Initial Outcomes with Customized Myopic LASIK, Guided by Automated Ray Tracing Optimization: A Novel Technique
PURPOSE: Safety and efficacy of a novel automated ray tracing optimization in customization of excimer ablation in myopic LASIK. METHODS: In a consecutive case series, 25 patients (50 eyes) undergoing femtosecond-laser-assisted myopic LASIK were evaluated. The novel, artificial-intelligence platform...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7680798/ https://www.ncbi.nlm.nih.gov/pubmed/33239861 http://dx.doi.org/10.2147/OPTH.S280560 |
Sumario: | PURPOSE: Safety and efficacy of a novel automated ray tracing optimization in customization of excimer ablation in myopic LASIK. METHODS: In a consecutive case series, 25 patients (50 eyes) undergoing femtosecond-laser-assisted myopic LASIK were evaluated. The novel, artificial-intelligence platform initially calculates the ablation profile based on a model eye for each case, based on interferometry axial length data. Low- and high-order aberration calculation is performed by raytracing based on wavefront and Scheimpflug tomography measurements, all from a single diagnostic device. Visual acuity, refractive error, keratometry, topography, high-order aberrations and contrast sensitivity were evaluated, over six months follow-up. RESULTS: Change from pre- to 6 months post-operative: mean refractive error improved from −5.06 ± 2.54 diopters (D) (range −8.0 to −0.50 D) to −0.11 ± 0.09 D (range −0.25 to + 0.25); refractive astigmatism from −1.07 ± 0.91 D (range −4.25 to 0 D) to −0.15 ± 0.04 D (range −0.25 to 0); and topographic astigmatism from −1.65 ± 0.85 D to −0.26 ± 0.11 D (range −0.60 to 0). About 65% of eyes gained one line of vision and 38% 2 lines. Pre- to post-operative high-order aberration average: RMSh changed from 0.25 um to 0.35 um. Contrast sensitivity improved post-operatively. CONCLUSION: We report safe and effective preliminary outcomes with a novel excimer laser customization by ray tracing optimization, for myopic LASIK treatments, employing several independent up-till-now diagnostics and a customized eye model reference for each case. It bears the potential advantage through total eye aberration data and ray tracing refraction calculation to offer improved and more predictable visual outcomes. |
---|