An Improved Passive CR-39-Based Direct (222)Rn/(220)Rn Progeny Detector

An improved passive CR-39-based direct (222)Rn/(220)Rn progeny detector with 3 detection channels was designed and tested in this study to measure and calculate equilibrium equivalent concentration (EEC) of both (222)Rn and (220)Rn without the equilibrium factor. A theoretical model was established...

Descripción completa

Detalles Bibliográficos
Autores principales: Hu, Jun, Yang, Guosheng, Kranrod, Chutima, Iwaoka, Kazuki, Hosoda, Masahiro, Tokonami, Shinji
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7699185/
https://www.ncbi.nlm.nih.gov/pubmed/33218216
http://dx.doi.org/10.3390/ijerph17228569
Descripción
Sumario:An improved passive CR-39-based direct (222)Rn/(220)Rn progeny detector with 3 detection channels was designed and tested in this study to measure and calculate equilibrium equivalent concentration (EEC) of both (222)Rn and (220)Rn without the equilibrium factor. A theoretical model was established to calculate the EEC with optimization. Subsequently, an exposure experiment was carried out to test the performance of this detector, and we compared the chamber experiment and the theoretical model by estimating and measuring various parameters. The deposition flux of progeny derived from the prediction agreed well with the value measured in the exposure chamber. The energy-weighted net track density (NTD) measured by this detector is much more reliable to reflect the linear relation between NTD and time-integrated EEC. Since the detector is sensitive to the exposure environmental condition, it is recommended to apply the detector to measure the EEC after its calibration in a typical indoor environment.