Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield

BACKGROUND: Some organs in the body are sensitive to radiation such as eyes, breast, and gonads. Protection of sensitive organs against radiation is necessary. Recently, many sensitive organ shields have been developed from different materials. OBJECTIVE: The aim of this study is to evaluate the dos...

Descripción completa

Detalles Bibliográficos
Autores principales: F., Zahroh, C., Anam, H., Sutanto, Y., Irdawati, Z., Arifin, Y., Kartikasari
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Shiraz University of Medical Sciences 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7753262/
https://www.ncbi.nlm.nih.gov/pubmed/33364207
http://dx.doi.org/10.31661/jbpe.v0i0.1912-1007
_version_ 1783626035485474816
author F., Zahroh
C., Anam
H., Sutanto
Y., Irdawati
Z., Arifin
Y., Kartikasari
author_facet F., Zahroh
C., Anam
H., Sutanto
Y., Irdawati
Z., Arifin
Y., Kartikasari
author_sort F., Zahroh
collection PubMed
description BACKGROUND: Some organs in the body are sensitive to radiation such as eyes, breast, and gonads. Protection of sensitive organs against radiation is necessary. Recently, many sensitive organ shields have been developed from different materials. OBJECTIVE: The aim of this study is to evaluate the dose reduction and image quality from implementation of Silicone Rubber-Lead (SR-Pb) as an alternative gonad shield in digital radiography (DR). MATERIAL AND METHODS: In this experimental study, the SR-Pb gonad shields with various thicknesses of 2, 4, 6, 8, and 10 mm were synthesized. This study used the Pb percentage of 5 wt%. An anthropomorphic phantom was used in abdomen plain examinations. The results obtained from the use of the SR-Pb was compared with standard gonad shield, i.e. lead apron. To measure the dose reduction, the Piranha detector was used. The image quality assessment was evaluated with the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR). RESULTS: This study showed the dose reduction was significant for all SR-Pb thicknesses, and incrementally increased with the increase of the SR-Pb thickness. The minimum and maximum of dose reduction were 22.8% for 2 mm and 66.9% for 10 mm SR-Pb, respectively. CONCLUSION: Compared to the reference image without gonad shield, the SNR and CNR do not significantly change. Hence, the SR-Pb is probably to be used as an alternative gonad shield.
format Online
Article
Text
id pubmed-7753262
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Shiraz University of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-77532622020-12-23 Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield F., Zahroh C., Anam H., Sutanto Y., Irdawati Z., Arifin Y., Kartikasari J Biomed Phys Eng Original Article BACKGROUND: Some organs in the body are sensitive to radiation such as eyes, breast, and gonads. Protection of sensitive organs against radiation is necessary. Recently, many sensitive organ shields have been developed from different materials. OBJECTIVE: The aim of this study is to evaluate the dose reduction and image quality from implementation of Silicone Rubber-Lead (SR-Pb) as an alternative gonad shield in digital radiography (DR). MATERIAL AND METHODS: In this experimental study, the SR-Pb gonad shields with various thicknesses of 2, 4, 6, 8, and 10 mm were synthesized. This study used the Pb percentage of 5 wt%. An anthropomorphic phantom was used in abdomen plain examinations. The results obtained from the use of the SR-Pb was compared with standard gonad shield, i.e. lead apron. To measure the dose reduction, the Piranha detector was used. The image quality assessment was evaluated with the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR). RESULTS: This study showed the dose reduction was significant for all SR-Pb thicknesses, and incrementally increased with the increase of the SR-Pb thickness. The minimum and maximum of dose reduction were 22.8% for 2 mm and 66.9% for 10 mm SR-Pb, respectively. CONCLUSION: Compared to the reference image without gonad shield, the SNR and CNR do not significantly change. Hence, the SR-Pb is probably to be used as an alternative gonad shield. Shiraz University of Medical Sciences 2020-12-01 /pmc/articles/PMC7753262/ /pubmed/33364207 http://dx.doi.org/10.31661/jbpe.v0i0.1912-1007 Text en Copyright: © Journal of Biomedical Physics and Engineering http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 Unported License, ( http://creativecommons.org/licenses/by-nc/4.0/ ) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
F., Zahroh
C., Anam
H., Sutanto
Y., Irdawati
Z., Arifin
Y., Kartikasari
Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield
title Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield
title_full Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield
title_fullStr Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield
title_full_unstemmed Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield
title_short Effect of Silicone Rubber-Lead (SR-Pb) Thickness on Dose Reduction and Image Quality as Gonad Shield
title_sort effect of silicone rubber-lead (sr-pb) thickness on dose reduction and image quality as gonad shield
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7753262/
https://www.ncbi.nlm.nih.gov/pubmed/33364207
http://dx.doi.org/10.31661/jbpe.v0i0.1912-1007
work_keys_str_mv AT fzahroh effectofsiliconerubberleadsrpbthicknessondosereductionandimagequalityasgonadshield
AT canam effectofsiliconerubberleadsrpbthicknessondosereductionandimagequalityasgonadshield
AT hsutanto effectofsiliconerubberleadsrpbthicknessondosereductionandimagequalityasgonadshield
AT yirdawati effectofsiliconerubberleadsrpbthicknessondosereductionandimagequalityasgonadshield
AT zarifin effectofsiliconerubberleadsrpbthicknessondosereductionandimagequalityasgonadshield
AT ykartikasari effectofsiliconerubberleadsrpbthicknessondosereductionandimagequalityasgonadshield