Investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans

BACKGROUND AND PURPOSE: Conical scintillation detectors are frequently used to measure geometric characteristics of radiotherapy modalities. However, their application to verify intensity-modulated radiotherapy plan delivery has not been investigated and requires a more detailed understanding of dev...

Descripción completa

Detalles Bibliográficos
Autores principales: Yock, Adam D., Johnson, Levi S., Price, Michael
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807557/
https://www.ncbi.nlm.nih.gov/pubmed/33458317
http://dx.doi.org/10.1016/j.phro.2020.06.001
_version_ 1783636766471749632
author Yock, Adam D.
Johnson, Levi S.
Price, Michael
author_facet Yock, Adam D.
Johnson, Levi S.
Price, Michael
author_sort Yock, Adam D.
collection PubMed
description BACKGROUND AND PURPOSE: Conical scintillation detectors are frequently used to measure geometric characteristics of radiotherapy modalities. However, their application to verify intensity-modulated radiotherapy plan delivery has not been investigated and requires a more detailed understanding of device response. This work evaluated the novel application of a conical scintillation detector to plan-specific quality assurance (QA) for intensity-modulated photon plans by evaluating device dependence on beam delivery and device acquisition parameters. MATERIALS AND METHODS: Measurements were made with a conical scintillation detector using beam delivery parameters of five photon beams (6–15 MV, including flattening filter free), three field sizes (1 × 1–5 × 5 cm(2)), and several dose rates (100–2000 MU/min) combined with device acquisition parameters of two frame rates (10 and 20 fps) and three gains (18–22 dB). A standardization equation to correct for gain and frame rate was investigated, and the remaining dose rate dependence was characterized. Device precision was evaluated using replicate measurements, and spatial uniformity was determined by irradiating different parts of the device. RESULTS: For each parameter combination, measurement reproducibility was 1.3%, and spatial uniformity was 1–2%. Scintillation intensity varied with gain, frame rate, and dose rate. Standardizing measurements for gain and frame rate was effective, but a dependence on dose rate caused errors at non-reference conditions (root mean squared error, RMSE: 0–152%). An additional dose rate correction specific to each combination of gain and frame rate improved accuracy (RMSE 0–17%). CONCLUSIONS: To consider the detector for plan-specific QA of intensity-modulated radiotherapy plans, correction factors are imperative to mitigate effects of delivery and acquisition parameters.
format Online
Article
Text
id pubmed-7807557
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-78075572021-01-14 Investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans Yock, Adam D. Johnson, Levi S. Price, Michael Phys Imaging Radiat Oncol Original Research Article BACKGROUND AND PURPOSE: Conical scintillation detectors are frequently used to measure geometric characteristics of radiotherapy modalities. However, their application to verify intensity-modulated radiotherapy plan delivery has not been investigated and requires a more detailed understanding of device response. This work evaluated the novel application of a conical scintillation detector to plan-specific quality assurance (QA) for intensity-modulated photon plans by evaluating device dependence on beam delivery and device acquisition parameters. MATERIALS AND METHODS: Measurements were made with a conical scintillation detector using beam delivery parameters of five photon beams (6–15 MV, including flattening filter free), three field sizes (1 × 1–5 × 5 cm(2)), and several dose rates (100–2000 MU/min) combined with device acquisition parameters of two frame rates (10 and 20 fps) and three gains (18–22 dB). A standardization equation to correct for gain and frame rate was investigated, and the remaining dose rate dependence was characterized. Device precision was evaluated using replicate measurements, and spatial uniformity was determined by irradiating different parts of the device. RESULTS: For each parameter combination, measurement reproducibility was 1.3%, and spatial uniformity was 1–2%. Scintillation intensity varied with gain, frame rate, and dose rate. Standardizing measurements for gain and frame rate was effective, but a dependence on dose rate caused errors at non-reference conditions (root mean squared error, RMSE: 0–152%). An additional dose rate correction specific to each combination of gain and frame rate improved accuracy (RMSE 0–17%). CONCLUSIONS: To consider the detector for plan-specific QA of intensity-modulated radiotherapy plans, correction factors are imperative to mitigate effects of delivery and acquisition parameters. Elsevier 2020-06-11 /pmc/articles/PMC7807557/ /pubmed/33458317 http://dx.doi.org/10.1016/j.phro.2020.06.001 Text en © 2020 Published by Elsevier B.V. on behalf of European Society of Radiotherapy & Oncology. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Research Article
Yock, Adam D.
Johnson, Levi S.
Price, Michael
Investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans
title Investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans
title_full Investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans
title_fullStr Investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans
title_full_unstemmed Investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans
title_short Investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans
title_sort investigating the potential of conical scintillation detectors for patient-specific verification of intensity-modulated radiotherapy plans
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807557/
https://www.ncbi.nlm.nih.gov/pubmed/33458317
http://dx.doi.org/10.1016/j.phro.2020.06.001
work_keys_str_mv AT yockadamd investigatingthepotentialofconicalscintillationdetectorsforpatientspecificverificationofintensitymodulatedradiotherapyplans
AT johnsonlevis investigatingthepotentialofconicalscintillationdetectorsforpatientspecificverificationofintensitymodulatedradiotherapyplans
AT pricemichael investigatingthepotentialofconicalscintillationdetectorsforpatientspecificverificationofintensitymodulatedradiotherapyplans