An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures

Background: The effect of transient electromagnetic fields on the function of pacemakers is not well-evaluated. There is a lack of effective methods for clinicians to reduce electromagnetic susceptibility (EMS) during implantation of pacemakers. This study aimed to evaluate whether a novel method of...

Descripción completa

Detalles Bibliográficos
Autores principales: Huang, Jing, Lin, Kaibin, Lu, Wu, Ding, Ranran, Wu, Bingjie, Cai, Mingqi, Nazarian, Saman, Zhao, Wenbin, Li, Jingbo, Huang, Dong
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7785788/
https://www.ncbi.nlm.nih.gov/pubmed/33426004
http://dx.doi.org/10.3389/fcvm.2020.607604
_version_ 1783632495295594496
author Huang, Jing
Lin, Kaibin
Lu, Wu
Ding, Ranran
Wu, Bingjie
Cai, Mingqi
Nazarian, Saman
Zhao, Wenbin
Li, Jingbo
Huang, Dong
author_facet Huang, Jing
Lin, Kaibin
Lu, Wu
Ding, Ranran
Wu, Bingjie
Cai, Mingqi
Nazarian, Saman
Zhao, Wenbin
Li, Jingbo
Huang, Dong
author_sort Huang, Jing
collection PubMed
description Background: The effect of transient electromagnetic fields on the function of pacemakers is not well-evaluated. There is a lack of effective methods for clinicians to reduce electromagnetic susceptibility (EMS) during implantation of pacemakers. This study aimed to evaluate whether a novel method of handling the excess leads in the pocket can lower the EMS of pacemakers and consequently reduce the effect of electromagnetic interference caused by transient electromagnetic fields on pacemakers. Methods: An in vitro chest model was established to simulate the clinical condition of dual-chamber pacemaker implantation. Three different intertwining patterns of excess leads were examined: parallel, twisted once, and multiple twisted-pair. Oscillated currents were injected into a copper electrical wire set horizontally above the model to create a radiated magnetic field to simulate the transient daily electromagnetic exposure of pacemakers. The electromagnetic induction of current was measured. The occurrence of EMS-related adverse events was evaluated when the induced pulsed voltage was applied. Results: Transient electromagnetic fields can induce electromagnetic noise in the pacing loop and inhibit the release of pacing pulses. The multiple twisted-pair intertwining pattern of excess leads was associated with a lower induced voltage amplitude than both the parallel and once-twisted patterns (P < 0.001). Even once twisted could significantly reduce induced voltage amplitude compared to not twisted (P < 0.001). A lower incidence of pacing inhibition was also observed in the multiple twisted-pair group than in the other two groups (P < 0.001). Conclusions: Transient electromagnetic fields can cause pacing inhibition. Twisting the excess leads for multiple turns in the pocket is an effective method to reduce the EMS of the dual-chamber pacemaker.
format Online
Article
Text
id pubmed-7785788
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-77857882021-01-07 An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures Huang, Jing Lin, Kaibin Lu, Wu Ding, Ranran Wu, Bingjie Cai, Mingqi Nazarian, Saman Zhao, Wenbin Li, Jingbo Huang, Dong Front Cardiovasc Med Cardiovascular Medicine Background: The effect of transient electromagnetic fields on the function of pacemakers is not well-evaluated. There is a lack of effective methods for clinicians to reduce electromagnetic susceptibility (EMS) during implantation of pacemakers. This study aimed to evaluate whether a novel method of handling the excess leads in the pocket can lower the EMS of pacemakers and consequently reduce the effect of electromagnetic interference caused by transient electromagnetic fields on pacemakers. Methods: An in vitro chest model was established to simulate the clinical condition of dual-chamber pacemaker implantation. Three different intertwining patterns of excess leads were examined: parallel, twisted once, and multiple twisted-pair. Oscillated currents were injected into a copper electrical wire set horizontally above the model to create a radiated magnetic field to simulate the transient daily electromagnetic exposure of pacemakers. The electromagnetic induction of current was measured. The occurrence of EMS-related adverse events was evaluated when the induced pulsed voltage was applied. Results: Transient electromagnetic fields can induce electromagnetic noise in the pacing loop and inhibit the release of pacing pulses. The multiple twisted-pair intertwining pattern of excess leads was associated with a lower induced voltage amplitude than both the parallel and once-twisted patterns (P < 0.001). Even once twisted could significantly reduce induced voltage amplitude compared to not twisted (P < 0.001). A lower incidence of pacing inhibition was also observed in the multiple twisted-pair group than in the other two groups (P < 0.001). Conclusions: Transient electromagnetic fields can cause pacing inhibition. Twisting the excess leads for multiple turns in the pocket is an effective method to reduce the EMS of the dual-chamber pacemaker. Frontiers Media S.A. 2020-12-23 /pmc/articles/PMC7785788/ /pubmed/33426004 http://dx.doi.org/10.3389/fcvm.2020.607604 Text en Copyright © 2020 Huang, Lin, Lu, Ding, Wu, Cai, Nazarian, Zhao, Li and Huang. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Cardiovascular Medicine
Huang, Jing
Lin, Kaibin
Lu, Wu
Ding, Ranran
Wu, Bingjie
Cai, Mingqi
Nazarian, Saman
Zhao, Wenbin
Li, Jingbo
Huang, Dong
An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures
title An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures
title_full An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures
title_fullStr An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures
title_full_unstemmed An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures
title_short An in vitro Evaluation of the Effect of Transient Electromagnetic Fields on Pacemakers and Clinical Mitigation Measures
title_sort in vitro evaluation of the effect of transient electromagnetic fields on pacemakers and clinical mitigation measures
topic Cardiovascular Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7785788/
https://www.ncbi.nlm.nih.gov/pubmed/33426004
http://dx.doi.org/10.3389/fcvm.2020.607604
work_keys_str_mv AT huangjing aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT linkaibin aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT luwu aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT dingranran aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT wubingjie aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT caimingqi aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT nazariansaman aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT zhaowenbin aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT lijingbo aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT huangdong aninvitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT huangjing invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT linkaibin invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT luwu invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT dingranran invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT wubingjie invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT caimingqi invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT nazariansaman invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT zhaowenbin invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT lijingbo invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures
AT huangdong invitroevaluationoftheeffectoftransientelectromagneticfieldsonpacemakersandclinicalmitigationmeasures