Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens

PURPOSE: To investigate the agreement and reliability of anterior segment optical coherence tomography (AS-OCT) and Scheimpflug imaging in measuring the distance from the anterior edge of an iris-fixated phakic intraocular lens (IF-pIOL) to the corneal endothelium. METHODS: Anterior segment configur...

Descripción completa

Detalles Bibliográficos
Autores principales: Gaurisankar, Zoraida S., van Rijn, Gwyneth A., Luyten, Gregorius P. M., Beenakker, Jan-Willem M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7790774/
https://www.ncbi.nlm.nih.gov/pubmed/32766949
http://dx.doi.org/10.1007/s00417-020-04874-7
_version_ 1783633492348764160
author Gaurisankar, Zoraida S.
van Rijn, Gwyneth A.
Luyten, Gregorius P. M.
Beenakker, Jan-Willem M.
author_facet Gaurisankar, Zoraida S.
van Rijn, Gwyneth A.
Luyten, Gregorius P. M.
Beenakker, Jan-Willem M.
author_sort Gaurisankar, Zoraida S.
collection PubMed
description PURPOSE: To investigate the agreement and reliability of anterior segment optical coherence tomography (AS-OCT) and Scheimpflug imaging in measuring the distance from the anterior edge of an iris-fixated phakic intraocular lens (IF-pIOL) to the corneal endothelium. METHODS: Anterior segment configuration was assessed in a total of 62 eyes of which 25 hyperopic and 37 myopic eyes, all corrected with an IF-pIOL. Measurements were performed by two independent observers using AS-OCT (Visante, Model 1000, Carl Zeiss Meditec Inc.) and Scheimpflug imaging (Pentacam HR, Oculus Optikgerate). The distance from the anterior edge of the pIOL to the endothelium was measured in five different positions using both modalities with their corresponding pIOL software. The measurements as well as the inter- and intra-observer reliability of the two imaging modalities were then compared. RESULTS: Distance measurements for all positions performed by AS-OCT were found to be significantly larger than those performed by Scheimpflug imaging, with mean differences ranging from 0.11 to 0.22 mm. Both instruments exhibited good inter- and intra-observer reliability. CONCLUSION: Anterior pIOL edge to endothelium distance measurements by AS-OCT and Scheimpflug imaging have good intra- and inter-observer reliability. However, as AS-OCT provides larger measurements, these two modalities cannot be used interchangeably. Correction of this difference might be essential for proper decision-making during pre-operative screening for pIOL implantation and post-operative safety monitoring. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1007/s00417-020-04874-7) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-7790774
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-77907742021-01-11 Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens Gaurisankar, Zoraida S. van Rijn, Gwyneth A. Luyten, Gregorius P. M. Beenakker, Jan-Willem M. Graefes Arch Clin Exp Ophthalmol Refractive Surgery PURPOSE: To investigate the agreement and reliability of anterior segment optical coherence tomography (AS-OCT) and Scheimpflug imaging in measuring the distance from the anterior edge of an iris-fixated phakic intraocular lens (IF-pIOL) to the corneal endothelium. METHODS: Anterior segment configuration was assessed in a total of 62 eyes of which 25 hyperopic and 37 myopic eyes, all corrected with an IF-pIOL. Measurements were performed by two independent observers using AS-OCT (Visante, Model 1000, Carl Zeiss Meditec Inc.) and Scheimpflug imaging (Pentacam HR, Oculus Optikgerate). The distance from the anterior edge of the pIOL to the endothelium was measured in five different positions using both modalities with their corresponding pIOL software. The measurements as well as the inter- and intra-observer reliability of the two imaging modalities were then compared. RESULTS: Distance measurements for all positions performed by AS-OCT were found to be significantly larger than those performed by Scheimpflug imaging, with mean differences ranging from 0.11 to 0.22 mm. Both instruments exhibited good inter- and intra-observer reliability. CONCLUSION: Anterior pIOL edge to endothelium distance measurements by AS-OCT and Scheimpflug imaging have good intra- and inter-observer reliability. However, as AS-OCT provides larger measurements, these two modalities cannot be used interchangeably. Correction of this difference might be essential for proper decision-making during pre-operative screening for pIOL implantation and post-operative safety monitoring. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1007/s00417-020-04874-7) contains supplementary material, which is available to authorized users. Springer Berlin Heidelberg 2020-08-06 2021 /pmc/articles/PMC7790774/ /pubmed/32766949 http://dx.doi.org/10.1007/s00417-020-04874-7 Text en © The Author(s) 2020 Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
spellingShingle Refractive Surgery
Gaurisankar, Zoraida S.
van Rijn, Gwyneth A.
Luyten, Gregorius P. M.
Beenakker, Jan-Willem M.
Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens
title Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens
title_full Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens
title_fullStr Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens
title_full_unstemmed Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens
title_short Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens
title_sort differences between scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens
topic Refractive Surgery
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7790774/
https://www.ncbi.nlm.nih.gov/pubmed/32766949
http://dx.doi.org/10.1007/s00417-020-04874-7
work_keys_str_mv AT gaurisankarzoraidas differencesbetweenscheimpflugandopticalcoherencetomographyindeterminingsafetydistancesineyeswithanirisfixatingphakicintraocularlens
AT vanrijngwynetha differencesbetweenscheimpflugandopticalcoherencetomographyindeterminingsafetydistancesineyeswithanirisfixatingphakicintraocularlens
AT luytengregoriuspm differencesbetweenscheimpflugandopticalcoherencetomographyindeterminingsafetydistancesineyeswithanirisfixatingphakicintraocularlens
AT beenakkerjanwillemm differencesbetweenscheimpflugandopticalcoherencetomographyindeterminingsafetydistancesineyeswithanirisfixatingphakicintraocularlens