Aptamer-Based Lateral Flow Biosensor for Rapid Detection of Salivary Cortisol

[Image: see text] We have developed a disposable point-of-care (POC) aptamer-based biosensor for the detection of salivary cortisol. Nonstressful and noninvasive sampling of saliva compared to that of blood makes saliva an attractive biological matrix in developing POC devices for biomarker monitori...

Descripción completa

Detalles Bibliográficos
Autores principales: Dalirirad, Shima, Han, Daewoo, Steckl, Andrew J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2020
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7774066/
https://www.ncbi.nlm.nih.gov/pubmed/33403250
http://dx.doi.org/10.1021/acsomega.0c03223
Descripción
Sumario:[Image: see text] We have developed a disposable point-of-care (POC) aptamer-based biosensor for the detection of salivary cortisol. Nonstressful and noninvasive sampling of saliva compared to that of blood makes saliva an attractive biological matrix in developing POC devices for biomarker monitoring. Aptamers are attractive as recognition elements for multiple reasons, including their specific chemical synthesis, high stability, lack of immunogenicity, and cell-free evolution. A duplex aptamer conjugated to the surface of Au nanoparticles (AuNPs) by Au–S bonds is utilized as the sensor probe in a lateral flow assay (LFA) device. The addition of saliva samples containing cortisol makes the cortisol–aptamer undergo conformational changes and dissociate from the capture probe. Increasing cortisol concentration in the dispensed saliva sample results in increased dissociation and leads to increased binding of AuNP conjugate on the test line. Therefore, the color intensity of the test line on the LFA is a direct function of the concentration of cortisol in saliva. This simple and fast method provides detection in the cortisol range of ∼0.5–15 ng/mL, which is in the clinically accepted range for salivary cortisol. The limit of detection was 0.37 ng/mL, and the accuracy was confirmed by enzyme-linked immunosorbent assay (ELISA) testing results. High selectivity was observed for salivary cortisol against other closely related steroids and stress biomarkers present in saliva.