Validation of the Omron HEM-1040 Blood Pressure Monitor According to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 Protocol

OBJECTIVE: The accuracy of the Omron HEM-1040 automated oscillometric upper-arm blood pressure (BP) monitoring device, designed for home self-measurements in adult populations, was tested according to the American National Standards Institute/Association for the Advancement of Medical Instrumentatio...

Descripción completa

Detalles Bibliográficos
Autores principales: Takahashi, Hakuo, Saito, Kanako
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7778435/
https://www.ncbi.nlm.nih.gov/pubmed/33402825
http://dx.doi.org/10.2147/VHRM.S287016
_version_ 1783631127451271168
author Takahashi, Hakuo
Saito, Kanako
author_facet Takahashi, Hakuo
Saito, Kanako
author_sort Takahashi, Hakuo
collection PubMed
description OBJECTIVE: The accuracy of the Omron HEM-1040 automated oscillometric upper-arm blood pressure (BP) monitoring device, designed for home self-measurements in adult populations, was tested according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060–2:2013 protocol. METHODS: The requirements of this protocol were precisely followed with the recruitment of 85 individuals in whom sequential systolic BP (SBP) and diastolic BP (DBP) were measured in the same (left) arm and who fulfilled the inclusion criteria involving the ranges of arm circumference, SBP, and DBP. RESULTS: The mean device-observer difference in the 255 separate pairs of BP data was −2.7 ± 7.14 mmHg for SBP and −3.3 ± 5.70 mmHg for DBP. The data were in accordance with criterion 1 of the ANSI/AAMI/ISO 81060–2:2013 standard requirements (≤ 5 ± ≤ 8 mmHg). In addition, the mean device-observer difference in the 85 participants was −2.7 ± 5.89 mmHg for SBP and −3.3 ± 4.99 mmHg for DBP, fulfilling criterion 2 with a standard deviation of ≤ 6.39 mmHg for SBP and ≤ 6.09 mmHg for DBP. CONCLUSION: The OMRON HEM-1040 BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation standard.
format Online
Article
Text
id pubmed-7778435
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Dove
record_format MEDLINE/PubMed
spelling pubmed-77784352021-01-04 Validation of the Omron HEM-1040 Blood Pressure Monitor According to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 Protocol Takahashi, Hakuo Saito, Kanako Vasc Health Risk Manag Original Research OBJECTIVE: The accuracy of the Omron HEM-1040 automated oscillometric upper-arm blood pressure (BP) monitoring device, designed for home self-measurements in adult populations, was tested according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060–2:2013 protocol. METHODS: The requirements of this protocol were precisely followed with the recruitment of 85 individuals in whom sequential systolic BP (SBP) and diastolic BP (DBP) were measured in the same (left) arm and who fulfilled the inclusion criteria involving the ranges of arm circumference, SBP, and DBP. RESULTS: The mean device-observer difference in the 255 separate pairs of BP data was −2.7 ± 7.14 mmHg for SBP and −3.3 ± 5.70 mmHg for DBP. The data were in accordance with criterion 1 of the ANSI/AAMI/ISO 81060–2:2013 standard requirements (≤ 5 ± ≤ 8 mmHg). In addition, the mean device-observer difference in the 85 participants was −2.7 ± 5.89 mmHg for SBP and −3.3 ± 4.99 mmHg for DBP, fulfilling criterion 2 with a standard deviation of ≤ 6.39 mmHg for SBP and ≤ 6.09 mmHg for DBP. CONCLUSION: The OMRON HEM-1040 BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation standard. Dove 2020-12-29 /pmc/articles/PMC7778435/ /pubmed/33402825 http://dx.doi.org/10.2147/VHRM.S287016 Text en © 2020 Takahashi and Saito. http://creativecommons.org/licenses/by-nc/3.0/ This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
spellingShingle Original Research
Takahashi, Hakuo
Saito, Kanako
Validation of the Omron HEM-1040 Blood Pressure Monitor According to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 Protocol
title Validation of the Omron HEM-1040 Blood Pressure Monitor According to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 Protocol
title_full Validation of the Omron HEM-1040 Blood Pressure Monitor According to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 Protocol
title_fullStr Validation of the Omron HEM-1040 Blood Pressure Monitor According to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 Protocol
title_full_unstemmed Validation of the Omron HEM-1040 Blood Pressure Monitor According to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 Protocol
title_short Validation of the Omron HEM-1040 Blood Pressure Monitor According to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2013 Protocol
title_sort validation of the omron hem-1040 blood pressure monitor according to the american national standards institute/association for the advancement of medical instrumentation/international organization for standardization 81060-2:2013 protocol
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7778435/
https://www.ncbi.nlm.nih.gov/pubmed/33402825
http://dx.doi.org/10.2147/VHRM.S287016
work_keys_str_mv AT takahashihakuo validationoftheomronhem1040bloodpressuremonitoraccordingtotheamericannationalstandardsinstituteassociationfortheadvancementofmedicalinstrumentationinternationalorganizationforstandardization8106022013protocol
AT saitokanako validationoftheomronhem1040bloodpressuremonitoraccordingtotheamericannationalstandardsinstituteassociationfortheadvancementofmedicalinstrumentationinternationalorganizationforstandardization8106022013protocol