Agreement and repeatability of Icare ic100 tonometer

PURPOSE: To find the agreement and repeatability of Icare ic100 tonometer. METHODS: We included 150 subjects above the age of 18 years for this cross-sectional, multicenter study with intraocular pressure (IOP) ≥7 mmHg. After the initial ophthalmic examination, two masked examiners took five IOP mea...

Descripción completa

Detalles Bibliográficos
Autores principales: Jose, Judy, Ve, Ramesh S, Pai, H Vijaya, Biswas, Sayantan, Parimi, Vamsi, Poojary, Paresh, Nagarajan, T
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7728033/
https://www.ncbi.nlm.nih.gov/pubmed/32971622
http://dx.doi.org/10.4103/ijo.IJO_546_19
_version_ 1783621183061622784
author Jose, Judy
Ve, Ramesh S
Pai, H Vijaya
Biswas, Sayantan
Parimi, Vamsi
Poojary, Paresh
Nagarajan, T
author_facet Jose, Judy
Ve, Ramesh S
Pai, H Vijaya
Biswas, Sayantan
Parimi, Vamsi
Poojary, Paresh
Nagarajan, T
author_sort Jose, Judy
collection PubMed
description PURPOSE: To find the agreement and repeatability of Icare ic100 tonometer. METHODS: We included 150 subjects above the age of 18 years for this cross-sectional, multicenter study with intraocular pressure (IOP) ≥7 mmHg. After the initial ophthalmic examination, two masked examiners took five IOP measurements using three different instruments; Icare ic100, Icare TA01i, and Goldmann applanation tonometer (GAT) in only one eye of the participants. Comparison of agreement of IOP using different instruments was quantified with intraclass correlation coefficient (ICC) using the two-way random effects models of absolute agreement and Cronbach's alpha. The test-retest variability of the instruments was assessed by deriving repeatability coefficient (RC) and coefficient of variation (CV). RESULTS: Agreement between the tonometers across the different IOP groups had no statistically significant difference in their mean IOP. Icare ic100 was found to have good reliability across all IOP groups (ICC value >0.78) when compared with Icare TA01i. In comparison with GAT, Icare ic100 showed good reliability across all IOP groups (ICC >0.87) except >16 to <23 mmHg group where it showed moderate reliability (ICC = 0.52). Icare ic100 showed good repeatability with RC and CV of 2.67 and 4.89, respectively. CONCLUSION: Icare ic100 rebound tonometer can measure IOP with relatively small measurement error and can provide a reliable and repeatable reading in comparison with GAT across a wide pressure range without hampering corneal health.
format Online
Article
Text
id pubmed-7728033
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-77280332020-12-11 Agreement and repeatability of Icare ic100 tonometer Jose, Judy Ve, Ramesh S Pai, H Vijaya Biswas, Sayantan Parimi, Vamsi Poojary, Paresh Nagarajan, T Indian J Ophthalmol Original Article PURPOSE: To find the agreement and repeatability of Icare ic100 tonometer. METHODS: We included 150 subjects above the age of 18 years for this cross-sectional, multicenter study with intraocular pressure (IOP) ≥7 mmHg. After the initial ophthalmic examination, two masked examiners took five IOP measurements using three different instruments; Icare ic100, Icare TA01i, and Goldmann applanation tonometer (GAT) in only one eye of the participants. Comparison of agreement of IOP using different instruments was quantified with intraclass correlation coefficient (ICC) using the two-way random effects models of absolute agreement and Cronbach's alpha. The test-retest variability of the instruments was assessed by deriving repeatability coefficient (RC) and coefficient of variation (CV). RESULTS: Agreement between the tonometers across the different IOP groups had no statistically significant difference in their mean IOP. Icare ic100 was found to have good reliability across all IOP groups (ICC value >0.78) when compared with Icare TA01i. In comparison with GAT, Icare ic100 showed good reliability across all IOP groups (ICC >0.87) except >16 to <23 mmHg group where it showed moderate reliability (ICC = 0.52). Icare ic100 showed good repeatability with RC and CV of 2.67 and 4.89, respectively. CONCLUSION: Icare ic100 rebound tonometer can measure IOP with relatively small measurement error and can provide a reliable and repeatable reading in comparison with GAT across a wide pressure range without hampering corneal health. Wolters Kluwer - Medknow 2020-10 2020-09-23 /pmc/articles/PMC7728033/ /pubmed/32971622 http://dx.doi.org/10.4103/ijo.IJO_546_19 Text en Copyright: © 2020 Indian Journal of Ophthalmology http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Jose, Judy
Ve, Ramesh S
Pai, H Vijaya
Biswas, Sayantan
Parimi, Vamsi
Poojary, Paresh
Nagarajan, T
Agreement and repeatability of Icare ic100 tonometer
title Agreement and repeatability of Icare ic100 tonometer
title_full Agreement and repeatability of Icare ic100 tonometer
title_fullStr Agreement and repeatability of Icare ic100 tonometer
title_full_unstemmed Agreement and repeatability of Icare ic100 tonometer
title_short Agreement and repeatability of Icare ic100 tonometer
title_sort agreement and repeatability of icare ic100 tonometer
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7728033/
https://www.ncbi.nlm.nih.gov/pubmed/32971622
http://dx.doi.org/10.4103/ijo.IJO_546_19
work_keys_str_mv AT josejudy agreementandrepeatabilityoficareic100tonometer
AT verameshs agreementandrepeatabilityoficareic100tonometer
AT paihvijaya agreementandrepeatabilityoficareic100tonometer
AT biswassayantan agreementandrepeatabilityoficareic100tonometer
AT parimivamsi agreementandrepeatabilityoficareic100tonometer
AT poojaryparesh agreementandrepeatabilityoficareic100tonometer
AT nagarajant agreementandrepeatabilityoficareic100tonometer