Higher incidence of aseptic loosening caused by a lower canal filling ratio with a modified modular stem in total hip arthroplasty
PURPOSE: Although a cementless modular prosthesis has shown reliable results, cases of unstable fixation and revision due to aseptic loosening were observed in our institute. The purpose of this study was to clarify the causes of unstable fixation of the prosthesis. METHODS: A total of 144 patients...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7706283/ https://www.ncbi.nlm.nih.gov/pubmed/33256767 http://dx.doi.org/10.1186/s13018-020-02101-x |
Sumario: | PURPOSE: Although a cementless modular prosthesis has shown reliable results, cases of unstable fixation and revision due to aseptic loosening were observed in our institute. The purpose of this study was to clarify the causes of unstable fixation of the prosthesis. METHODS: A total of 144 patients (154 hips) who underwent total hip arthroplasty using the modular prosthesis were retrospectively investigated. For the cohort study, 97 patients (104 hips) were included. The femoral component survival rate and sleeve fixation were assessed at a minimum follow-up of 5 years. Patients were divided into 2 groups, including stable and unstable fixation groups, by sleeve fixation. Clinical and radiographic outcomes were compared. RESULTS: The Kaplan-Meier survival rate at 9 years was 93% with revision for any reason as the endpoint in study cohort. The reasons for revision were recurrent dislocation (1 hip) and aseptic loosening of the stem (5 hips). A total of 88 hips (84.6%) showed stable fixation, and 16 hips (15.4%) showed unstable fixation at final follow-up. There was no significant difference in clinical outcomes between the 2 groups at final follow-up. The canal flare index was significantly higher, and the canal filling ratio was significantly lower in the unstable fixation group. CONCLUSION: Although the modified modular prosthesis was useful for treating anatomically difficult patients, we need to pay attention to both proximal/distal mismatch of the intramedullary canal and the canal filling ratio to achieve stable fixation and good long-term results. |
---|