Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film

PURPOSE: To compare the repeatability of keratometry between different instruments in patients with hyperosmolar tear film and a control group. PATIENTS AND METHODS: Subjects with tear-film osmolarity of 316 mOsm/L or more in either eye or 308 m/Osm/L or lower in both eyes were assigned to the hyper...

Descripción completa

Detalles Bibliográficos
Autores principales: Gjerdrum, Bjørn, Gundersen, Kjell Gunnar, Lundmark, Per Olof, Aakre, Bente Monica
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7680682/
https://www.ncbi.nlm.nih.gov/pubmed/33239862
http://dx.doi.org/10.2147/OPTH.S280868
_version_ 1783612482455076864
author Gjerdrum, Bjørn
Gundersen, Kjell Gunnar
Lundmark, Per Olof
Aakre, Bente Monica
author_facet Gjerdrum, Bjørn
Gundersen, Kjell Gunnar
Lundmark, Per Olof
Aakre, Bente Monica
author_sort Gjerdrum, Bjørn
collection PubMed
description PURPOSE: To compare the repeatability of keratometry between different instruments in patients with hyperosmolar tear film and a control group. PATIENTS AND METHODS: Subjects with tear-film osmolarity of 316 mOsm/L or more in either eye or 308 m/Osm/L or lower in both eyes were assigned to the hyperosmolar and the control group, respectively. The test eye was the eye with higher osmolarity in the hyperosmolar group and randomly chosen in the control group. The repeatability of keratometry was compared between a reflectometry device (Haag-Streit Lenstar 900), a Scheimpflug device (Oculus Pentacam HR) and two optical coherence tomography (OCT) devices (Tomey Casia SS-1000 and Heidelberg Anterion), based on two measurements from each device. RESULTS: The study included 94 subjects (31 hyperosmolar and 63 controls). Both OCT devices had higher mean differences of average simulated keratometry (SimK) vs the Lenstar in both groups, though all differences in means were <0.07 D. The Casia had the highest mean vector difference of SimK astigmatism in the control group (differences in means <0.11 D). These differences of the instruments were statistically significant (p < 0.02), except for the Anterion in the control group. With all subjects, the coefficient of repeatability varied from 0.1 to 0.3 for average SimK (highest for both OCT devices) and from 0.4 to 0.7 for SimK astigmatism (highest for the Casia). Similar results were found for total corneal power (OCT devices compared to the Pentacam). CONCLUSION: Both OCT devices show more variability in average SimK and the Casia more variability in SimK astigmatism compared to the Lenstar and the Pentacam. However, the results suggested that repeatability was not influenced by osmolarity.
format Online
Article
Text
id pubmed-7680682
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Dove
record_format MEDLINE/PubMed
spelling pubmed-76806822020-11-24 Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film Gjerdrum, Bjørn Gundersen, Kjell Gunnar Lundmark, Per Olof Aakre, Bente Monica Clin Ophthalmol Original Research PURPOSE: To compare the repeatability of keratometry between different instruments in patients with hyperosmolar tear film and a control group. PATIENTS AND METHODS: Subjects with tear-film osmolarity of 316 mOsm/L or more in either eye or 308 m/Osm/L or lower in both eyes were assigned to the hyperosmolar and the control group, respectively. The test eye was the eye with higher osmolarity in the hyperosmolar group and randomly chosen in the control group. The repeatability of keratometry was compared between a reflectometry device (Haag-Streit Lenstar 900), a Scheimpflug device (Oculus Pentacam HR) and two optical coherence tomography (OCT) devices (Tomey Casia SS-1000 and Heidelberg Anterion), based on two measurements from each device. RESULTS: The study included 94 subjects (31 hyperosmolar and 63 controls). Both OCT devices had higher mean differences of average simulated keratometry (SimK) vs the Lenstar in both groups, though all differences in means were <0.07 D. The Casia had the highest mean vector difference of SimK astigmatism in the control group (differences in means <0.11 D). These differences of the instruments were statistically significant (p < 0.02), except for the Anterion in the control group. With all subjects, the coefficient of repeatability varied from 0.1 to 0.3 for average SimK (highest for both OCT devices) and from 0.4 to 0.7 for SimK astigmatism (highest for the Casia). Similar results were found for total corneal power (OCT devices compared to the Pentacam). CONCLUSION: Both OCT devices show more variability in average SimK and the Casia more variability in SimK astigmatism compared to the Lenstar and the Pentacam. However, the results suggested that repeatability was not influenced by osmolarity. Dove 2020-11-18 /pmc/articles/PMC7680682/ /pubmed/33239862 http://dx.doi.org/10.2147/OPTH.S280868 Text en © 2020 Gjerdrum et al. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
spellingShingle Original Research
Gjerdrum, Bjørn
Gundersen, Kjell Gunnar
Lundmark, Per Olof
Aakre, Bente Monica
Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film
title Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film
title_full Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film
title_fullStr Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film
title_full_unstemmed Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film
title_short Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film
title_sort repeatability of oct-based versus scheimpflug- and reflection-based keratometry in patients with hyperosmolar and normal tear film
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7680682/
https://www.ncbi.nlm.nih.gov/pubmed/33239862
http://dx.doi.org/10.2147/OPTH.S280868
work_keys_str_mv AT gjerdrumbjørn repeatabilityofoctbasedversusscheimpflugandreflectionbasedkeratometryinpatientswithhyperosmolarandnormaltearfilm
AT gundersenkjellgunnar repeatabilityofoctbasedversusscheimpflugandreflectionbasedkeratometryinpatientswithhyperosmolarandnormaltearfilm
AT lundmarkperolof repeatabilityofoctbasedversusscheimpflugandreflectionbasedkeratometryinpatientswithhyperosmolarandnormaltearfilm
AT aakrebentemonica repeatabilityofoctbasedversusscheimpflugandreflectionbasedkeratometryinpatientswithhyperosmolarandnormaltearfilm