Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes

BACKGROUND: The performance of the pulse oximeter was evaluated based on the ISO 80601-2-61:2011 (E) guidelines. This study aimed to determine whether the various finger probes of the MP570T pulse oximeter (MEK-ICS Co., Ltd., Korea) would provide clinically reliable peripheral oxygen saturation (SpO...

Descripción completa

Detalles Bibliográficos
Autores principales: Choi, Byung-Moon, Kang, Bong Jin, Yun, Ho-Yong, Jeon, Bokyoung, Bang, Ji-Yeon, Noh, Gyu-Jeong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society of Anesthesiologists 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7713831/
https://www.ncbi.nlm.nih.gov/pubmed/33329838
http://dx.doi.org/10.17085/apm.20028
_version_ 1783618625597341696
author Choi, Byung-Moon
Kang, Bong Jin
Yun, Ho-Yong
Jeon, Bokyoung
Bang, Ji-Yeon
Noh, Gyu-Jeong
author_facet Choi, Byung-Moon
Kang, Bong Jin
Yun, Ho-Yong
Jeon, Bokyoung
Bang, Ji-Yeon
Noh, Gyu-Jeong
author_sort Choi, Byung-Moon
collection PubMed
description BACKGROUND: The performance of the pulse oximeter was evaluated based on the ISO 80601-2-61:2011 (E) guidelines. This study aimed to determine whether the various finger probes of the MP570T pulse oximeter (MEK-ICS Co., Ltd., Korea) would provide clinically reliable peripheral oxygen saturation (SpO(2)) readings over a range of 70-100% arterial oxygen saturation (SaO(2)) during non-motion conditions. METHODS: Each volunteer (n = 12) was connected to a breathing circuit for the administration of a hypoxic gas mixture. For frequent blood sampling, an arterial cannula was placed in a radial artery. The following seven pulse oximeter probes were simultaneously attached to each volunteer’s fingers: (1) WA-100 reusable finger probe (MEDNIS Co., Ltd., Korea), (2) MDNA disposable finger probe (MEDNIS Co., Ltd.), (3) IS-1011 disposable finger probe (Insung Medical Co., Ltd., Korea), (4) CJ340NA disposable finger probe (CHUN JI IN Medical Co., Ltd., Korea), (5) Nellcor(TM) OxiMax DS-100A reusable finger probe (Medtronic, USA), (6) Nellcor(TM) OxiMax MAX-N disposable finger probe (Medtronic), and (7) OXI-PRO DA disposable finger probe (Bio-Protech Inc., Korea). RESULTS: A total of 275 SpO(2)-SaO(2) pairs were included in the analysis. The accuracy of the root mean square (A(rms)) of each probe was 2.83%, 3.98%, 3.75%, 6.84%, 3.43%, 5.17%, and 3.84%, respectively. CONCLUSIONS: The MP570T pulse oximeter with WA-100 reusable, MDNA disposable, IS-1011 disposable, Nellcor(TM) OxiMax DS-100A reusable, and OXI-PRO DA disposable finger probes meets an acceptable standard of SpO(2) accuracy under non-motion conditions.
format Online
Article
Text
id pubmed-7713831
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Korean Society of Anesthesiologists
record_format MEDLINE/PubMed
spelling pubmed-77138312020-12-15 Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes Choi, Byung-Moon Kang, Bong Jin Yun, Ho-Yong Jeon, Bokyoung Bang, Ji-Yeon Noh, Gyu-Jeong Anesth Pain Med (Seoul) General Article BACKGROUND: The performance of the pulse oximeter was evaluated based on the ISO 80601-2-61:2011 (E) guidelines. This study aimed to determine whether the various finger probes of the MP570T pulse oximeter (MEK-ICS Co., Ltd., Korea) would provide clinically reliable peripheral oxygen saturation (SpO(2)) readings over a range of 70-100% arterial oxygen saturation (SaO(2)) during non-motion conditions. METHODS: Each volunteer (n = 12) was connected to a breathing circuit for the administration of a hypoxic gas mixture. For frequent blood sampling, an arterial cannula was placed in a radial artery. The following seven pulse oximeter probes were simultaneously attached to each volunteer’s fingers: (1) WA-100 reusable finger probe (MEDNIS Co., Ltd., Korea), (2) MDNA disposable finger probe (MEDNIS Co., Ltd.), (3) IS-1011 disposable finger probe (Insung Medical Co., Ltd., Korea), (4) CJ340NA disposable finger probe (CHUN JI IN Medical Co., Ltd., Korea), (5) Nellcor(TM) OxiMax DS-100A reusable finger probe (Medtronic, USA), (6) Nellcor(TM) OxiMax MAX-N disposable finger probe (Medtronic), and (7) OXI-PRO DA disposable finger probe (Bio-Protech Inc., Korea). RESULTS: A total of 275 SpO(2)-SaO(2) pairs were included in the analysis. The accuracy of the root mean square (A(rms)) of each probe was 2.83%, 3.98%, 3.75%, 6.84%, 3.43%, 5.17%, and 3.84%, respectively. CONCLUSIONS: The MP570T pulse oximeter with WA-100 reusable, MDNA disposable, IS-1011 disposable, Nellcor(TM) OxiMax DS-100A reusable, and OXI-PRO DA disposable finger probes meets an acceptable standard of SpO(2) accuracy under non-motion conditions. Korean Society of Anesthesiologists 2020-07-31 2020-07-31 /pmc/articles/PMC7713831/ /pubmed/33329838 http://dx.doi.org/10.17085/apm.20028 Text en Copyright © the Korean Society of Anesthesiologists, 2020 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle General Article
Choi, Byung-Moon
Kang, Bong Jin
Yun, Ho-Yong
Jeon, Bokyoung
Bang, Ji-Yeon
Noh, Gyu-Jeong
Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes
title Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes
title_full Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes
title_fullStr Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes
title_full_unstemmed Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes
title_short Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes
title_sort performance of the mp570t pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes
topic General Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7713831/
https://www.ncbi.nlm.nih.gov/pubmed/33329838
http://dx.doi.org/10.17085/apm.20028
work_keys_str_mv AT choibyungmoon performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes
AT kangbongjin performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes
AT yunhoyong performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes
AT jeonbokyoung performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes
AT bangjiyeon performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes
AT nohgyujeong performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes