Cargando…
Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes
BACKGROUND: The performance of the pulse oximeter was evaluated based on the ISO 80601-2-61:2011 (E) guidelines. This study aimed to determine whether the various finger probes of the MP570T pulse oximeter (MEK-ICS Co., Ltd., Korea) would provide clinically reliable peripheral oxygen saturation (SpO...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Society of Anesthesiologists
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7713831/ https://www.ncbi.nlm.nih.gov/pubmed/33329838 http://dx.doi.org/10.17085/apm.20028 |
_version_ | 1783618625597341696 |
---|---|
author | Choi, Byung-Moon Kang, Bong Jin Yun, Ho-Yong Jeon, Bokyoung Bang, Ji-Yeon Noh, Gyu-Jeong |
author_facet | Choi, Byung-Moon Kang, Bong Jin Yun, Ho-Yong Jeon, Bokyoung Bang, Ji-Yeon Noh, Gyu-Jeong |
author_sort | Choi, Byung-Moon |
collection | PubMed |
description | BACKGROUND: The performance of the pulse oximeter was evaluated based on the ISO 80601-2-61:2011 (E) guidelines. This study aimed to determine whether the various finger probes of the MP570T pulse oximeter (MEK-ICS Co., Ltd., Korea) would provide clinically reliable peripheral oxygen saturation (SpO(2)) readings over a range of 70-100% arterial oxygen saturation (SaO(2)) during non-motion conditions. METHODS: Each volunteer (n = 12) was connected to a breathing circuit for the administration of a hypoxic gas mixture. For frequent blood sampling, an arterial cannula was placed in a radial artery. The following seven pulse oximeter probes were simultaneously attached to each volunteer’s fingers: (1) WA-100 reusable finger probe (MEDNIS Co., Ltd., Korea), (2) MDNA disposable finger probe (MEDNIS Co., Ltd.), (3) IS-1011 disposable finger probe (Insung Medical Co., Ltd., Korea), (4) CJ340NA disposable finger probe (CHUN JI IN Medical Co., Ltd., Korea), (5) Nellcor(TM) OxiMax DS-100A reusable finger probe (Medtronic, USA), (6) Nellcor(TM) OxiMax MAX-N disposable finger probe (Medtronic), and (7) OXI-PRO DA disposable finger probe (Bio-Protech Inc., Korea). RESULTS: A total of 275 SpO(2)-SaO(2) pairs were included in the analysis. The accuracy of the root mean square (A(rms)) of each probe was 2.83%, 3.98%, 3.75%, 6.84%, 3.43%, 5.17%, and 3.84%, respectively. CONCLUSIONS: The MP570T pulse oximeter with WA-100 reusable, MDNA disposable, IS-1011 disposable, Nellcor(TM) OxiMax DS-100A reusable, and OXI-PRO DA disposable finger probes meets an acceptable standard of SpO(2) accuracy under non-motion conditions. |
format | Online Article Text |
id | pubmed-7713831 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Korean Society of Anesthesiologists |
record_format | MEDLINE/PubMed |
spelling | pubmed-77138312020-12-15 Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes Choi, Byung-Moon Kang, Bong Jin Yun, Ho-Yong Jeon, Bokyoung Bang, Ji-Yeon Noh, Gyu-Jeong Anesth Pain Med (Seoul) General Article BACKGROUND: The performance of the pulse oximeter was evaluated based on the ISO 80601-2-61:2011 (E) guidelines. This study aimed to determine whether the various finger probes of the MP570T pulse oximeter (MEK-ICS Co., Ltd., Korea) would provide clinically reliable peripheral oxygen saturation (SpO(2)) readings over a range of 70-100% arterial oxygen saturation (SaO(2)) during non-motion conditions. METHODS: Each volunteer (n = 12) was connected to a breathing circuit for the administration of a hypoxic gas mixture. For frequent blood sampling, an arterial cannula was placed in a radial artery. The following seven pulse oximeter probes were simultaneously attached to each volunteer’s fingers: (1) WA-100 reusable finger probe (MEDNIS Co., Ltd., Korea), (2) MDNA disposable finger probe (MEDNIS Co., Ltd.), (3) IS-1011 disposable finger probe (Insung Medical Co., Ltd., Korea), (4) CJ340NA disposable finger probe (CHUN JI IN Medical Co., Ltd., Korea), (5) Nellcor(TM) OxiMax DS-100A reusable finger probe (Medtronic, USA), (6) Nellcor(TM) OxiMax MAX-N disposable finger probe (Medtronic), and (7) OXI-PRO DA disposable finger probe (Bio-Protech Inc., Korea). RESULTS: A total of 275 SpO(2)-SaO(2) pairs were included in the analysis. The accuracy of the root mean square (A(rms)) of each probe was 2.83%, 3.98%, 3.75%, 6.84%, 3.43%, 5.17%, and 3.84%, respectively. CONCLUSIONS: The MP570T pulse oximeter with WA-100 reusable, MDNA disposable, IS-1011 disposable, Nellcor(TM) OxiMax DS-100A reusable, and OXI-PRO DA disposable finger probes meets an acceptable standard of SpO(2) accuracy under non-motion conditions. Korean Society of Anesthesiologists 2020-07-31 2020-07-31 /pmc/articles/PMC7713831/ /pubmed/33329838 http://dx.doi.org/10.17085/apm.20028 Text en Copyright © the Korean Society of Anesthesiologists, 2020 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | General Article Choi, Byung-Moon Kang, Bong Jin Yun, Ho-Yong Jeon, Bokyoung Bang, Ji-Yeon Noh, Gyu-Jeong Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes |
title | Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes |
title_full | Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes |
title_fullStr | Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes |
title_full_unstemmed | Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes |
title_short | Performance of the MP570T pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes |
title_sort | performance of the mp570t pulse oximeter in volunteers participating in the controlled desaturation study: a comparison of seven probes |
topic | General Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7713831/ https://www.ncbi.nlm.nih.gov/pubmed/33329838 http://dx.doi.org/10.17085/apm.20028 |
work_keys_str_mv | AT choibyungmoon performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes AT kangbongjin performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes AT yunhoyong performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes AT jeonbokyoung performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes AT bangjiyeon performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes AT nohgyujeong performanceofthemp570tpulseoximeterinvolunteersparticipatinginthecontrolleddesaturationstudyacomparisonofsevenprobes |