Cargando…
Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
BACKGROUND AND PURPOSE: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807542/ https://www.ncbi.nlm.nih.gov/pubmed/33458386 http://dx.doi.org/10.1016/j.phro.2018.04.005 |
_version_ | 1783636762984185856 |
---|---|
author | Wyatt, Jonathan Hedley, Stephen Johnstone, Emily Speight, Richard Kelly, Charles Henry, Ann Short, Susan Murray, Louise Sebag-Montefiore, David McCallum, Hazel |
author_facet | Wyatt, Jonathan Hedley, Stephen Johnstone, Emily Speight, Richard Kelly, Charles Henry, Ann Short, Susan Murray, Louise Sebag-Montefiore, David McCallum, Hazel |
author_sort | Wyatt, Jonathan |
collection | PubMed |
description | BACKGROUND AND PURPOSE: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the sensitivity of these measurements to small set-up errors. MATERIALS AND METHODS: Geometric distortion was measured using a commercial phantom and software with 2D and 3D acquisition sequences on three different MR scanners. Two sets of repeatability measurements were made: three scans acquired without moving the phantom between scans (single set-up) and five scans acquired with the phantom re-set up in between each scan (repeated set-up). The set-up sensitivity was assessed by scanning the phantom with an intentional 1 mm lateral offset and independently an intentional 1° rotation. RESULTS: The mean standard deviation of distortion for all phantom markers for the repeated set-up scans was [Formula: see text] for all scanners and sequences. For the [Formula: see text] lateral offset scan [Formula: see text] of the markers agreed within two standard deviations of the mean of the repeated set-up scan (median of all scanners and sequences, range 78%–93%). For the 1° rotation scan, 80% of markers agreed within two standard deviations of the mean (range 69%–93%). CONCLUSIONS: Geometric distortion measurements using a commercial phantom and associated software appear repeatable, although with some sensitivity to set-up errors. This suggests the phantom and software are appropriate for commissioning a MR-only radiotherapy workflow. |
format | Online Article Text |
id | pubmed-7807542 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-78075422021-01-14 Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy Wyatt, Jonathan Hedley, Stephen Johnstone, Emily Speight, Richard Kelly, Charles Henry, Ann Short, Susan Murray, Louise Sebag-Montefiore, David McCallum, Hazel Phys Imaging Radiat Oncol Original Research Article BACKGROUND AND PURPOSE: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the sensitivity of these measurements to small set-up errors. MATERIALS AND METHODS: Geometric distortion was measured using a commercial phantom and software with 2D and 3D acquisition sequences on three different MR scanners. Two sets of repeatability measurements were made: three scans acquired without moving the phantom between scans (single set-up) and five scans acquired with the phantom re-set up in between each scan (repeated set-up). The set-up sensitivity was assessed by scanning the phantom with an intentional 1 mm lateral offset and independently an intentional 1° rotation. RESULTS: The mean standard deviation of distortion for all phantom markers for the repeated set-up scans was [Formula: see text] for all scanners and sequences. For the [Formula: see text] lateral offset scan [Formula: see text] of the markers agreed within two standard deviations of the mean of the repeated set-up scan (median of all scanners and sequences, range 78%–93%). For the 1° rotation scan, 80% of markers agreed within two standard deviations of the mean (range 69%–93%). CONCLUSIONS: Geometric distortion measurements using a commercial phantom and associated software appear repeatable, although with some sensitivity to set-up errors. This suggests the phantom and software are appropriate for commissioning a MR-only radiotherapy workflow. Elsevier 2018-05-01 /pmc/articles/PMC7807542/ /pubmed/33458386 http://dx.doi.org/10.1016/j.phro.2018.04.005 Text en © 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Original Research Article Wyatt, Jonathan Hedley, Stephen Johnstone, Emily Speight, Richard Kelly, Charles Henry, Ann Short, Susan Murray, Louise Sebag-Montefiore, David McCallum, Hazel Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy |
title | Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy |
title_full | Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy |
title_fullStr | Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy |
title_full_unstemmed | Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy |
title_short | Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy |
title_sort | evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy |
topic | Original Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807542/ https://www.ncbi.nlm.nih.gov/pubmed/33458386 http://dx.doi.org/10.1016/j.phro.2018.04.005 |
work_keys_str_mv | AT wyattjonathan evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT hedleystephen evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT johnstoneemily evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT speightrichard evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT kellycharles evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT henryann evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT shortsusan evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT murraylouise evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT sebagmontefioredavid evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy AT mccallumhazel evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy |