Cargando…

The Effect of Corneal Thickness, Densitometry and Curvature on Intraocular Pressure Measurements Obtained by Applanation, Rebound and Dynamic Contour Tonometry

Evaluate the effect of corneal thickness, densitometry and curvature on intraocular pressure (IOP) measurements obtained by Goldmann applanation tonometry (GAT), non-contact tonometry (NCT), rebound tonometry (RT), and dynamic contour tonometry (DCT). A cross-sectional prospective study involving 40...

Descripción completa

Detalles Bibliográficos
Autores principales: Olyntho Junior, Marco Antonio de Castro, Augusto, Lucas Bertazzi, Gracitelli, Carolina P. B., Tatham, Andrew J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7711663/
https://www.ncbi.nlm.nih.gov/pubmed/33096873
http://dx.doi.org/10.3390/vision4040045
Descripción
Sumario:Evaluate the effect of corneal thickness, densitometry and curvature on intraocular pressure (IOP) measurements obtained by Goldmann applanation tonometry (GAT), non-contact tonometry (NCT), rebound tonometry (RT), and dynamic contour tonometry (DCT). A cross-sectional prospective study involving 40 participants was performed. Corneal measurements were obtained using Pentacam (Oculus GMbH, Wetzlar, Germany), densitometry was measured at annuli of 0–2, 2–6, 6–10 and 10–12 mm. The relationship between corneal thickness (central, 4 and 6 mm), corneal astigmatism and corneal densitometry and IOP was examined. There was a significant relationship between corneal thickness (central, 4 and 6 mm) and GAT180, GAT90, RT, and NCT (P < 0.001 for all comparisons) but not for DCT. Higher corneal densitometry (6–10 mm and 10–12 mm zones) was associated with higher IOP from GAT180 and GAT90, and higher densitometry in the 6–10 mm zone correlated with higher IOP from NCT, however corneal densitometry increased with age. Accounting for age, the relationship between corneal densitometry and IOP measurements was not significant. In eyes with greater corneal astigmatism there was a greater difference between GAT90 and GAT180 measurements. IOP measurements may be affected by corneal thickness, densitometry and curvature. DCT was less affected by properties of the cornea compared to other devices.