In Vitro Study of a Stentless Aortic Bioprosthesis Made of Bacterial Cellulose

PURPOSE: The paper present findings from an in vitro experimental study of a stentless human aortic bioprosthesis (HAB) made of bacterial cellulose (BC). Three variants of the basic model were designed and tested to identify the valve prosthesis with the best performance parameters. The modified mod...

Descripción completa

Detalles Bibliográficos
Autores principales: Dawidowska, Kinga, Siondalski, Piotr, Kołaczkowska, Magdalena
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7782396/
https://www.ncbi.nlm.nih.gov/pubmed/33205361
http://dx.doi.org/10.1007/s13239-020-00500-z
Descripción
Sumario:PURPOSE: The paper present findings from an in vitro experimental study of a stentless human aortic bioprosthesis (HAB) made of bacterial cellulose (BC). Three variants of the basic model were designed and tested to identify the valve prosthesis with the best performance parameters. The modified models were made of BC, and the basic model of pericardium. METHODS: Each model (named V(1), V(2) and V(3)) was implanted into a 90 mm porcine aorta. Effective Orifice Area (EOA), rapid valve opening time (RVOT) and rapid valve closing time (RVCT) were determined. The flow resistance of each bioprosthesis model during the simulated heart systole, i.e. for the mean differential pressure (ΔP) at the time of full valve opening was measured. All experimental specimens were exposed to a mean blood pressure (MBP) of 90.5 ± 2.3 mmHg. RESULTS: The V(3) model demonstrated the best performance. The index defining the maximum opening of the bioprosthesis during systole for models V(1), V(2) and V(3) was 2.67 ± 0.59, 2.04 ± 0.23 and 2.85 ± 0.59 cm(2), respectively. The mean flow rate through the V(3) valve was 5.7 ± 1, 6.9 ± 0.7 and 8.9 ± 1.4 l/min for stroke volume (SV) of 65, 90 and 110 mL, respectively. The phase of immediate opening and closure for models V(1), V(2) and V(3) was 8, 7 and 5% of the cycle duration, respectively. The mean flow resistance of the models was: 4.07 ± 2.1, 4.28 ± 2.51 and 5.6 ± 2.32 mmHg. CONCLUSIONS: The V(3) model of the aortic valve prosthesis is the most effective. In vivo tests using BC as a structural material for this model are recommended. The response time of the V(3) model to changed work conditions is comparable to that of a healthy human heart. The model functions as an aortic valve prosthesis in in vitro conditions.