Experimental evaluation of the deadtime phenomenon for GM detector: deadtime dependence on operating voltages

A detailed analysis of Geiger Mueller counter deadtime dependence on operating voltage is presented in the manuscript using four pairs of radiation sources. Based on two-source method, detector deadtime is calculated for a wide range of operating voltages which revealed a peculiar relationship betwe...

Descripción completa

Detalles Bibliográficos
Autores principales: Almutairi, Bader, Alam, Syed, Akyurek, Tayfun, Goodwin, Cameron S., Usman, Shoaib
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7672071/
https://www.ncbi.nlm.nih.gov/pubmed/33203933
http://dx.doi.org/10.1038/s41598-020-75310-3
Descripción
Sumario:A detailed analysis of Geiger Mueller counter deadtime dependence on operating voltage is presented in the manuscript using four pairs of radiation sources. Based on two-source method, detector deadtime is calculated for a wide range of operating voltages which revealed a peculiar relationship between the operating voltage and the detector deadtime. In the low voltage range, a distinct drop in deadtime was observed where deadtime reached a value as low as a few microseconds (22 µs for (204)Tl, 26 µs for (137)Cs, 9 µs for (22)Na). This sharp drop in the deadtime is possibly due to reduced recombination with increasing voltage. After the lowest point, the deadtime generally increased rapidly to reach a maximum (292 µs for (204)Tl, 277 µs for (137)Cs, 258 µs for (22)Na). This rapid increase in the deadtime is mainly due to the on-set of charge multiplication. After the maximum deadtime values, there was an exponential decrease in the deadtime reaching an asymptotic low where the manufacturer recommended voltage for operation falls. This pattern of deadtime voltage dependence was repeated for all sources tested with the exception of (54)Mn. Low count rates leading to a negative deadtime suggested poor statistical nature of the data collected for (54)Mn and the data while being presented here is not used for any inference.