Deployment of a Covered Stent for Left Internal Mammary Perforation From Percutaneous Pacemaker Implantation
A 79-year-old woman developed a hemothorax 2 days after implantation of a permanent pacemaker. Computed tomography angiography revealed active extravasation from the left internal mammary artery. A covered stent was deployed to manage the arterial perforation. This case report explores different ven...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7710994/ https://www.ncbi.nlm.nih.gov/pubmed/33305234 http://dx.doi.org/10.1016/j.cjco.2020.06.018 |
Sumario: | A 79-year-old woman developed a hemothorax 2 days after implantation of a permanent pacemaker. Computed tomography angiography revealed active extravasation from the left internal mammary artery. A covered stent was deployed to manage the arterial perforation. This case report explores different venous access techniques to minimize the risk of arterial injuries and describes the use of a covered stent in managing a non-grafted left internal mammary artery injury from a pacemaker implantation procedure. |
---|