Cargando…

Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy

BACKGROUND AND PURPOSE: Independent dosimetry audits improve quality and safety of radiation therapy. This work reports on design and findings of a comprehensive 3D conformal radiotherapy (3D-CRT) Level III audit. MATERIALS AND METHODS: The audit was conducted as onsite audit using an anthropomorphi...

Descripción completa

Detalles Bibliográficos
Autores principales: Lehmann, Joerg, Alves, Andrew, Dunn, Leon, Shaw, Maddison, Kenny, John, Keehan, Stephanie, Supple, Jeremy, Gibbons, Francis, Manktelow, Sophie, Oliver, Chris, Kron, Tomas, Williams, Ivan, Lye, Jessica
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807562/
https://www.ncbi.nlm.nih.gov/pubmed/33458381
http://dx.doi.org/10.1016/j.phro.2018.03.006
_version_ 1783636767523471360
author Lehmann, Joerg
Alves, Andrew
Dunn, Leon
Shaw, Maddison
Kenny, John
Keehan, Stephanie
Supple, Jeremy
Gibbons, Francis
Manktelow, Sophie
Oliver, Chris
Kron, Tomas
Williams, Ivan
Lye, Jessica
author_facet Lehmann, Joerg
Alves, Andrew
Dunn, Leon
Shaw, Maddison
Kenny, John
Keehan, Stephanie
Supple, Jeremy
Gibbons, Francis
Manktelow, Sophie
Oliver, Chris
Kron, Tomas
Williams, Ivan
Lye, Jessica
author_sort Lehmann, Joerg
collection PubMed
description BACKGROUND AND PURPOSE: Independent dosimetry audits improve quality and safety of radiation therapy. This work reports on design and findings of a comprehensive 3D conformal radiotherapy (3D-CRT) Level III audit. MATERIALS AND METHODS: The audit was conducted as onsite audit using an anthropomorphic thorax phantom in an end-to-end test by the Australian Clinical Dosimetry Service (ACDS). Absolute dose point measurements were performed with Farmer-type ionization chambers. The audited treatment plans included open and half blocked fields, wedges and lung inhomogeneities. Audit results were determined as Pass Optimal Level (deviations within 3.3%), Pass Action Level (greater than 3.3% but within 5%) and Out of Tolerance (beyond 5%), as well as Reported Not Scored (RNS). The audit has been performed between July 2012 and January 2018 on 94 occasions, covering approximately 90% of all Australian facilities. RESULTS: The audit pass rate was 87% (53% optimal). Fifty recommendations were given, mainly related to planning system commissioning. Dose overestimation behind low density inhomogeneities by the analytical anisotropic algorithm (AAA) was identified across facilities and found to extend to beam setups which resemble a typical breast cancer treatment beam placement. RNS measurements inside lung showed a variation in the opposite direction: AAA under-dosed a target beyond lung and over-dosed the lung upstream and downstream of the target. Results also highlighted shortcomings of some superposition and convolution algorithms in modelling large angle wedges. CONCLUSIONS: This audit showed that 3D-CRT dosimetry audits remain relevant and can identify fundamental global and local problems that also affect advanced treatments.
format Online
Article
Text
id pubmed-7807562
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-78075622021-01-14 Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy Lehmann, Joerg Alves, Andrew Dunn, Leon Shaw, Maddison Kenny, John Keehan, Stephanie Supple, Jeremy Gibbons, Francis Manktelow, Sophie Oliver, Chris Kron, Tomas Williams, Ivan Lye, Jessica Phys Imaging Radiat Oncol Original Research Article BACKGROUND AND PURPOSE: Independent dosimetry audits improve quality and safety of radiation therapy. This work reports on design and findings of a comprehensive 3D conformal radiotherapy (3D-CRT) Level III audit. MATERIALS AND METHODS: The audit was conducted as onsite audit using an anthropomorphic thorax phantom in an end-to-end test by the Australian Clinical Dosimetry Service (ACDS). Absolute dose point measurements were performed with Farmer-type ionization chambers. The audited treatment plans included open and half blocked fields, wedges and lung inhomogeneities. Audit results were determined as Pass Optimal Level (deviations within 3.3%), Pass Action Level (greater than 3.3% but within 5%) and Out of Tolerance (beyond 5%), as well as Reported Not Scored (RNS). The audit has been performed between July 2012 and January 2018 on 94 occasions, covering approximately 90% of all Australian facilities. RESULTS: The audit pass rate was 87% (53% optimal). Fifty recommendations were given, mainly related to planning system commissioning. Dose overestimation behind low density inhomogeneities by the analytical anisotropic algorithm (AAA) was identified across facilities and found to extend to beam setups which resemble a typical breast cancer treatment beam placement. RNS measurements inside lung showed a variation in the opposite direction: AAA under-dosed a target beyond lung and over-dosed the lung upstream and downstream of the target. Results also highlighted shortcomings of some superposition and convolution algorithms in modelling large angle wedges. CONCLUSIONS: This audit showed that 3D-CRT dosimetry audits remain relevant and can identify fundamental global and local problems that also affect advanced treatments. Elsevier 2018-04-24 /pmc/articles/PMC7807562/ /pubmed/33458381 http://dx.doi.org/10.1016/j.phro.2018.03.006 Text en Crown Copyright © 2018 Published by Elsevier B.V. on behalf of European Society of Radiotherapy & Oncology. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Research Article
Lehmann, Joerg
Alves, Andrew
Dunn, Leon
Shaw, Maddison
Kenny, John
Keehan, Stephanie
Supple, Jeremy
Gibbons, Francis
Manktelow, Sophie
Oliver, Chris
Kron, Tomas
Williams, Ivan
Lye, Jessica
Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy
title Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy
title_full Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy
title_fullStr Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy
title_full_unstemmed Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy
title_short Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy
title_sort dosimetric end-to-end tests in a national audit of 3d conformal radiotherapy
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807562/
https://www.ncbi.nlm.nih.gov/pubmed/33458381
http://dx.doi.org/10.1016/j.phro.2018.03.006
work_keys_str_mv AT lehmannjoerg dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT alvesandrew dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT dunnleon dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT shawmaddison dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT kennyjohn dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT keehanstephanie dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT supplejeremy dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT gibbonsfrancis dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT manktelowsophie dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT oliverchris dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT krontomas dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT williamsivan dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy
AT lyejessica dosimetricendtoendtestsinanationalauditof3dconformalradiotherapy