Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy

BACKGROUND AND PURPOSE: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the...

Descripción completa

Detalles Bibliográficos
Autores principales: Wyatt, Jonathan, Hedley, Stephen, Johnstone, Emily, Speight, Richard, Kelly, Charles, Henry, Ann, Short, Susan, Murray, Louise, Sebag-Montefiore, David, McCallum, Hazel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807542/
https://www.ncbi.nlm.nih.gov/pubmed/33458386
http://dx.doi.org/10.1016/j.phro.2018.04.005
_version_ 1783636762984185856
author Wyatt, Jonathan
Hedley, Stephen
Johnstone, Emily
Speight, Richard
Kelly, Charles
Henry, Ann
Short, Susan
Murray, Louise
Sebag-Montefiore, David
McCallum, Hazel
author_facet Wyatt, Jonathan
Hedley, Stephen
Johnstone, Emily
Speight, Richard
Kelly, Charles
Henry, Ann
Short, Susan
Murray, Louise
Sebag-Montefiore, David
McCallum, Hazel
author_sort Wyatt, Jonathan
collection PubMed
description BACKGROUND AND PURPOSE: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the sensitivity of these measurements to small set-up errors. MATERIALS AND METHODS: Geometric distortion was measured using a commercial phantom and software with 2D and 3D acquisition sequences on three different MR scanners. Two sets of repeatability measurements were made: three scans acquired without moving the phantom between scans (single set-up) and five scans acquired with the phantom re-set up in between each scan (repeated set-up). The set-up sensitivity was assessed by scanning the phantom with an intentional 1 mm lateral offset and independently an intentional 1° rotation. RESULTS: The mean standard deviation of distortion for all phantom markers for the repeated set-up scans was [Formula: see text] for all scanners and sequences. For the [Formula: see text] lateral offset scan [Formula: see text] of the markers agreed within two standard deviations of the mean of the repeated set-up scan (median of all scanners and sequences, range 78%–93%). For the 1° rotation scan, 80% of markers agreed within two standard deviations of the mean (range 69%–93%). CONCLUSIONS: Geometric distortion measurements using a commercial phantom and associated software appear repeatable, although with some sensitivity to set-up errors. This suggests the phantom and software are appropriate for commissioning a MR-only radiotherapy workflow.
format Online
Article
Text
id pubmed-7807542
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-78075422021-01-14 Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy Wyatt, Jonathan Hedley, Stephen Johnstone, Emily Speight, Richard Kelly, Charles Henry, Ann Short, Susan Murray, Louise Sebag-Montefiore, David McCallum, Hazel Phys Imaging Radiat Oncol Original Research Article BACKGROUND AND PURPOSE: Magnetic Resonance (MR)-only radiotherapy requires geometrically accurate MR images over the full scanner Field of View (FoV). This study aimed to investigate the repeatability of distortion measurements made using a commercial large FoV phantom and analysis software and the sensitivity of these measurements to small set-up errors. MATERIALS AND METHODS: Geometric distortion was measured using a commercial phantom and software with 2D and 3D acquisition sequences on three different MR scanners. Two sets of repeatability measurements were made: three scans acquired without moving the phantom between scans (single set-up) and five scans acquired with the phantom re-set up in between each scan (repeated set-up). The set-up sensitivity was assessed by scanning the phantom with an intentional 1 mm lateral offset and independently an intentional 1° rotation. RESULTS: The mean standard deviation of distortion for all phantom markers for the repeated set-up scans was [Formula: see text] for all scanners and sequences. For the [Formula: see text] lateral offset scan [Formula: see text] of the markers agreed within two standard deviations of the mean of the repeated set-up scan (median of all scanners and sequences, range 78%–93%). For the 1° rotation scan, 80% of markers agreed within two standard deviations of the mean (range 69%–93%). CONCLUSIONS: Geometric distortion measurements using a commercial phantom and associated software appear repeatable, although with some sensitivity to set-up errors. This suggests the phantom and software are appropriate for commissioning a MR-only radiotherapy workflow. Elsevier 2018-05-01 /pmc/articles/PMC7807542/ /pubmed/33458386 http://dx.doi.org/10.1016/j.phro.2018.04.005 Text en © 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Research Article
Wyatt, Jonathan
Hedley, Stephen
Johnstone, Emily
Speight, Richard
Kelly, Charles
Henry, Ann
Short, Susan
Murray, Louise
Sebag-Montefiore, David
McCallum, Hazel
Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
title Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
title_full Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
title_fullStr Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
title_full_unstemmed Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
title_short Evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
title_sort evaluating the repeatability and set-up sensitivity of a large field of view distortion phantom and software for magnetic resonance-only radiotherapy
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807542/
https://www.ncbi.nlm.nih.gov/pubmed/33458386
http://dx.doi.org/10.1016/j.phro.2018.04.005
work_keys_str_mv AT wyattjonathan evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT hedleystephen evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT johnstoneemily evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT speightrichard evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT kellycharles evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT henryann evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT shortsusan evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT murraylouise evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT sebagmontefioredavid evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy
AT mccallumhazel evaluatingtherepeatabilityandsetupsensitivityofalargefieldofviewdistortionphantomandsoftwareformagneticresonanceonlyradiotherapy