Cargando…

Feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with COVID-19: an observational study

BACKGROUND: Coronavirus disease 2019 (COVID-19) causes an atypical acute respiratory distress syndrome associated with thromboembolism and high shunt fraction. Shunt may be intrapulmonary, or extrapulmonary. Handheld devices are increasingly being used for point-of-care ultrasound, but their use to...

Descripción completa

Detalles Bibliográficos
Autores principales: Rajendram, Rajkumar, Hussain, Arif, Mahmood, Naveed, Kharal, Mubashar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7702202/
https://www.ncbi.nlm.nih.gov/pubmed/33252722
http://dx.doi.org/10.1186/s13089-020-00197-0
_version_ 1783616567637966848
author Rajendram, Rajkumar
Hussain, Arif
Mahmood, Naveed
Kharal, Mubashar
author_facet Rajendram, Rajkumar
Hussain, Arif
Mahmood, Naveed
Kharal, Mubashar
author_sort Rajendram, Rajkumar
collection PubMed
description BACKGROUND: Coronavirus disease 2019 (COVID-19) causes an atypical acute respiratory distress syndrome associated with thromboembolism and high shunt fraction. Shunt may be intrapulmonary, or extrapulmonary. Handheld devices are increasingly being used for point-of-care ultrasound, but their use to characterize shunt has not been reported. OBJECTIVES: Determine the feasibility of using handheld ultrasound to detect and characterize anatomical substrates of hypoxia and deep vein thrombosis (DVT) in patients with COVID-19 suspected to have severe shunt. METHODS: A handheld ultrasound device (iQ, Butterfly, USA) was used to perform lung ultrasound, vascular assessment for DVT, and limited transthoracic echocardiography (TTE) with color Doppler and saline microbubble contrast in patients with COVID-19 suspected to have severe shunt. Images were reassessed by an independent reviewer. RESULTS: After screening 40 patients, six patients who fulfilled the inclusion criteria were identified. Two were excluded because palliation had been initiated. So, four patients were studied. Interpretable images were obtained in all cases. Interobserver agreement was good. All patients had abnormal lung ultrasound (lung ultrasound score range 17–22). Identified lung pathology included interstitial syndrome with light beams and small peripheral consolidation (4), lobar consolidation (1), and pleural effusion (1). Abnormal echocardiographic findings included interatrial shunt (2), intrapulmonary shunt (1), and dilated right ventricle with tricuspid valve regurgitation (1). Significant DVT was not detected. CONCLUSION: Use of handheld ultrasound to perform combined lung ultrasound, DVT ultrasound, and limited TTE with color Doppler and saline microbubble contrast is feasible, and may be able to characterize shunt in critically hypoxic patients. Serial studies could be used to monitor changes in shunt. Further studies are required to determine whether this can guide treatment to improve the outcomes of patients with refractory hypoxia.
format Online
Article
Text
id pubmed-7702202
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-77022022020-12-01 Feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with COVID-19: an observational study Rajendram, Rajkumar Hussain, Arif Mahmood, Naveed Kharal, Mubashar Ultrasound J Original Article BACKGROUND: Coronavirus disease 2019 (COVID-19) causes an atypical acute respiratory distress syndrome associated with thromboembolism and high shunt fraction. Shunt may be intrapulmonary, or extrapulmonary. Handheld devices are increasingly being used for point-of-care ultrasound, but their use to characterize shunt has not been reported. OBJECTIVES: Determine the feasibility of using handheld ultrasound to detect and characterize anatomical substrates of hypoxia and deep vein thrombosis (DVT) in patients with COVID-19 suspected to have severe shunt. METHODS: A handheld ultrasound device (iQ, Butterfly, USA) was used to perform lung ultrasound, vascular assessment for DVT, and limited transthoracic echocardiography (TTE) with color Doppler and saline microbubble contrast in patients with COVID-19 suspected to have severe shunt. Images were reassessed by an independent reviewer. RESULTS: After screening 40 patients, six patients who fulfilled the inclusion criteria were identified. Two were excluded because palliation had been initiated. So, four patients were studied. Interpretable images were obtained in all cases. Interobserver agreement was good. All patients had abnormal lung ultrasound (lung ultrasound score range 17–22). Identified lung pathology included interstitial syndrome with light beams and small peripheral consolidation (4), lobar consolidation (1), and pleural effusion (1). Abnormal echocardiographic findings included interatrial shunt (2), intrapulmonary shunt (1), and dilated right ventricle with tricuspid valve regurgitation (1). Significant DVT was not detected. CONCLUSION: Use of handheld ultrasound to perform combined lung ultrasound, DVT ultrasound, and limited TTE with color Doppler and saline microbubble contrast is feasible, and may be able to characterize shunt in critically hypoxic patients. Serial studies could be used to monitor changes in shunt. Further studies are required to determine whether this can guide treatment to improve the outcomes of patients with refractory hypoxia. Springer International Publishing 2020-11-30 /pmc/articles/PMC7702202/ /pubmed/33252722 http://dx.doi.org/10.1186/s13089-020-00197-0 Text en © The Author(s) 2020 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Original Article
Rajendram, Rajkumar
Hussain, Arif
Mahmood, Naveed
Kharal, Mubashar
Feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with COVID-19: an observational study
title Feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with COVID-19: an observational study
title_full Feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with COVID-19: an observational study
title_fullStr Feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with COVID-19: an observational study
title_full_unstemmed Feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with COVID-19: an observational study
title_short Feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with COVID-19: an observational study
title_sort feasibility of using a handheld ultrasound device to detect and characterize shunt and deep vein thrombosis in patients with covid-19: an observational study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7702202/
https://www.ncbi.nlm.nih.gov/pubmed/33252722
http://dx.doi.org/10.1186/s13089-020-00197-0
work_keys_str_mv AT rajendramrajkumar feasibilityofusingahandheldultrasounddevicetodetectandcharacterizeshuntanddeepveinthrombosisinpatientswithcovid19anobservationalstudy
AT hussainarif feasibilityofusingahandheldultrasounddevicetodetectandcharacterizeshuntanddeepveinthrombosisinpatientswithcovid19anobservationalstudy
AT mahmoodnaveed feasibilityofusingahandheldultrasounddevicetodetectandcharacterizeshuntanddeepveinthrombosisinpatientswithcovid19anobservationalstudy
AT kharalmubashar feasibilityofusingahandheldultrasounddevicetodetectandcharacterizeshuntanddeepveinthrombosisinpatientswithcovid19anobservationalstudy