Using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold
PURPOSE: In order to reduce heart dose, DIBH has become a common practice in left‐sided whole breast irradiation. This technique involves a significant strain on patients due to the breath‐hold requirements. We hereby investigate the dosimetric and delivery feasibility of using flattening filter fre...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7769397/ https://www.ncbi.nlm.nih.gov/pubmed/33270988 http://dx.doi.org/10.1002/acm2.13109 |
_version_ | 1783629316282646528 |
---|---|
author | Wisnoskie, Sarah B. Liang, Xiaoying Wahl, Andrew O. Bennion, Nathan R. Granatowicz, Andrew D. Zhou, Sumin Zheng, Dandan |
author_facet | Wisnoskie, Sarah B. Liang, Xiaoying Wahl, Andrew O. Bennion, Nathan R. Granatowicz, Andrew D. Zhou, Sumin Zheng, Dandan |
author_sort | Wisnoskie, Sarah B. |
collection | PubMed |
description | PURPOSE: In order to reduce heart dose, DIBH has become a common practice in left‐sided whole breast irradiation. This technique involves a significant strain on patients due to the breath‐hold requirements. We hereby investigate the dosimetric and delivery feasibility of using flattening filter free (FFF) energies with electronic tissue compensation (ECOMP) planning technique to reduce the required breath‐hold lengths and increase patient compatibility. METHODS: Fifteen left‐sided, postlumpectomy patients previously receiving DIBH whole‐breast radiotherapy (266cGy x 16fx) were retrospectively planned using ECOMP for both 6X and 6X‐FFF. A dosimetric comparison was made between the two plans for each patient using various dosimetric constraints. Delivery feasibility was analyzed by recalculating the 6X ECOMP plan with 6X‐FFF without replanning (6X‐FFF QA) and delivering both plans for a one‐to‐one comparison using Gamma analysis. Beam‐on times for the 6X and 6X‐FFF plans were measured. For all tests, Wilcoxon signed‐rank test was used with P < 0.05 as significant. RESULTS: No statistical difference was observed between 6X and 6X‐FFF plans for most dosimetric endpoints except contralateral breast D(max) (P = 0.0008) and skin D(max) (p = 0.03) and D(min) (P = 0.01) for which 6X‐FFF showed favorable results when compared with 6X. 6X‐FFF significantly reduced beam‐on times for all patients by 22%–42% (average 32%). All plan QAs passed departmental gamma criteria (10% low‐dose threshold, 3%/3mm, >95% passing). CONCLUSION: ECOMP planning with FFF was found feasible for left‐sided breast patients with DIBH. Plan quality is comparable, if not better, than plans using flattened beams. FFF ECOMP could significantly reduce beam‐on time and required breath‐hold lengths thereby increasing patient compatibility for this treatment while offering satisfactory plan quality and delivery accuracy. |
format | Online Article Text |
id | pubmed-7769397 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-77693972020-12-31 Using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold Wisnoskie, Sarah B. Liang, Xiaoying Wahl, Andrew O. Bennion, Nathan R. Granatowicz, Andrew D. Zhou, Sumin Zheng, Dandan J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: In order to reduce heart dose, DIBH has become a common practice in left‐sided whole breast irradiation. This technique involves a significant strain on patients due to the breath‐hold requirements. We hereby investigate the dosimetric and delivery feasibility of using flattening filter free (FFF) energies with electronic tissue compensation (ECOMP) planning technique to reduce the required breath‐hold lengths and increase patient compatibility. METHODS: Fifteen left‐sided, postlumpectomy patients previously receiving DIBH whole‐breast radiotherapy (266cGy x 16fx) were retrospectively planned using ECOMP for both 6X and 6X‐FFF. A dosimetric comparison was made between the two plans for each patient using various dosimetric constraints. Delivery feasibility was analyzed by recalculating the 6X ECOMP plan with 6X‐FFF without replanning (6X‐FFF QA) and delivering both plans for a one‐to‐one comparison using Gamma analysis. Beam‐on times for the 6X and 6X‐FFF plans were measured. For all tests, Wilcoxon signed‐rank test was used with P < 0.05 as significant. RESULTS: No statistical difference was observed between 6X and 6X‐FFF plans for most dosimetric endpoints except contralateral breast D(max) (P = 0.0008) and skin D(max) (p = 0.03) and D(min) (P = 0.01) for which 6X‐FFF showed favorable results when compared with 6X. 6X‐FFF significantly reduced beam‐on times for all patients by 22%–42% (average 32%). All plan QAs passed departmental gamma criteria (10% low‐dose threshold, 3%/3mm, >95% passing). CONCLUSION: ECOMP planning with FFF was found feasible for left‐sided breast patients with DIBH. Plan quality is comparable, if not better, than plans using flattened beams. FFF ECOMP could significantly reduce beam‐on time and required breath‐hold lengths thereby increasing patient compatibility for this treatment while offering satisfactory plan quality and delivery accuracy. John Wiley and Sons Inc. 2020-12-03 /pmc/articles/PMC7769397/ /pubmed/33270988 http://dx.doi.org/10.1002/acm2.13109 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Wisnoskie, Sarah B. Liang, Xiaoying Wahl, Andrew O. Bennion, Nathan R. Granatowicz, Andrew D. Zhou, Sumin Zheng, Dandan Using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold |
title | Using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold |
title_full | Using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold |
title_fullStr | Using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold |
title_full_unstemmed | Using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold |
title_short | Using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold |
title_sort | using flattening filter free beams in electronic tissue compensation whole breast irradiation with deep inspiration breath hold |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7769397/ https://www.ncbi.nlm.nih.gov/pubmed/33270988 http://dx.doi.org/10.1002/acm2.13109 |
work_keys_str_mv | AT wisnoskiesarahb usingflatteningfilterfreebeamsinelectronictissuecompensationwholebreastirradiationwithdeepinspirationbreathhold AT liangxiaoying usingflatteningfilterfreebeamsinelectronictissuecompensationwholebreastirradiationwithdeepinspirationbreathhold AT wahlandrewo usingflatteningfilterfreebeamsinelectronictissuecompensationwholebreastirradiationwithdeepinspirationbreathhold AT bennionnathanr usingflatteningfilterfreebeamsinelectronictissuecompensationwholebreastirradiationwithdeepinspirationbreathhold AT granatowiczandrewd usingflatteningfilterfreebeamsinelectronictissuecompensationwholebreastirradiationwithdeepinspirationbreathhold AT zhousumin usingflatteningfilterfreebeamsinelectronictissuecompensationwholebreastirradiationwithdeepinspirationbreathhold AT zhengdandan usingflatteningfilterfreebeamsinelectronictissuecompensationwholebreastirradiationwithdeepinspirationbreathhold |