Cargando…

A multi-centre analytical study of small field output factor calculations in radiotherapy

An audit methodology was developed and applied for output factor (OF) calculations in radiotherapy. The auditees were asked to calculate OFs for field sizes from 10 × 10 cm(2) to 2 × 2 cm(2). Sixty five beams were audited; missing reference OFs were interpolated. The calculated OFs were in 73% of ca...

Descripción completa

Detalles Bibliográficos
Autores principales: Chełmiński, Krzysztof, Bulski, Wojciech
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807585/
https://www.ncbi.nlm.nih.gov/pubmed/33458380
http://dx.doi.org/10.1016/j.phro.2018.03.001