Cargando…
An intraoperative 3D image-based navigation error during cervical pedicle screw insertion
A 67-year-old man underwent posterior cervical decompression surgery for ossification of the posterior longitudinal ligament (OPLL) with fixation using cervical pedicle screws (CPSs) guided by intraoperative 3D image-based navigation. Intraoperatively, while creating the screw hole using the navigat...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nagoya University
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7719458/ https://www.ncbi.nlm.nih.gov/pubmed/33311810 http://dx.doi.org/10.18999/nagjms.82.4.799 |
_version_ | 1783619684008984576 |
---|---|
author | Nakashima, Hiroaki Ishikawa, Yoshimoto Ando, Kei Kobayashi, Kazuyoshi Ishiguro, Naoki Imagama, Shiro |
author_facet | Nakashima, Hiroaki Ishikawa, Yoshimoto Ando, Kei Kobayashi, Kazuyoshi Ishiguro, Naoki Imagama, Shiro |
author_sort | Nakashima, Hiroaki |
collection | PubMed |
description | A 67-year-old man underwent posterior cervical decompression surgery for ossification of the posterior longitudinal ligament (OPLL) with fixation using cervical pedicle screws (CPSs) guided by intraoperative 3D image-based navigation. Intraoperatively, while creating the screw hole using the navigation probe, the virtual trajectory on the intraoperative navigation screen showed a 10-degree angle discrepancy in the axial plane depending on whether a probing force was or was not applied for making the hole. This was potentially caused by vertebra rotation and a bent probe. Consequently, the CPSs were placed more laterally than the ideal trajectory, which resulted in <2 mm lateral perforation to the foramen transversarium. There were no screw insertion-related perioperative complications. Based on this case, we conclude that navigation error during CPS insertion can occur even with intraoperative 3D image-based navigation. The risk of a bowed navigation probe caused by posterior cervical muscle and vertebra rotation should be considered, even with use of a navigation reference frame. |
format | Online Article Text |
id | pubmed-7719458 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Nagoya University |
record_format | MEDLINE/PubMed |
spelling | pubmed-77194582020-12-11 An intraoperative 3D image-based navigation error during cervical pedicle screw insertion Nakashima, Hiroaki Ishikawa, Yoshimoto Ando, Kei Kobayashi, Kazuyoshi Ishiguro, Naoki Imagama, Shiro Nagoya J Med Sci Case Report A 67-year-old man underwent posterior cervical decompression surgery for ossification of the posterior longitudinal ligament (OPLL) with fixation using cervical pedicle screws (CPSs) guided by intraoperative 3D image-based navigation. Intraoperatively, while creating the screw hole using the navigation probe, the virtual trajectory on the intraoperative navigation screen showed a 10-degree angle discrepancy in the axial plane depending on whether a probing force was or was not applied for making the hole. This was potentially caused by vertebra rotation and a bent probe. Consequently, the CPSs were placed more laterally than the ideal trajectory, which resulted in <2 mm lateral perforation to the foramen transversarium. There were no screw insertion-related perioperative complications. Based on this case, we conclude that navigation error during CPS insertion can occur even with intraoperative 3D image-based navigation. The risk of a bowed navigation probe caused by posterior cervical muscle and vertebra rotation should be considered, even with use of a navigation reference frame. Nagoya University 2020-11 /pmc/articles/PMC7719458/ /pubmed/33311810 http://dx.doi.org/10.18999/nagjms.82.4.799 Text en http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an Open Access article distributed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. To view the details of this license, please visit (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Case Report Nakashima, Hiroaki Ishikawa, Yoshimoto Ando, Kei Kobayashi, Kazuyoshi Ishiguro, Naoki Imagama, Shiro An intraoperative 3D image-based navigation error during cervical pedicle screw insertion |
title | An intraoperative 3D image-based navigation error during cervical pedicle screw insertion |
title_full | An intraoperative 3D image-based navigation error during cervical pedicle screw insertion |
title_fullStr | An intraoperative 3D image-based navigation error during cervical pedicle screw insertion |
title_full_unstemmed | An intraoperative 3D image-based navigation error during cervical pedicle screw insertion |
title_short | An intraoperative 3D image-based navigation error during cervical pedicle screw insertion |
title_sort | intraoperative 3d image-based navigation error during cervical pedicle screw insertion |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7719458/ https://www.ncbi.nlm.nih.gov/pubmed/33311810 http://dx.doi.org/10.18999/nagjms.82.4.799 |
work_keys_str_mv | AT nakashimahiroaki anintraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT ishikawayoshimoto anintraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT andokei anintraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT kobayashikazuyoshi anintraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT ishiguronaoki anintraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT imagamashiro anintraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT nakashimahiroaki intraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT ishikawayoshimoto intraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT andokei intraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT kobayashikazuyoshi intraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT ishiguronaoki intraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion AT imagamashiro intraoperative3dimagebasednavigationerrorduringcervicalpediclescrewinsertion |