Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment

Purpose A single-isocenter volumetric modulated arc therapy (VMAT) treatment to multiple brain metastatic patients is an efficient stereotactic radiosurgery (SRS) option. However, the current clinical practice of single-isocenter SRS does not account for patient setup uncertainty, which degrades tre...

Descripción completa

Detalles Bibliográficos
Autores principales: Palmiero, Allison N, Critchfield, Lana, St Clair, William, Randall, Marcus, Pokhrel, Damodar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7707911/
https://www.ncbi.nlm.nih.gov/pubmed/33274143
http://dx.doi.org/10.7759/cureus.11267
_version_ 1783617453884964864
author Palmiero, Allison N
Critchfield, Lana
St Clair, William
Randall, Marcus
Pokhrel, Damodar
author_facet Palmiero, Allison N
Critchfield, Lana
St Clair, William
Randall, Marcus
Pokhrel, Damodar
author_sort Palmiero, Allison N
collection PubMed
description Purpose A single-isocenter volumetric modulated arc therapy (VMAT) treatment to multiple brain metastatic patients is an efficient stereotactic radiosurgery (SRS) option. However, the current clinical practice of single-isocenter SRS does not account for patient setup uncertainty, which degrades treatment delivery accuracy. This study quantifies the loss of target coverage and potential collateral dose to normal tissue due to clinically observable isocenter misalignment. Methods and materials Nine patients with 61 total tumors (2-16 tumors/patient) who underwent Gamma Knife® SRS were replanned in Eclipse™ using 10 megavoltages (MV) flattening-filter-free (FFF) bream (2400 MU/min), using a single-isocenter VMAT plan, similar to HyperArc™ VMAT plan. Isocenter was placed in the geometric center of the tumors. The prescription was 20 Gy to each tumor. Average gross tumor volume (GTV) and planning target volume (PTV) were 1.1 cc (0.02-11.5 cc) and 1.9 cc (0.11-18.8 cc), respectively, derived from MRI images. The average isocenter to tumor distance was 5.5 cm (1.6-10.1 cm). Six-degrees of freedom (6DoF) random and systematic residual set up errors within [±2 mm, ±2(o)] were generated using an in-house script in Eclipse based on our pre-treatment daily cone-beam CT imaging shifts and recomputed for the simulated VMAT plan. Relative loss of target coverage as a function of tumor size and distance to isocenter were evaluated as well as collateral dose to organs-at risk (OAR). Results The average beam-on time was less than six minutes. However, loss of target coverage for clinically observable setup errors were, on average, 7.9% (up to 73.1%) for the GTV (p < 0.001) and 21.5% for the PTV (up to 93.7%; p < 0.001). The correlation was found for both random and systematic residual setup errors with tumor sizes; there was a greater loss of target coverage for small tumors. Due to isocenter misalignment, OAR doses fluctuated and potentially receive higher doses than the original plan. Conclusion A single-isocenter VMAT SRS treatment (similar to HyperArc™ VMAT) to multiple brain metastases was fast with < 6 min of beam-on time. However, due to small residual set up errors, single-isocenter VMAT, in its current use, is not an accurate SRS treatment modality for multiple brain metastases. Loss of target coverage was statistically significant, especially for smaller lesions, and may not be clinically acceptable if left uncorrected. Further investigation of correction strategies is underway.
format Online
Article
Text
id pubmed-7707911
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-77079112020-12-02 Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment Palmiero, Allison N Critchfield, Lana St Clair, William Randall, Marcus Pokhrel, Damodar Cureus Medical Physics Purpose A single-isocenter volumetric modulated arc therapy (VMAT) treatment to multiple brain metastatic patients is an efficient stereotactic radiosurgery (SRS) option. However, the current clinical practice of single-isocenter SRS does not account for patient setup uncertainty, which degrades treatment delivery accuracy. This study quantifies the loss of target coverage and potential collateral dose to normal tissue due to clinically observable isocenter misalignment. Methods and materials Nine patients with 61 total tumors (2-16 tumors/patient) who underwent Gamma Knife® SRS were replanned in Eclipse™ using 10 megavoltages (MV) flattening-filter-free (FFF) bream (2400 MU/min), using a single-isocenter VMAT plan, similar to HyperArc™ VMAT plan. Isocenter was placed in the geometric center of the tumors. The prescription was 20 Gy to each tumor. Average gross tumor volume (GTV) and planning target volume (PTV) were 1.1 cc (0.02-11.5 cc) and 1.9 cc (0.11-18.8 cc), respectively, derived from MRI images. The average isocenter to tumor distance was 5.5 cm (1.6-10.1 cm). Six-degrees of freedom (6DoF) random and systematic residual set up errors within [±2 mm, ±2(o)] were generated using an in-house script in Eclipse based on our pre-treatment daily cone-beam CT imaging shifts and recomputed for the simulated VMAT plan. Relative loss of target coverage as a function of tumor size and distance to isocenter were evaluated as well as collateral dose to organs-at risk (OAR). Results The average beam-on time was less than six minutes. However, loss of target coverage for clinically observable setup errors were, on average, 7.9% (up to 73.1%) for the GTV (p < 0.001) and 21.5% for the PTV (up to 93.7%; p < 0.001). The correlation was found for both random and systematic residual setup errors with tumor sizes; there was a greater loss of target coverage for small tumors. Due to isocenter misalignment, OAR doses fluctuated and potentially receive higher doses than the original plan. Conclusion A single-isocenter VMAT SRS treatment (similar to HyperArc™ VMAT) to multiple brain metastases was fast with < 6 min of beam-on time. However, due to small residual set up errors, single-isocenter VMAT, in its current use, is not an accurate SRS treatment modality for multiple brain metastases. Loss of target coverage was statistically significant, especially for smaller lesions, and may not be clinically acceptable if left uncorrected. Further investigation of correction strategies is underway. Cureus 2020-10-30 /pmc/articles/PMC7707911/ /pubmed/33274143 http://dx.doi.org/10.7759/cureus.11267 Text en Copyright © 2020, Palmiero et al. http://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Medical Physics
Palmiero, Allison N
Critchfield, Lana
St Clair, William
Randall, Marcus
Pokhrel, Damodar
Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment
title Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment
title_full Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment
title_fullStr Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment
title_full_unstemmed Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment
title_short Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment
title_sort single-isocenter volumetric modulated arc therapy (vmat) radiosurgery for multiple brain metastases: potential loss of target(s) coverage due to isocenter misalignment
topic Medical Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7707911/
https://www.ncbi.nlm.nih.gov/pubmed/33274143
http://dx.doi.org/10.7759/cureus.11267
work_keys_str_mv AT palmieroallisonn singleisocentervolumetricmodulatedarctherapyvmatradiosurgeryformultiplebrainmetastasespotentiallossoftargetscoverageduetoisocentermisalignment
AT critchfieldlana singleisocentervolumetricmodulatedarctherapyvmatradiosurgeryformultiplebrainmetastasespotentiallossoftargetscoverageduetoisocentermisalignment
AT stclairwilliam singleisocentervolumetricmodulatedarctherapyvmatradiosurgeryformultiplebrainmetastasespotentiallossoftargetscoverageduetoisocentermisalignment
AT randallmarcus singleisocentervolumetricmodulatedarctherapyvmatradiosurgeryformultiplebrainmetastasespotentiallossoftargetscoverageduetoisocentermisalignment
AT pokhreldamodar singleisocentervolumetricmodulatedarctherapyvmatradiosurgeryformultiplebrainmetastasespotentiallossoftargetscoverageduetoisocentermisalignment