Cargando…

Quantification and correction of the scattered X-rays from a megavoltage photon beam to a linac-mounted kilovoltage imaging subsystem

OBJECTIVE: To quantify and correct megavoltage (MV) scattered X-rays (MV-scatter) on an image acquired using a linac-mounted kilovoltage (kV) imaging subsystem. METHODS AND MATERIALS: A linac-mounted flat-panel detector (FPD) was used to acquire an image containing MV-scatter by activating the FPD o...

Descripción completa

Detalles Bibliográficos
Autores principales: Iramina, Hiraku, Nakamura, Mitsuhiro, Miyabe, Yuki, Mukumoto, Nobutaka, Ono, Tomohiro, Hirashima, Hideaki, Mizowaki, Takashi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Institute of Radiology. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7731796/
https://www.ncbi.nlm.nih.gov/pubmed/33324865
http://dx.doi.org/10.1259/bjro.20190048
Descripción
Sumario:OBJECTIVE: To quantify and correct megavoltage (MV) scattered X-rays (MV-scatter) on an image acquired using a linac-mounted kilovoltage (kV) imaging subsystem. METHODS AND MATERIALS: A linac-mounted flat-panel detector (FPD) was used to acquire an image containing MV-scatter by activating the FPD only during MV beam irradiation. 6-, 10-, and 15 MV with a flattening-filter (FF; 6X-FF, 10X-FF, 15X-FF), and 6- and 10 MV without an FF (6X-FFF, 10X-FFF) were used. The maps were acquired by changing one of the irradiation parameters while the others remained fixed. The mean pixel values of the MV-scatter were normalized to the 6X-FF reference condition (MV-scatter value). An MV-scatter database was constructed using these values. An MV-scatter correction experiment with one full arc image acquisition and two square field sizes (FSs) was conducted. Measurement- and estimation-based corrections were performed using the database. The image contrast was calculated at each angle. RESULTS: The MV-scatter increased with a larger FS and dose rate. The MV-scatter value factor varied substantially depending on the FPD position or collimator rotation. The median relative error ranges of the contrast for the image without, and with the measurement- and estimation-based correction were −10.9 to −2.9, and −1.5 to 4.8 and −7.4 to 2.6, respectively, for an FS of 10.0 × 10.0 cm(2). CONCLUSIONS: The MV-scatter was strongly dependent on the FS, dose rate, and FPD position. The MV-scatter correction improved the image contrast. ADVANCES IN KNOWLEDGE: The MV-scatters on the TrueBeam linac kV imaging subsystem were quantified with various MV beam parameters, and strongly depended on the fieldsize, dose rate, and flat panel detector position. The MV-scatter correction using the constructed database improved the image quality.