Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data

PURPOSE: Treatment planning system (TPS) dose calculations have previously been shown to be sensitive to modeling errors, especially when treating with complex strategies like intensity‐modulated radiation therapy (IMRT). This work investigates the dosimetric impact of several dosimetric and nondosi...

Descripción completa

Detalles Bibliográficos
Autores principales: Glenn, Mallory C., Peterson, Christine B., Howell, Rebecca M., Followill, David S., Pollard‐Larkin, Julianne M., Kry, Stephen F.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7689833/
https://www.ncbi.nlm.nih.gov/pubmed/32677052
http://dx.doi.org/10.1002/mp.14396
_version_ 1783613941042118656
author Glenn, Mallory C.
Peterson, Christine B.
Howell, Rebecca M.
Followill, David S.
Pollard‐Larkin, Julianne M.
Kry, Stephen F.
author_facet Glenn, Mallory C.
Peterson, Christine B.
Howell, Rebecca M.
Followill, David S.
Pollard‐Larkin, Julianne M.
Kry, Stephen F.
author_sort Glenn, Mallory C.
collection PubMed
description PURPOSE: Treatment planning system (TPS) dose calculations have previously been shown to be sensitive to modeling errors, especially when treating with complex strategies like intensity‐modulated radiation therapy (IMRT). This work investigates the dosimetric impact of several dosimetric and nondosimetric beam modeling parameters, based on their distribution in the radiotherapy community, in two commercial TPSs in order to understand the realistic potential for dose deviations and their clinical effects. METHODS AND MATERIALS: Beam models representing standard 120‐leaf Varian Clinac‐type machines were developed in Eclipse 13.5 (AAA algorithm) and RayStation 9A (v8.99, collapsed‐cone algorithm) based upon median values of dosimetric measurements from Imaging and Radiation Oncology Core (IROC) Houston site visit data and community beam modeling parameter survey data in order to represent a baseline linear accelerator. Five clinically acceptable treatment plans (three IMRT, two VMAT) were developed for the IROC head and neck phantom. Dose distributions for each plan were recalculated after individually modifying parameters of interest (e.g., MLC transmission, percent depth doses [PDDs], and output factors) according to the 2.5th to 97.5(th) percentiles of community survey and machine performance data to encompass the realistic extent of variance in the radiotherapy community. The resultant dose distributions were evaluated by examining relative changes in average dose for thermoluminescent dosimeter (TLD) locations across the two target volumes and organ at risk (OAR). Interplay was also examined for parameters generating changes in target dose greater than 1%. RESULTS: For Eclipse, dose calculations were sensitive to changes in the dosimetric leaf gap (DLG), which resulted in differences from −5% to +3% to the targets relative to the baseline beam model. Modifying the MLC transmission factor introduced differences up to ± 1%. For RayStation, parameters determining MLC behaviors likewise contributed substantially; the MLC offset introduced changes in dose from −4% to +7%, and the MLC transmission caused changes of −4% to +2%. Among the dosimetric qualities examined, changes in PDD implementation resulted in the most substantial changes, but these were only up to ±1%. Other dosimetric factors had <1% impact on dose accuracy. Interplay between impactful parameters was found to be minimal. CONCLUSION: Factors related to the modeling of the MLC, particularly relating to the leaf offset, can cause clinically significant changes in the calculated dose for IMRT and VMAT plans. This should be of concern to the radiotherapy community because the clinical effects of poor TPS commissioning were based on reported data from clinically implemented beam models. These results further reinforce that dose errors caused by poor TPS calculations are often involved in IROC phantom failures.
format Online
Article
Text
id pubmed-7689833
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-76898332020-12-08 Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data Glenn, Mallory C. Peterson, Christine B. Howell, Rebecca M. Followill, David S. Pollard‐Larkin, Julianne M. Kry, Stephen F. Med Phys COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY PURPOSE: Treatment planning system (TPS) dose calculations have previously been shown to be sensitive to modeling errors, especially when treating with complex strategies like intensity‐modulated radiation therapy (IMRT). This work investigates the dosimetric impact of several dosimetric and nondosimetric beam modeling parameters, based on their distribution in the radiotherapy community, in two commercial TPSs in order to understand the realistic potential for dose deviations and their clinical effects. METHODS AND MATERIALS: Beam models representing standard 120‐leaf Varian Clinac‐type machines were developed in Eclipse 13.5 (AAA algorithm) and RayStation 9A (v8.99, collapsed‐cone algorithm) based upon median values of dosimetric measurements from Imaging and Radiation Oncology Core (IROC) Houston site visit data and community beam modeling parameter survey data in order to represent a baseline linear accelerator. Five clinically acceptable treatment plans (three IMRT, two VMAT) were developed for the IROC head and neck phantom. Dose distributions for each plan were recalculated after individually modifying parameters of interest (e.g., MLC transmission, percent depth doses [PDDs], and output factors) according to the 2.5th to 97.5(th) percentiles of community survey and machine performance data to encompass the realistic extent of variance in the radiotherapy community. The resultant dose distributions were evaluated by examining relative changes in average dose for thermoluminescent dosimeter (TLD) locations across the two target volumes and organ at risk (OAR). Interplay was also examined for parameters generating changes in target dose greater than 1%. RESULTS: For Eclipse, dose calculations were sensitive to changes in the dosimetric leaf gap (DLG), which resulted in differences from −5% to +3% to the targets relative to the baseline beam model. Modifying the MLC transmission factor introduced differences up to ± 1%. For RayStation, parameters determining MLC behaviors likewise contributed substantially; the MLC offset introduced changes in dose from −4% to +7%, and the MLC transmission caused changes of −4% to +2%. Among the dosimetric qualities examined, changes in PDD implementation resulted in the most substantial changes, but these were only up to ±1%. Other dosimetric factors had <1% impact on dose accuracy. Interplay between impactful parameters was found to be minimal. CONCLUSION: Factors related to the modeling of the MLC, particularly relating to the leaf offset, can cause clinically significant changes in the calculated dose for IMRT and VMAT plans. This should be of concern to the radiotherapy community because the clinical effects of poor TPS commissioning were based on reported data from clinically implemented beam models. These results further reinforce that dose errors caused by poor TPS calculations are often involved in IROC phantom failures. John Wiley and Sons Inc. 2020-08-16 2020-10 /pmc/articles/PMC7689833/ /pubmed/32677052 http://dx.doi.org/10.1002/mp.14396 Text en © 2020 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.
spellingShingle COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY
Glenn, Mallory C.
Peterson, Christine B.
Howell, Rebecca M.
Followill, David S.
Pollard‐Larkin, Julianne M.
Kry, Stephen F.
Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data
title Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data
title_full Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data
title_fullStr Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data
title_full_unstemmed Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data
title_short Sensitivity of IROC phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data
title_sort sensitivity of iroc phantom performance to radiotherapy treatment planning system beam modeling parameters based on community‐driven data
topic COMPUTATIONAL AND EXPERIMENTAL DOSIMETRY
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7689833/
https://www.ncbi.nlm.nih.gov/pubmed/32677052
http://dx.doi.org/10.1002/mp.14396
work_keys_str_mv AT glennmalloryc sensitivityofirocphantomperformancetoradiotherapytreatmentplanningsystembeammodelingparametersbasedoncommunitydrivendata
AT petersonchristineb sensitivityofirocphantomperformancetoradiotherapytreatmentplanningsystembeammodelingparametersbasedoncommunitydrivendata
AT howellrebeccam sensitivityofirocphantomperformancetoradiotherapytreatmentplanningsystembeammodelingparametersbasedoncommunitydrivendata
AT followilldavids sensitivityofirocphantomperformancetoradiotherapytreatmentplanningsystembeammodelingparametersbasedoncommunitydrivendata
AT pollardlarkinjuliannem sensitivityofirocphantomperformancetoradiotherapytreatmentplanningsystembeammodelingparametersbasedoncommunitydrivendata
AT krystephenf sensitivityofirocphantomperformancetoradiotherapytreatmentplanningsystembeammodelingparametersbasedoncommunitydrivendata