound imaging. CEUS images were obtained using a Vantage(TM) (Verasonics Inc.) and a pulse-inversion (PI) algorithm on a flow phantom. Contrast (C) and contrast reduction (CR) were calculated, where C was the initial ratio of signal in vessel to signal in background and CR was its reduction after 200 frames (acquired in 20 s). Two transducer orientations were used: parallel and perpendicular to the vessel direction. Similar C and CR was achievable for PW and FI by choosing optimal parameter values. PW imaging suffered from high frequency grating lobe artefacts, which may lead to degraded image quality and misinterpretation of data. Flow rate influenced the contrast based on: (1) false contrast increase due to the bubble motion between the PI positive and negative pulses (for both PW and FI), and (2) contrast reduction due to the incoherency caused by bubble motion between the compounding angles (for PW only). The effects were less pronounced for perpendicular transducer orientation compared to a parallel one. Although both effects are undesirable, it may be more straight forward to account for artefacts in FI as it only suffers from the former effect. In conclusion, if higher frame rate imaging is not required (a benefit of PW), FI appears to be a better choice of imaging mode for CEUS, providing greater image quality over PW for similar rates of contrast reduction. IOP Publishing 2019-05 2019-04-23 /pmc/articles/PMC7655116/ /pubmed/30917360 http://dx.doi.org/10.1088/1361-6560/ab13f2 Text en © 2019 Institute of Physics and Engineering in Medicine http://creativecommons.org/licenses/by/3.0/ Original content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence (http://creativecommons.org/licenses/by/3.0) . Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. |
spellingShingle | Paper Moghimirad, Elahe Bamber, Jeffrey Harris, Emma Plane wave versus focused transmissions for contrast enhanced ultrasound imaging: the role of parameter settings and the effects of flow rate on contrast measurements |
title | Plane wave versus focused transmissions for contrast enhanced
ultrasound imaging: the role of parameter settings and the effects of flow rate
on contrast measurements |
title_full | Plane wave versus focused transmissions for contrast enhanced
ultrasound imaging: the role of parameter settings and the effects of flow rate
on contrast measurements |
title_fullStr | Plane wave versus focused transmissions for contrast enhanced
ultrasound imaging: the role of parameter settings and the effects of flow rate
on contrast measurements |
title_full_unstemmed | Plane wave versus focused transmissions for contrast enhanced
ultrasound imaging: the role of parameter settings and the effects of flow rate
on contrast measurements |
title_short | Plane wave versus focused transmissions for contrast enhanced
ultrasound imaging: the role of parameter settings and the effects of flow rate
on contrast measurements |
title_sort | plane wave versus focused transmissions for contrast enhanced
ultrasound imaging: the role of parameter settings and the effects of flow rate
on contrast measurements |
topic | Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7655116/ https://www.ncbi.nlm.nih.gov/pubmed/30917360 http://dx.doi.org/10.1088/1361-6560/ab13f2 |
work_keys_str_mv | AT moghimiradelahe planewaveversusfocusedtransmissionsforcontrastenhancedultrasoundimagingtheroleofparametersettingsandtheeffectsofflowrateoncontrastmeasurements AT bamberjeffrey planewaveversusfocusedtransmissionsforcontrastenhancedultrasoundimagingtheroleofparametersettingsandtheeffectsofflowrateoncontrastmeasurements AT harrisemma planewaveversusfocusedtransmissionsforcontrastenhancedultrasoundimagingtheroleofparametersettingsandtheeffectsofflowrateoncontrastmeasurements |