Cargando…
Coil embolization of renal artery aneurysms: Simultaneous delivery of three microdevices with a novel 6-Fr guiding sheath
Here, we report the case of a 59-year-old male patient who underwent transcatheter embolization of bilateral renal artery aneurysms, using 2 microballoons and 1 microcatheter, all carried within a single guiding sheath. During coil embolization in a visceral artery, there are situations that can req...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7773766/ https://www.ncbi.nlm.nih.gov/pubmed/33408801 http://dx.doi.org/10.1016/j.radcr.2020.12.010 |