A High-Precision Method for In Vitro Proton Irradiation

PURPOSE: Although proton therapy has become a well-established radiation modality, continued efforts are needed to improve our understanding of the molecular and cellular mechanisms occurring during treatment. Such studies are challenging, requiring many resources. The purpose of this study was to c...

Descripción completa

Detalles Bibliográficos
Autores principales: Howard, Michelle E., Denbeigh, Janet M., Debrot, Emily K., Remmes, Nicholas B., Herman, Michael G., Beltran, Chris J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Particle Therapy Co-operative Group 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7707323/
https://www.ncbi.nlm.nih.gov/pubmed/33274258
http://dx.doi.org/10.14338/IJPT-20-00007.1
_version_ 1783617321320841216
author Howard, Michelle E.
Denbeigh, Janet M.
Debrot, Emily K.
Remmes, Nicholas B.
Herman, Michael G.
Beltran, Chris J.
author_facet Howard, Michelle E.
Denbeigh, Janet M.
Debrot, Emily K.
Remmes, Nicholas B.
Herman, Michael G.
Beltran, Chris J.
author_sort Howard, Michelle E.
collection PubMed
description PURPOSE: Although proton therapy has become a well-established radiation modality, continued efforts are needed to improve our understanding of the molecular and cellular mechanisms occurring during treatment. Such studies are challenging, requiring many resources. The purpose of this study was to create a phantom that would allow multiple in vitro experiments to be irradiated simultaneously with a spot-scanning proton beam. MATERIALS AND METHODS: The setup included a modified patient-couch top coupled with a high-precision robotic arm for positioning. An acrylic phantom was created to hold 4 6-well cell-culture plates at 2 different positions along the Bragg curve in a reproducible manner. The proton treatment plan consisted of 1 large field encompassing all 4 plates with a monoenergetic 76.8-MeV posterior beam. For robust delivery, a mini pyramid filter was used to broaden the Bragg peak (BP) in the depth direction. Both a Markus ionization chamber and EBT3 radiochromic film measurements were used to verify absolute dose. RESULTS: A treatment plan for the simultaneous irradiation of 2 plates irradiated with high linear energy transfer protons (BP, 7 keV/μm) and 2 plates irradiated with low linear energy transfer protons (entrance, 2.2 keV/μm) was created. Dose uncertainty was larger across the setup for cell plates positioned at the BP because of beam divergence and, subsequently, variable proton-path lengths. Markus chamber measurements resulted in uncertainty values of ±1.8% from the mean dose. Negligible differences were seen in the entrance region (<0.3%). CONCLUSION: The proposed proton irradiation setup allows 4 plates to be simultaneously irradiated with 2 different portions (entrance and BP) of a 76.8-MeV beam. Dosimetric uncertainties across the setup are within ±1.8% of the mean dose.
format Online
Article
Text
id pubmed-7707323
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher The Particle Therapy Co-operative Group
record_format MEDLINE/PubMed
spelling pubmed-77073232020-12-02 A High-Precision Method for In Vitro Proton Irradiation Howard, Michelle E. Denbeigh, Janet M. Debrot, Emily K. Remmes, Nicholas B. Herman, Michael G. Beltran, Chris J. Int J Part Ther Technical Note PURPOSE: Although proton therapy has become a well-established radiation modality, continued efforts are needed to improve our understanding of the molecular and cellular mechanisms occurring during treatment. Such studies are challenging, requiring many resources. The purpose of this study was to create a phantom that would allow multiple in vitro experiments to be irradiated simultaneously with a spot-scanning proton beam. MATERIALS AND METHODS: The setup included a modified patient-couch top coupled with a high-precision robotic arm for positioning. An acrylic phantom was created to hold 4 6-well cell-culture plates at 2 different positions along the Bragg curve in a reproducible manner. The proton treatment plan consisted of 1 large field encompassing all 4 plates with a monoenergetic 76.8-MeV posterior beam. For robust delivery, a mini pyramid filter was used to broaden the Bragg peak (BP) in the depth direction. Both a Markus ionization chamber and EBT3 radiochromic film measurements were used to verify absolute dose. RESULTS: A treatment plan for the simultaneous irradiation of 2 plates irradiated with high linear energy transfer protons (BP, 7 keV/μm) and 2 plates irradiated with low linear energy transfer protons (entrance, 2.2 keV/μm) was created. Dose uncertainty was larger across the setup for cell plates positioned at the BP because of beam divergence and, subsequently, variable proton-path lengths. Markus chamber measurements resulted in uncertainty values of ±1.8% from the mean dose. Negligible differences were seen in the entrance region (<0.3%). CONCLUSION: The proposed proton irradiation setup allows 4 plates to be simultaneously irradiated with 2 different portions (entrance and BP) of a 76.8-MeV beam. Dosimetric uncertainties across the setup are within ±1.8% of the mean dose. The Particle Therapy Co-operative Group 2020-10-01 /pmc/articles/PMC7707323/ /pubmed/33274258 http://dx.doi.org/10.14338/IJPT-20-00007.1 Text en ©Copyright 2020 The Author(s) Distributed under Creative Commons CC-BY (https://creativecommons.org/licenses/cc-by/4.0/)
spellingShingle Technical Note
Howard, Michelle E.
Denbeigh, Janet M.
Debrot, Emily K.
Remmes, Nicholas B.
Herman, Michael G.
Beltran, Chris J.
A High-Precision Method for In Vitro Proton Irradiation
title A High-Precision Method for In Vitro Proton Irradiation
title_full A High-Precision Method for In Vitro Proton Irradiation
title_fullStr A High-Precision Method for In Vitro Proton Irradiation
title_full_unstemmed A High-Precision Method for In Vitro Proton Irradiation
title_short A High-Precision Method for In Vitro Proton Irradiation
title_sort high-precision method for in vitro proton irradiation
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7707323/
https://www.ncbi.nlm.nih.gov/pubmed/33274258
http://dx.doi.org/10.14338/IJPT-20-00007.1
work_keys_str_mv AT howardmichellee ahighprecisionmethodforinvitroprotonirradiation
AT denbeighjanetm ahighprecisionmethodforinvitroprotonirradiation
AT debrotemilyk ahighprecisionmethodforinvitroprotonirradiation
AT remmesnicholasb ahighprecisionmethodforinvitroprotonirradiation
AT hermanmichaelg ahighprecisionmethodforinvitroprotonirradiation
AT beltranchrisj ahighprecisionmethodforinvitroprotonirradiation
AT howardmichellee highprecisionmethodforinvitroprotonirradiation
AT denbeighjanetm highprecisionmethodforinvitroprotonirradiation
AT debrotemilyk highprecisionmethodforinvitroprotonirradiation
AT remmesnicholasb highprecisionmethodforinvitroprotonirradiation
AT hermanmichaelg highprecisionmethodforinvitroprotonirradiation
AT beltranchrisj highprecisionmethodforinvitroprotonirradiation