Cargando…
Alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization
INTRODUCTION: In Africa and other Low Resource Settings (LRS), the guideline-based and thus in most cases mesh-based treatment of inguinal hernias is only feasible to a very limited extent. This has led to an increased use of low cost meshes (LCMs, mostly mosquito meshes) for patients in LRS. Most o...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Paris
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7701087/ https://www.ncbi.nlm.nih.gov/pubmed/32975699 http://dx.doi.org/10.1007/s10029-020-02272-w |
_version_ | 1783616419385049088 |
---|---|
author | Wiessner, Reiko Lorenz, R. Gehring, A. Kleber, T. Benz, C. Sander, M. Richter, D.-U. Philipp, M. |
author_facet | Wiessner, Reiko Lorenz, R. Gehring, A. Kleber, T. Benz, C. Sander, M. Richter, D.-U. Philipp, M. |
author_sort | Wiessner, Reiko |
collection | PubMed |
description | INTRODUCTION: In Africa and other Low Resource Settings (LRS), the guideline-based and thus in most cases mesh-based treatment of inguinal hernias is only feasible to a very limited extent. This has led to an increased use of low cost meshes (LCMs, mostly mosquito meshes) for patients in LRS. Most of the LCMs used are made of polyethylene or polyester, which must be sterilized before use. The aim of our investigations was to determine changes in the biocompatibility of fibroblasts as well as mechanical and chemical properties of LCMs after steam sterilization. MATERIAL AND METHODS: Two large-pored LCMs made of polyester and polyethylene in a size of 11 x 6 cm were cut and steam sterilized at 100, 121 and 134 °C. These probes and non-sterile meshes were then subjected to mechanical tensile tests in vertical and horizontal tension, chemical analyses and biocompatibility tests with human fibroblasts. All meshes were examined by stereomicroscopy, scanning electron microscopy (SEM), LDH (cytotoxicity) measurement, viability testing, pH, lactate and glycolysis determination. RESULTS: Even macroscopically, polyethylene LCMs showed massive shrinkage after steam sterilization, especially at 121 and 134 °C. While polyester meshes showed no significant changes after sterilization with regard to deformation and damage as well as tensile force and stiffness, only the unsterile polyethylene mesh and the mesh sterilized at 100 °C could be tested mechanically due to the shrinkage of the other specimen. For these meshes the tensile forces were about four times higher than for polyester LCMs. Chemical analysis showed that the typical melting point of polyester LCMs was between 254 and 269 °C. Contrary to the specifications, the polyethylene LCM did not consist of low-density polyethylene, but rather high-density polyethylene and therefore had a melting point of 137 °C, so that the marked shrinkage described above occurred. Stereomicroscopy confirmed the shrinkage of polyethylene LCMs already after sterilization at 100 °C in contrast to polyester LCMs. Surprisingly, cytotoxicity (LDH measurement) was lowest for both non-sterile LCMs, while polyethylene LCMs sterilized at 100 and 121 °C in particular showed a significant increase in cytotoxicity 48 hours after incubation with fibroblasts. Glucose metabolism showed no significant changes between sterile and non-sterile polyethylene and polyester LCMs. CONCLUSION: The process of steam sterilization significantly alters mechanical and structural properties of synthetic hernia mesh implants. Our findings do not support a use of low-cost meshes because of their unpredictable properties after steam sterilization. |
format | Online Article Text |
id | pubmed-7701087 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Springer Paris |
record_format | MEDLINE/PubMed |
spelling | pubmed-77010872020-12-03 Alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization Wiessner, Reiko Lorenz, R. Gehring, A. Kleber, T. Benz, C. Sander, M. Richter, D.-U. Philipp, M. Hernia Original Article INTRODUCTION: In Africa and other Low Resource Settings (LRS), the guideline-based and thus in most cases mesh-based treatment of inguinal hernias is only feasible to a very limited extent. This has led to an increased use of low cost meshes (LCMs, mostly mosquito meshes) for patients in LRS. Most of the LCMs used are made of polyethylene or polyester, which must be sterilized before use. The aim of our investigations was to determine changes in the biocompatibility of fibroblasts as well as mechanical and chemical properties of LCMs after steam sterilization. MATERIAL AND METHODS: Two large-pored LCMs made of polyester and polyethylene in a size of 11 x 6 cm were cut and steam sterilized at 100, 121 and 134 °C. These probes and non-sterile meshes were then subjected to mechanical tensile tests in vertical and horizontal tension, chemical analyses and biocompatibility tests with human fibroblasts. All meshes were examined by stereomicroscopy, scanning electron microscopy (SEM), LDH (cytotoxicity) measurement, viability testing, pH, lactate and glycolysis determination. RESULTS: Even macroscopically, polyethylene LCMs showed massive shrinkage after steam sterilization, especially at 121 and 134 °C. While polyester meshes showed no significant changes after sterilization with regard to deformation and damage as well as tensile force and stiffness, only the unsterile polyethylene mesh and the mesh sterilized at 100 °C could be tested mechanically due to the shrinkage of the other specimen. For these meshes the tensile forces were about four times higher than for polyester LCMs. Chemical analysis showed that the typical melting point of polyester LCMs was between 254 and 269 °C. Contrary to the specifications, the polyethylene LCM did not consist of low-density polyethylene, but rather high-density polyethylene and therefore had a melting point of 137 °C, so that the marked shrinkage described above occurred. Stereomicroscopy confirmed the shrinkage of polyethylene LCMs already after sterilization at 100 °C in contrast to polyester LCMs. Surprisingly, cytotoxicity (LDH measurement) was lowest for both non-sterile LCMs, while polyethylene LCMs sterilized at 100 and 121 °C in particular showed a significant increase in cytotoxicity 48 hours after incubation with fibroblasts. Glucose metabolism showed no significant changes between sterile and non-sterile polyethylene and polyester LCMs. CONCLUSION: The process of steam sterilization significantly alters mechanical and structural properties of synthetic hernia mesh implants. Our findings do not support a use of low-cost meshes because of their unpredictable properties after steam sterilization. Springer Paris 2020-09-25 2020 /pmc/articles/PMC7701087/ /pubmed/32975699 http://dx.doi.org/10.1007/s10029-020-02272-w Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. |
spellingShingle | Original Article Wiessner, Reiko Lorenz, R. Gehring, A. Kleber, T. Benz, C. Sander, M. Richter, D.-U. Philipp, M. Alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization |
title | Alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization |
title_full | Alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization |
title_fullStr | Alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization |
title_full_unstemmed | Alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization |
title_short | Alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization |
title_sort | alterations in the mechanical, chemical and biocompatibility properties of low-cost polyethylene and polyester meshes after steam sterilization |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7701087/ https://www.ncbi.nlm.nih.gov/pubmed/32975699 http://dx.doi.org/10.1007/s10029-020-02272-w |
work_keys_str_mv | AT wiessnerreiko alterationsinthemechanicalchemicalandbiocompatibilitypropertiesoflowcostpolyethyleneandpolyestermeshesaftersteamsterilization AT lorenzr alterationsinthemechanicalchemicalandbiocompatibilitypropertiesoflowcostpolyethyleneandpolyestermeshesaftersteamsterilization AT gehringa alterationsinthemechanicalchemicalandbiocompatibilitypropertiesoflowcostpolyethyleneandpolyestermeshesaftersteamsterilization AT klebert alterationsinthemechanicalchemicalandbiocompatibilitypropertiesoflowcostpolyethyleneandpolyestermeshesaftersteamsterilization AT benzc alterationsinthemechanicalchemicalandbiocompatibilitypropertiesoflowcostpolyethyleneandpolyestermeshesaftersteamsterilization AT sanderm alterationsinthemechanicalchemicalandbiocompatibilitypropertiesoflowcostpolyethyleneandpolyestermeshesaftersteamsterilization AT richterdu alterationsinthemechanicalchemicalandbiocompatibilitypropertiesoflowcostpolyethyleneandpolyestermeshesaftersteamsterilization AT philippm alterationsinthemechanicalchemicalandbiocompatibilitypropertiesoflowcostpolyethyleneandpolyestermeshesaftersteamsterilization |