Cargando…

A Contactless Laser Doppler Strain Sensor for Fatigue Testing with Resonance-Testing Machine

In this article, a non-contact laser Doppler strain sensor designed for fatigue testing with the resonance-testing machine is presented. The compact sensor measures in-plane displacements simultaneously from two adjacent points using the principle of in-plane, laser-Doppler vibrometry. The strain is...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Fangjian, Krause, Steffen, Hug, Joachim, Rembe, Christian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7796465/
https://www.ncbi.nlm.nih.gov/pubmed/33466507
http://dx.doi.org/10.3390/s21010319
Descripción
Sumario:In this article, a non-contact laser Doppler strain sensor designed for fatigue testing with the resonance-testing machine is presented. The compact sensor measures in-plane displacements simultaneously from two adjacent points using the principle of in-plane, laser-Doppler vibrometry. The strain is computed from the relative displacements divided by the distance between these two points. The optical design, the mathematical model for estimating noise-limited resolution, the simulation results of this model, and the first measurement results are presented. The comparison of the measurement results of our sensor with the results of a conventional strain gauge shows that our design meets the measurement requirements. The maximum strain deviation compared to conventional strain gauges of the laser-Doppler extensometer is below [Formula: see text] in all performed experiments.