Guide sheath breakage during endobronchial ultrasonography

This report describes a case of guide sheath breakage during endobronchial ultrasonography. While steering the guide sheath in the direction of a peripheral lung nodule using a guiding device/curette, the radiopaque band (RB) attached to the head of the guide sheath dislodged and remained in the per...

Descripción completa

Detalles Bibliográficos
Autores principales: Yanagihara, Takahiro, Yamaoka, Masatoshi, Kawakami, Naoki, Onuki, Takuya, Inagaki, Masaharu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley & Sons, Ltd 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7678640/
https://www.ncbi.nlm.nih.gov/pubmed/33251013
http://dx.doi.org/10.1002/rcr2.690
_version_ 1783612198885523456
author Yanagihara, Takahiro
Yamaoka, Masatoshi
Kawakami, Naoki
Onuki, Takuya
Inagaki, Masaharu
author_facet Yanagihara, Takahiro
Yamaoka, Masatoshi
Kawakami, Naoki
Onuki, Takuya
Inagaki, Masaharu
author_sort Yanagihara, Takahiro
collection PubMed
description This report describes a case of guide sheath breakage during endobronchial ultrasonography. While steering the guide sheath in the direction of a peripheral lung nodule using a guiding device/curette, the radiopaque band (RB) attached to the head of the guide sheath dislodged and remained in the peripheral bronchus near the tumour. The band could not be removed endoscopically. As the tumour was diagnosed as a colon cancer metastasis, we performed a partial lung resection to remove the RB and nodule together four months after bronchoscopy.
format Online
Article
Text
id pubmed-7678640
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley & Sons, Ltd
record_format MEDLINE/PubMed
spelling pubmed-76786402020-11-27 Guide sheath breakage during endobronchial ultrasonography Yanagihara, Takahiro Yamaoka, Masatoshi Kawakami, Naoki Onuki, Takuya Inagaki, Masaharu Respirol Case Rep Case Reports This report describes a case of guide sheath breakage during endobronchial ultrasonography. While steering the guide sheath in the direction of a peripheral lung nodule using a guiding device/curette, the radiopaque band (RB) attached to the head of the guide sheath dislodged and remained in the peripheral bronchus near the tumour. The band could not be removed endoscopically. As the tumour was diagnosed as a colon cancer metastasis, we performed a partial lung resection to remove the RB and nodule together four months after bronchoscopy. John Wiley & Sons, Ltd 2020-11-20 /pmc/articles/PMC7678640/ /pubmed/33251013 http://dx.doi.org/10.1002/rcr2.690 Text en © 2020 The Authors. Respirology Case Reports published by John Wiley & Sons Australia, Ltd on behalf of The Asian Pacific Society of Respirology. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Reports
Yanagihara, Takahiro
Yamaoka, Masatoshi
Kawakami, Naoki
Onuki, Takuya
Inagaki, Masaharu
Guide sheath breakage during endobronchial ultrasonography
title Guide sheath breakage during endobronchial ultrasonography
title_full Guide sheath breakage during endobronchial ultrasonography
title_fullStr Guide sheath breakage during endobronchial ultrasonography
title_full_unstemmed Guide sheath breakage during endobronchial ultrasonography
title_short Guide sheath breakage during endobronchial ultrasonography
title_sort guide sheath breakage during endobronchial ultrasonography
topic Case Reports
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7678640/
https://www.ncbi.nlm.nih.gov/pubmed/33251013
http://dx.doi.org/10.1002/rcr2.690
work_keys_str_mv AT yanagiharatakahiro guidesheathbreakageduringendobronchialultrasonography
AT yamaokamasatoshi guidesheathbreakageduringendobronchialultrasonography
AT kawakaminaoki guidesheathbreakageduringendobronchialultrasonography
AT onukitakuya guidesheathbreakageduringendobronchialultrasonography
AT inagakimasaharu guidesheathbreakageduringendobronchialultrasonography