Coagulation testing: Comparison of portable (CoaguChek(®) XS) and automated coagulation analyzer in healthy cats
BACKGROUND AND AIM: The CoaguChek(®) XS (CCX) is a portable coagulation analyzer that is widely used to monitor prothrombin time (PT) in human patients taking oral anticoagulants. It can also be reliably used for screening dogs when PT is in the normal range. Efficacy of the portable CCX coagulation...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Veterinary World
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7750227/ https://www.ncbi.nlm.nih.gov/pubmed/33363352 http://dx.doi.org/10.14202/vetworld.2020.2541-2545 |
_version_ | 1783625450912743424 |
---|---|
author | Tonthong, Sireeporn Rungpupradit, Jetsada |
author_facet | Tonthong, Sireeporn Rungpupradit, Jetsada |
author_sort | Tonthong, Sireeporn |
collection | PubMed |
description | BACKGROUND AND AIM: The CoaguChek(®) XS (CCX) is a portable coagulation analyzer that is widely used to monitor prothrombin time (PT) in human patients taking oral anticoagulants. It can also be reliably used for screening dogs when PT is in the normal range. Efficacy of the portable CCX coagulation analyzer was evaluated for testing PT in healthy cats and the normal range was established. MATERIALS AND METHODS: Blood samples of 82 cats were collected from the jugular vein and PT was measured using both the CCX and an automated coagulation analyzer (ACA). Spearman’s correlation was used to measure the strength and direction of association between the two analyzers, while limits of agreement were assessed utilizing Bland-Altman analysis. RESULTS: Range of PT using the CCX was 10.1-14.1 s. Correlation between the two analyzers was moderate but significant (r=0.3465, p=0.0014). Mean difference between CCX-PT and ACA-PT was 1.624 s and standard deviation was 0.890 with 95.1% of the samples falling within the limits of agreement. CONCLUSION: The CCX is a portable, easy to use coagulation analyzer that requires a small volume of blood and gives results within 1 min. Results showed moderate correlation and good agreement with a standard automated laboratory analyzer. The CCX can be used for screening coagulation testing when PT is in the normal range for cats. However, testing accuracy of the CCX in abnormal PT cats should be further investigated before diagnostic coagulopathy applications. |
format | Online Article Text |
id | pubmed-7750227 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Veterinary World |
record_format | MEDLINE/PubMed |
spelling | pubmed-77502272020-12-23 Coagulation testing: Comparison of portable (CoaguChek(®) XS) and automated coagulation analyzer in healthy cats Tonthong, Sireeporn Rungpupradit, Jetsada Vet World Research Article BACKGROUND AND AIM: The CoaguChek(®) XS (CCX) is a portable coagulation analyzer that is widely used to monitor prothrombin time (PT) in human patients taking oral anticoagulants. It can also be reliably used for screening dogs when PT is in the normal range. Efficacy of the portable CCX coagulation analyzer was evaluated for testing PT in healthy cats and the normal range was established. MATERIALS AND METHODS: Blood samples of 82 cats were collected from the jugular vein and PT was measured using both the CCX and an automated coagulation analyzer (ACA). Spearman’s correlation was used to measure the strength and direction of association between the two analyzers, while limits of agreement were assessed utilizing Bland-Altman analysis. RESULTS: Range of PT using the CCX was 10.1-14.1 s. Correlation between the two analyzers was moderate but significant (r=0.3465, p=0.0014). Mean difference between CCX-PT and ACA-PT was 1.624 s and standard deviation was 0.890 with 95.1% of the samples falling within the limits of agreement. CONCLUSION: The CCX is a portable, easy to use coagulation analyzer that requires a small volume of blood and gives results within 1 min. Results showed moderate correlation and good agreement with a standard automated laboratory analyzer. The CCX can be used for screening coagulation testing when PT is in the normal range for cats. However, testing accuracy of the CCX in abnormal PT cats should be further investigated before diagnostic coagulopathy applications. Veterinary World 2020-11 2020-11-27 /pmc/articles/PMC7750227/ /pubmed/33363352 http://dx.doi.org/10.14202/vetworld.2020.2541-2545 Text en Copyright: © Tonthong and Rungpupradit. http://creativecommons.org/licenses/by/4.0 Open Access. This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Article Tonthong, Sireeporn Rungpupradit, Jetsada Coagulation testing: Comparison of portable (CoaguChek(®) XS) and automated coagulation analyzer in healthy cats |
title | Coagulation testing: Comparison of portable (CoaguChek(®) XS) and automated coagulation analyzer in healthy cats |
title_full | Coagulation testing: Comparison of portable (CoaguChek(®) XS) and automated coagulation analyzer in healthy cats |
title_fullStr | Coagulation testing: Comparison of portable (CoaguChek(®) XS) and automated coagulation analyzer in healthy cats |
title_full_unstemmed | Coagulation testing: Comparison of portable (CoaguChek(®) XS) and automated coagulation analyzer in healthy cats |
title_short | Coagulation testing: Comparison of portable (CoaguChek(®) XS) and automated coagulation analyzer in healthy cats |
title_sort | coagulation testing: comparison of portable (coaguchek(®) xs) and automated coagulation analyzer in healthy cats |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7750227/ https://www.ncbi.nlm.nih.gov/pubmed/33363352 http://dx.doi.org/10.14202/vetworld.2020.2541-2545 |
work_keys_str_mv | AT tonthongsireeporn coagulationtestingcomparisonofportablecoaguchekxsandautomatedcoagulationanalyzerinhealthycats AT rungpupraditjetsada coagulationtestingcomparisonofportablecoaguchekxsandautomatedcoagulationanalyzerinhealthycats |