An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures
Background: The effect of transient electromagnetic fields on the function of pacemakers is not well-evaluated. There is a lack of effective methods for clinicians to reduce electromagnetic susceptibility (EMS) during implantation of pacemakers. This study aimed to evaluate whether a novel method of...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7785788/ https://www.ncbi.nlm.nih.gov/pubmed/33426004 http://dx.doi.org/10.3389/fcvm.2020.607604 |
_version_ | 1783632495295594496 |
---|---|
author | Huang, Jing Lin, Kaibin Lu, Wu Ding, Ranran Wu, Bingjie Cai, Mingqi Nazarian, Saman Zhao, Wenbin Li, Jingbo Huang, Dong |
author_facet | Huang, Jing Lin, Kaibin Lu, Wu Ding, Ranran Wu, Bingjie Cai, Mingqi Nazarian, Saman Zhao, Wenbin Li, Jingbo Huang, Dong |
author_sort | Huang, Jing |
collection | PubMed |
description | Background: The effect of transient electromagnetic fields on the function of pacemakers is not well-evaluated. There is a lack of effective methods for clinicians to reduce electromagnetic susceptibility (EMS) during implantation of pacemakers. This study aimed to evaluate whether a novel method of handling the excess leads in the pocket can lower the EMS of pacemakers and consequently reduce the effect of electromagnetic interference caused by transient electromagnetic fields on pacemakers. Methods: An in vitro chest model was established to simulate the clinical condition of dual-chamber pacemaker implantation. Three different intertwining patterns of excess leads were examined: parallel, twisted once, and multiple twisted-pair. Oscillated currents were injected into a copper electrical wire set horizontally above the model to create a radiated magnetic field to simulate the transient daily electromagnetic exposure of pacemakers. The electromagnetic induction of current was measured. The occurrence of EMS-related adverse events was evaluated when the induced pulsed voltage was applied. Results: Transient electromagnetic fields can induce electromagnetic noise in the pacing loop and inhibit the release of pacing pulses. The multiple twisted-pair intertwining pattern of excess leads was associated with a lower induced voltage amplitude than both the parallel and once-twisted patterns (P < 0.001). Even once twisted could significantly reduce induced voltage amplitude compared to not twisted (P < 0.001). A lower incidence of pacing inhibition was also observed in the multiple twisted-pair group than in the other two groups (P < 0.001). Conclusions: Transient electromagnetic fields can cause pacing inhibition. Twisting the excess leads for multiple turns in the pocket is an effective method to reduce the EMS of the dual-chamber pacemaker. |
format | Online Article Text |
id | pubmed-7785788 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-77857882021-01-07 An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures Huang, Jing Lin, Kaibin Lu, Wu Ding, Ranran Wu, Bingjie Cai, Mingqi Nazarian, Saman Zhao, Wenbin Li, Jingbo Huang, Dong Front Cardiovasc Med Cardiovascular Medicine Background: The effect of transient electromagnetic fields on the function of pacemakers is not well-evaluated. There is a lack of effective methods for clinicians to reduce electromagnetic susceptibility (EMS) during implantation of pacemakers. This study aimed to evaluate whether a novel method of handling the excess leads in the pocket can lower the EMS of pacemakers and consequently reduce the effect of electromagnetic interference caused by transient electromagnetic fields on pacemakers. Methods: An in vitro chest model was established to simulate the clinical condition of dual-chamber pacemaker implantation. Three different intertwining patterns of excess leads were examined: parallel, twisted once, and multiple twisted-pair. Oscillated currents were injected into a copper electrical wire set horizontally above the model to create a radiated magnetic field to simulate the transient daily electromagnetic exposure of pacemakers. The electromagnetic induction of current was measured. The occurrence of EMS-related adverse events was evaluated when the induced pulsed voltage was applied. Results: Transient electromagnetic fields can induce electromagnetic noise in the pacing loop and inhibit the release of pacing pulses. The multiple twisted-pair intertwining pattern of excess leads was associated with a lower induced voltage amplitude than both the parallel and once-twisted patterns (P < 0.001). Even once twisted could significantly reduce induced voltage amplitude compared to not twisted (P < 0.001). A lower incidence of pacing inhibition was also observed in the multiple twisted-pair group than in the other two groups (P < 0.001). Conclusions: Transient electromagnetic fields can cause pacing inhibition. Twisting the excess leads for multiple turns in the pocket is an effective method to reduce the EMS of the dual-chamber pacemaker. Frontiers Media S.A. 2020-12-23 /pmc/articles/PMC7785788/ /pubmed/33426004 http://dx.doi.org/10.3389/fcvm.2020.607604 Text en Copyright © 2020 Huang, Lin, Lu, Ding, Wu, Cai, Nazarian, Zhao, Li and Huang. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Cardiovascular Medicine Huang, Jing Lin, Kaibin Lu, Wu Ding, Ranran Wu, Bingjie Cai, Mingqi Nazarian, Saman Zhao, Wenbin Li, Jingbo Huang, Dong An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures |
title | An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures |
title_full | An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures |
title_fullStr | An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures |
title_full_unstemmed | An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures |
title_short | An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures |
title_sort | in vitro evaluation of the effect of transient electromagnetic fields on pacemakers and clinical mitigation measures |
topic | Cardiovascular Medicine |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7785788/ https://www.ncbi.nlm.nih.gov/pubmed/33426004 http://dx.doi.org/10.3389/fcvm.2020.607604 |
work_keys_str_mv | AT huangjing aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT linkaibin aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT luwu aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT dingranran aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT wubingjie aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT caimingqi aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT nazariansaman aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT zhaowenbin aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT lijingbo aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT huangdong aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT huangjing invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT linkaibin invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT luwu invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT dingranran invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT wubingjie invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT caimingqi invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT nazariansaman invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT zhaowenbin invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT lijingbo invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures AT huangdong invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures |