Cargando…

An additive manufacturing fabricated a split Pitot tube transducer for mechanical ventilator analyzers

INTRODUCTION: Frequent calibration of ventilators and anesthesia machines might reduce the risk of death and of sequelae in patients under mechanical ventilation. However, ventilator analyzers might be difficult to purchase due to high cost or even in trade-restrictive scenarios, such as the COVID-1...

Descripción completa

Detalles Bibliográficos
Autores principales: de Andrade, Luiz A. K., Manffra, Elisangela F.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7778415/
http://dx.doi.org/10.1007/s42600-020-00121-4
_version_ 1783631123245432832
author de Andrade, Luiz A. K.
Manffra, Elisangela F.
author_facet de Andrade, Luiz A. K.
Manffra, Elisangela F.
author_sort de Andrade, Luiz A. K.
collection PubMed
description INTRODUCTION: Frequent calibration of ventilators and anesthesia machines might reduce the risk of death and of sequelae in patients under mechanical ventilation. However, ventilator analyzers might be difficult to purchase due to high cost or even in trade-restrictive scenarios, such as the COVID-19 pandemic. To alleviate this problem, the aim of this article is to present the design and characterization of a Pitot tube transducer fabricated with additive manufacturing (AM), to be used in ventilator analyzers. METHOD: A split Pitot tube (SPT) transducer was designed using computer-aided design (CAD) and characterized using finite element method (FEM) simulations. Bernoulli’s equation was used to determine a transducer discharge coefficient. The sensor was fabricated with the acrylonitrile butadiene styrene (ABS) thermoplastic, with the fused deposition modeling after properly configuring the 3D printer. RESULTS: According to the simulated velocity profile, the transducer does not impose excessive resistance to the flow and the pressure profile revealed that the pressure is constant and stable inside the pressure lines. The characteristic curve of differential pressure (dP) versus flow is quadratic. For the minimal and maximal simulated flows 0.01 to 300 L/min, the corresponding dP values are 6.535·10(−4) Pa and 13.178·10(3) Pa. CONCLUSION: The split Pitot tube transducer developed here has appropriate characteristics for measuring air flows from mechanical ventilators, and the delivered pressures can be read by commercial electronic sensors. The AM is viable for fabricating the transducer, and the printing time is considerably low compared to that necessary to order a similar part from a sales representative.
format Online
Article
Text
id pubmed-7778415
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-77784152021-01-04 An additive manufacturing fabricated a split Pitot tube transducer for mechanical ventilator analyzers de Andrade, Luiz A. K. Manffra, Elisangela F. Res. Biomed. Eng. Technical Communication INTRODUCTION: Frequent calibration of ventilators and anesthesia machines might reduce the risk of death and of sequelae in patients under mechanical ventilation. However, ventilator analyzers might be difficult to purchase due to high cost or even in trade-restrictive scenarios, such as the COVID-19 pandemic. To alleviate this problem, the aim of this article is to present the design and characterization of a Pitot tube transducer fabricated with additive manufacturing (AM), to be used in ventilator analyzers. METHOD: A split Pitot tube (SPT) transducer was designed using computer-aided design (CAD) and characterized using finite element method (FEM) simulations. Bernoulli’s equation was used to determine a transducer discharge coefficient. The sensor was fabricated with the acrylonitrile butadiene styrene (ABS) thermoplastic, with the fused deposition modeling after properly configuring the 3D printer. RESULTS: According to the simulated velocity profile, the transducer does not impose excessive resistance to the flow and the pressure profile revealed that the pressure is constant and stable inside the pressure lines. The characteristic curve of differential pressure (dP) versus flow is quadratic. For the minimal and maximal simulated flows 0.01 to 300 L/min, the corresponding dP values are 6.535·10(−4) Pa and 13.178·10(3) Pa. CONCLUSION: The split Pitot tube transducer developed here has appropriate characteristics for measuring air flows from mechanical ventilators, and the delivered pressures can be read by commercial electronic sensors. The AM is viable for fabricating the transducer, and the printing time is considerably low compared to that necessary to order a similar part from a sales representative. Springer International Publishing 2021-01-02 2022 /pmc/articles/PMC7778415/ http://dx.doi.org/10.1007/s42600-020-00121-4 Text en © Sociedade Brasileira de Engenharia Biomedica 2021 This article is made available via the PMC Open Access Subset for unrestricted research re-use and secondary analysis in any form or by any means with acknowledgement of the original source. These permissions are granted for the duration of the World Health Organization (WHO) declaration of COVID-19 as a global pandemic.
spellingShingle Technical Communication
de Andrade, Luiz A. K.
Manffra, Elisangela F.
An additive manufacturing fabricated a split Pitot tube transducer for mechanical ventilator analyzers
title An additive manufacturing fabricated a split Pitot tube transducer for mechanical ventilator analyzers
title_full An additive manufacturing fabricated a split Pitot tube transducer for mechanical ventilator analyzers
title_fullStr An additive manufacturing fabricated a split Pitot tube transducer for mechanical ventilator analyzers
title_full_unstemmed An additive manufacturing fabricated a split Pitot tube transducer for mechanical ventilator analyzers
title_short An additive manufacturing fabricated a split Pitot tube transducer for mechanical ventilator analyzers
title_sort additive manufacturing fabricated a split pitot tube transducer for mechanical ventilator analyzers
topic Technical Communication
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7778415/
http://dx.doi.org/10.1007/s42600-020-00121-4
work_keys_str_mv AT deandradeluizak anadditivemanufacturingfabricatedasplitpitottubetransducerformechanicalventilatoranalyzers
AT manffraelisangelaf anadditivemanufacturingfabricatedasplitpitottubetransducerformechanicalventilatoranalyzers
AT deandradeluizak additivemanufacturingfabricatedasplitpitottubetransducerformechanicalventilatoranalyzers
AT manffraelisangelaf additivemanufacturingfabricatedasplitpitottubetransducerformechanicalventilatoranalyzers