Cargando…
Guide sheath breakage during endobronchial ultrasonography
This report describes a case of guide sheath breakage during endobronchial ultrasonography. While steering the guide sheath in the direction of a peripheral lung nodule using a guiding device/curette, the radiopaque band (RB) attached to the head of the guide sheath dislodged and remained in the per...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley & Sons, Ltd
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7678640/ https://www.ncbi.nlm.nih.gov/pubmed/33251013 http://dx.doi.org/10.1002/rcr2.690 |
_version_ | 1783612198885523456 |
---|---|
author | Yanagihara, Takahiro Yamaoka, Masatoshi Kawakami, Naoki Onuki, Takuya Inagaki, Masaharu |
author_facet | Yanagihara, Takahiro Yamaoka, Masatoshi Kawakami, Naoki Onuki, Takuya Inagaki, Masaharu |
author_sort | Yanagihara, Takahiro |
collection | PubMed |
description | This report describes a case of guide sheath breakage during endobronchial ultrasonography. While steering the guide sheath in the direction of a peripheral lung nodule using a guiding device/curette, the radiopaque band (RB) attached to the head of the guide sheath dislodged and remained in the peripheral bronchus near the tumour. The band could not be removed endoscopically. As the tumour was diagnosed as a colon cancer metastasis, we performed a partial lung resection to remove the RB and nodule together four months after bronchoscopy. |
format | Online Article Text |
id | pubmed-7678640 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | John Wiley & Sons, Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-76786402020-11-27 Guide sheath breakage during endobronchial ultrasonography Yanagihara, Takahiro Yamaoka, Masatoshi Kawakami, Naoki Onuki, Takuya Inagaki, Masaharu Respirol Case Rep Case Reports This report describes a case of guide sheath breakage during endobronchial ultrasonography. While steering the guide sheath in the direction of a peripheral lung nodule using a guiding device/curette, the radiopaque band (RB) attached to the head of the guide sheath dislodged and remained in the peripheral bronchus near the tumour. The band could not be removed endoscopically. As the tumour was diagnosed as a colon cancer metastasis, we performed a partial lung resection to remove the RB and nodule together four months after bronchoscopy. John Wiley & Sons, Ltd 2020-11-20 /pmc/articles/PMC7678640/ /pubmed/33251013 http://dx.doi.org/10.1002/rcr2.690 Text en © 2020 The Authors. Respirology Case Reports published by John Wiley & Sons Australia, Ltd on behalf of The Asian Pacific Society of Respirology. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Reports Yanagihara, Takahiro Yamaoka, Masatoshi Kawakami, Naoki Onuki, Takuya Inagaki, Masaharu Guide sheath breakage during endobronchial ultrasonography |
title |
Guide sheath breakage during endobronchial ultrasonography |
title_full |
Guide sheath breakage during endobronchial ultrasonography |
title_fullStr |
Guide sheath breakage during endobronchial ultrasonography |
title_full_unstemmed |
Guide sheath breakage during endobronchial ultrasonography |
title_short |
Guide sheath breakage during endobronchial ultrasonography |
title_sort | guide sheath breakage during endobronchial ultrasonography |
topic | Case Reports |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7678640/ https://www.ncbi.nlm.nih.gov/pubmed/33251013 http://dx.doi.org/10.1002/rcr2.690 |
work_keys_str_mv | AT yanagiharatakahiro guidesheathbreakageduringendobronchialultrasonography AT yamaokamasatoshi guidesheathbreakageduringendobronchialultrasonography AT kawakaminaoki guidesheathbreakageduringendobronchialultrasonography AT onukitakuya guidesheathbreakageduringendobronchialultrasonography AT inagakimasaharu guidesheathbreakageduringendobronchialultrasonography |