Cargando…

Reducing ECG Artifact From Left Ventricular Assist Device Electromagnetic Interference

BACKGROUND: Left ventricular assist devices (LVADs) generate electromagnetic interference that causes high‐frequency noise artifacts on 12‐lead ECGs. We describe the causes of this interference and potential solutions to aid ECG interpretation in patients with LVAD. METHODS AND RESULTS: Waveform dat...

Descripción completa

Detalles Bibliográficos
Autores principales: Loring, Zak, Sen, Sounok, Black-Maier, Eric, Atwater, Brett D., Russell, Stuart D., DeVore, Adam D., Piccini, Jonathan P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7660795/
https://www.ncbi.nlm.nih.gov/pubmed/32787630
http://dx.doi.org/10.1161/JAHA.120.017563
_version_ 1783609083501215744
author Loring, Zak
Sen, Sounok
Black-Maier, Eric
Atwater, Brett D.
Russell, Stuart D.
DeVore, Adam D.
Piccini, Jonathan P.
author_facet Loring, Zak
Sen, Sounok
Black-Maier, Eric
Atwater, Brett D.
Russell, Stuart D.
DeVore, Adam D.
Piccini, Jonathan P.
author_sort Loring, Zak
collection PubMed
description BACKGROUND: Left ventricular assist devices (LVADs) generate electromagnetic interference that causes high‐frequency noise artifacts on 12‐lead ECGs. We describe the causes of this interference and potential solutions to aid ECG interpretation in patients with LVAD. METHODS AND RESULTS: Waveform data from ECGs performed before and after LVAD implantation were passed through a fast Fourier transform to identify LVAD‐related changes in the spectral profile. ECGs recorded in 9 patients with HeartMate II, HeartMate 3, and HeartWare LVADs were analyzed to identify the LVAD model‐specific spectral patterns. Waveform data were then passed through digital low‐pass and bandstop filters and redisplayed to evaluate the effect of filtering on LVAD‐related electromagnetic interference. The spectral profile of patients with HeartMate II and HeartMate 3 LVADs demonstrated a prominent signal at the device‐specific frequency of impeller rotation. In patients with the HeartMate 3 LVAD, 2 additional peaks were observed at the frequencies equivalent to the LVAD's artificial pulsatility rotational speeds. Patients with HeartWare devices demonstrated a prominent signal peak at a frequency equal to double their LVAD's set rotational speed. Applying a low‐pass filter to a value below the observed frequency peak from the LVAD significantly improved the waveform tracing and quality of the ECG. Applying a speed‐specific bandstop filter to remove the observed LVAD frequency peak also improved the clarity of the ECG without compromising physiological high‐frequency signal components. CONCLUSIONS: LVADs create impeller rotational speed‐specific electromagnetic interference that can be ameliorated by application of low‐pass or bandstop filters to improve ECG clarity.
format Online
Article
Text
id pubmed-7660795
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-76607952020-11-17 Reducing ECG Artifact From Left Ventricular Assist Device Electromagnetic Interference Loring, Zak Sen, Sounok Black-Maier, Eric Atwater, Brett D. Russell, Stuart D. DeVore, Adam D. Piccini, Jonathan P. J Am Heart Assoc Original Research BACKGROUND: Left ventricular assist devices (LVADs) generate electromagnetic interference that causes high‐frequency noise artifacts on 12‐lead ECGs. We describe the causes of this interference and potential solutions to aid ECG interpretation in patients with LVAD. METHODS AND RESULTS: Waveform data from ECGs performed before and after LVAD implantation were passed through a fast Fourier transform to identify LVAD‐related changes in the spectral profile. ECGs recorded in 9 patients with HeartMate II, HeartMate 3, and HeartWare LVADs were analyzed to identify the LVAD model‐specific spectral patterns. Waveform data were then passed through digital low‐pass and bandstop filters and redisplayed to evaluate the effect of filtering on LVAD‐related electromagnetic interference. The spectral profile of patients with HeartMate II and HeartMate 3 LVADs demonstrated a prominent signal at the device‐specific frequency of impeller rotation. In patients with the HeartMate 3 LVAD, 2 additional peaks were observed at the frequencies equivalent to the LVAD's artificial pulsatility rotational speeds. Patients with HeartWare devices demonstrated a prominent signal peak at a frequency equal to double their LVAD's set rotational speed. Applying a low‐pass filter to a value below the observed frequency peak from the LVAD significantly improved the waveform tracing and quality of the ECG. Applying a speed‐specific bandstop filter to remove the observed LVAD frequency peak also improved the clarity of the ECG without compromising physiological high‐frequency signal components. CONCLUSIONS: LVADs create impeller rotational speed‐specific electromagnetic interference that can be ameliorated by application of low‐pass or bandstop filters to improve ECG clarity. John Wiley and Sons Inc. 2020-08-13 /pmc/articles/PMC7660795/ /pubmed/32787630 http://dx.doi.org/10.1161/JAHA.120.017563 Text en © 2020 The Authors. Published on behalf of the American Heart Association, Inc., by Wiley. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.
spellingShingle Original Research
Loring, Zak
Sen, Sounok
Black-Maier, Eric
Atwater, Brett D.
Russell, Stuart D.
DeVore, Adam D.
Piccini, Jonathan P.
Reducing ECG Artifact From Left Ventricular Assist Device Electromagnetic Interference
title Reducing ECG Artifact From Left Ventricular Assist Device Electromagnetic Interference
title_full Reducing ECG Artifact From Left Ventricular Assist Device Electromagnetic Interference
title_fullStr Reducing ECG Artifact From Left Ventricular Assist Device Electromagnetic Interference
title_full_unstemmed Reducing ECG Artifact From Left Ventricular Assist Device Electromagnetic Interference
title_short Reducing ECG Artifact From Left Ventricular Assist Device Electromagnetic Interference
title_sort reducing ecg artifact from left ventricular assist device electromagnetic interference
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7660795/
https://www.ncbi.nlm.nih.gov/pubmed/32787630
http://dx.doi.org/10.1161/JAHA.120.017563
work_keys_str_mv AT loringzak reducingecgartifactfromleftventricularassistdeviceelectromagneticinterference
AT sensounok reducingecgartifactfromleftventricularassistdeviceelectromagneticinterference
AT blackmaiereric reducingecgartifactfromleftventricularassistdeviceelectromagneticinterference
AT atwaterbrettd reducingecgartifactfromleftventricularassistdeviceelectromagneticinterference
AT russellstuartd reducingecgartifactfromleftventricularassistdeviceelectromagneticinterference
AT devoreadamd reducingecgartifactfromleftventricularassistdeviceelectromagneticinterference
AT piccinijonathanp reducingecgartifactfromleftventricularassistdeviceelectromagneticinterference