Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems
CARE Dose 4D modulates mAs through several mechanisms according to patient size and shape, whilst maintaining user-defined reference image quality on Siemens Symbia single-photon emission computed tomography (SPECT)-computed tomography (CT) systems. A 20 kg child reference was used in child protocol...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7725139/ https://www.ncbi.nlm.nih.gov/pubmed/33079892 http://dx.doi.org/10.1097/MNM.0000000000001312 |
_version_ | 1783620651115872256 |
---|---|
author | Hansen, Sofie L. Bebbington, Natalie A. |
author_facet | Hansen, Sofie L. Bebbington, Natalie A. |
author_sort | Hansen, Sofie L. |
collection | PubMed |
description | CARE Dose 4D modulates mAs through several mechanisms according to patient size and shape, whilst maintaining user-defined reference image quality on Siemens Symbia single-photon emission computed tomography (SPECT)-computed tomography (CT) systems. A 20 kg child reference was used in child protocols prior to software version VB10 and a 75 kg adult thereafter. Quality reference mAs conversion factors are estimated for delivering equivalent mAs to children between two comparable SPECT-CT systems using adult and child references for topogram-based patient-size-related dose level adaptations. METHODS: A child phantom was scanned using child protocols on a Siemens Symbia T16 (child reference) and a Siemens Symbia Intevo Bold (adult reference). On each system, scans of the thorax, abdomen and pelvis were acquired with arms up and down, at 80 and 110 kVp. Quality reference mAs settings of 10–50 were used on the Symbia T16 and 40–200 on the Symbia Intevo Bold. These data were used to propose quality reference mAs (adult/child reference) conversion factors according to scan range, arm position and tube voltage. RESULTS: Quality reference mAs for child protocols using the adult reference should multiply the child quality reference mAs by the following factors, to give comparable delivered mAs: arms up 80 kV: 3.8 (thorax), 3.8 (abdomen), 4.3 (pelvis); arms up at 110 kV: 3.8 (thorax), 4.1 (abdomen), 4.6 (pelvis); arms down at 80 kV: 4.0 (thorax), 3.7 (abdomen), 3.9 (pelvis); arms down at 110 kV: 4.3 (thorax), 4.0 (abdomen), 4.2 (pelvis). CONCLUSION: Conversion factors for child to adult dose modulation references are proposed, allowing comparable delivered mAs to a child. Video abstract: http://links.lww.com/NMC/A178 |
format | Online Article Text |
id | pubmed-7725139 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Lippincott Williams & Wilkins |
record_format | MEDLINE/PubMed |
spelling | pubmed-77251392020-12-15 Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems Hansen, Sofie L. Bebbington, Natalie A. Nucl Med Commun Technical Note CARE Dose 4D modulates mAs through several mechanisms according to patient size and shape, whilst maintaining user-defined reference image quality on Siemens Symbia single-photon emission computed tomography (SPECT)-computed tomography (CT) systems. A 20 kg child reference was used in child protocols prior to software version VB10 and a 75 kg adult thereafter. Quality reference mAs conversion factors are estimated for delivering equivalent mAs to children between two comparable SPECT-CT systems using adult and child references for topogram-based patient-size-related dose level adaptations. METHODS: A child phantom was scanned using child protocols on a Siemens Symbia T16 (child reference) and a Siemens Symbia Intevo Bold (adult reference). On each system, scans of the thorax, abdomen and pelvis were acquired with arms up and down, at 80 and 110 kVp. Quality reference mAs settings of 10–50 were used on the Symbia T16 and 40–200 on the Symbia Intevo Bold. These data were used to propose quality reference mAs (adult/child reference) conversion factors according to scan range, arm position and tube voltage. RESULTS: Quality reference mAs for child protocols using the adult reference should multiply the child quality reference mAs by the following factors, to give comparable delivered mAs: arms up 80 kV: 3.8 (thorax), 3.8 (abdomen), 4.3 (pelvis); arms up at 110 kV: 3.8 (thorax), 4.1 (abdomen), 4.6 (pelvis); arms down at 80 kV: 4.0 (thorax), 3.7 (abdomen), 3.9 (pelvis); arms down at 110 kV: 4.3 (thorax), 4.0 (abdomen), 4.2 (pelvis). CONCLUSION: Conversion factors for child to adult dose modulation references are proposed, allowing comparable delivered mAs to a child. Video abstract: http://links.lww.com/NMC/A178 Lippincott Williams & Wilkins 2020-10-16 2021-01 /pmc/articles/PMC7725139/ /pubmed/33079892 http://dx.doi.org/10.1097/MNM.0000000000001312 Text en Copyright © 2020 The Author(s). Published by Wolters Kluwer Health, Inc. This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (https://creativecommons.org/licenses/by-nc-nd/4.0/) (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal. |
spellingShingle | Technical Note Hansen, Sofie L. Bebbington, Natalie A. Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems |
title | Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems |
title_full | Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems |
title_fullStr | Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems |
title_full_unstemmed | Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems |
title_short | Estimation of CARE Dose 4D quality reference mAs conversion factors for child to adult reference patient in child protocols on Siemens Symbia SPECT-CT systems |
title_sort | estimation of care dose 4d quality reference mas conversion factors for child to adult reference patient in child protocols on siemens symbia spect-ct systems |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7725139/ https://www.ncbi.nlm.nih.gov/pubmed/33079892 http://dx.doi.org/10.1097/MNM.0000000000001312 |
work_keys_str_mv | AT hansensofiel estimationofcaredose4dqualityreferencemasconversionfactorsforchildtoadultreferencepatientinchildprotocolsonsiemenssymbiaspectctsystems AT bebbingtonnataliea estimationofcaredose4dqualityreferencemasconversionfactorsforchildtoadultreferencepatientinchildprotocolsonsiemenssymbiaspectctsystems |