Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours

BACKGROUND AND PURPOSE: In neuro-oncology, high spatial accuracy is needed for clinically acceptable high-precision radiation treatment planning (RTP). In this study, the clinical applicability of anatomically optimised 7-Tesla (7T) MR images for reliable RTP is assessed with respect to standard cli...

Descripción completa

Detalles Bibliográficos
Autores principales: Peerlings, Jurgen, Compter, Inge, Janssen, Fiere, Wiggins, Christopher J., Postma, Alida A., Mottaghy, Felix M., Lambin, Philippe, Hoffmann, Aswin L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807620/
https://www.ncbi.nlm.nih.gov/pubmed/33458423
http://dx.doi.org/10.1016/j.phro.2018.12.001
_version_ 1783636780624379904
author Peerlings, Jurgen
Compter, Inge
Janssen, Fiere
Wiggins, Christopher J.
Postma, Alida A.
Mottaghy, Felix M.
Lambin, Philippe
Hoffmann, Aswin L.
author_facet Peerlings, Jurgen
Compter, Inge
Janssen, Fiere
Wiggins, Christopher J.
Postma, Alida A.
Mottaghy, Felix M.
Lambin, Philippe
Hoffmann, Aswin L.
author_sort Peerlings, Jurgen
collection PubMed
description BACKGROUND AND PURPOSE: In neuro-oncology, high spatial accuracy is needed for clinically acceptable high-precision radiation treatment planning (RTP). In this study, the clinical applicability of anatomically optimised 7-Tesla (7T) MR images for reliable RTP is assessed with respect to standard clinical imaging modalities. MATERIALS AND METHODS: System- and phantom-related geometrical distortion (GD) were quantified on clinically-relevant MR sequences at 7T and 3T, and on CT images using a dedicated anthropomorphic head phantom incorporating a 3D grid-structure, creating 436 points-of-interest. Global GD was assessed by mean absolute deviation (MAD(Global)). Local GD relative to the magnetic isocentre was assessed by MAD(Local). Using 3D displacement vectors of individual points-of-interest, GD maps were created. For clinically acceptable radiotherapy, 7T images need to meet the criteria for accurate dose delivery (GD < 1 mm) and present comparable GD as tolerated in clinically standard 3T MR/CT-based RTP. RESULTS: MAD(Global) in 7T and 3T images ranged from 0.3 to 2.2 mm and 0.2–0.8 mm, respectively. MAD(Local) increased with increasing distance from the isocentre, showed an anisotropic distribution, and was significantly larger in 7T MR sequences (MAD(Local) = 0.2–1.2 mm) than in 3T (MAD(Local) = 0.1–0.7 mm) (p < 0.05). Significant differences in GD were detected between 7T images (p < 0.001). However, maximum MAD(Local) remained ≤1 mm within 68.7 mm diameter spherical volume. No significant differences in GD were found between 7T and 3T protocols near the isocentre. CONCLUSIONS: System- and phantom-related GD remained ≤1 mm in central brain regions, suggesting that 7T MR images could be implemented in radiotherapy with clinically acceptable spatial accuracy and equally tolerated GD as in 3T MR/CT-based RTP. For peripheral regions, GD should be incorporated in safety margins for treatment uncertainties. Moreover, the effects of sequence-related factors on GD needs further investigation to obtain RTP-specific MR protocols.
format Online
Article
Text
id pubmed-7807620
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-78076202021-01-14 Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours Peerlings, Jurgen Compter, Inge Janssen, Fiere Wiggins, Christopher J. Postma, Alida A. Mottaghy, Felix M. Lambin, Philippe Hoffmann, Aswin L. Phys Imaging Radiat Oncol Original Research Article BACKGROUND AND PURPOSE: In neuro-oncology, high spatial accuracy is needed for clinically acceptable high-precision radiation treatment planning (RTP). In this study, the clinical applicability of anatomically optimised 7-Tesla (7T) MR images for reliable RTP is assessed with respect to standard clinical imaging modalities. MATERIALS AND METHODS: System- and phantom-related geometrical distortion (GD) were quantified on clinically-relevant MR sequences at 7T and 3T, and on CT images using a dedicated anthropomorphic head phantom incorporating a 3D grid-structure, creating 436 points-of-interest. Global GD was assessed by mean absolute deviation (MAD(Global)). Local GD relative to the magnetic isocentre was assessed by MAD(Local). Using 3D displacement vectors of individual points-of-interest, GD maps were created. For clinically acceptable radiotherapy, 7T images need to meet the criteria for accurate dose delivery (GD < 1 mm) and present comparable GD as tolerated in clinically standard 3T MR/CT-based RTP. RESULTS: MAD(Global) in 7T and 3T images ranged from 0.3 to 2.2 mm and 0.2–0.8 mm, respectively. MAD(Local) increased with increasing distance from the isocentre, showed an anisotropic distribution, and was significantly larger in 7T MR sequences (MAD(Local) = 0.2–1.2 mm) than in 3T (MAD(Local) = 0.1–0.7 mm) (p < 0.05). Significant differences in GD were detected between 7T images (p < 0.001). However, maximum MAD(Local) remained ≤1 mm within 68.7 mm diameter spherical volume. No significant differences in GD were found between 7T and 3T protocols near the isocentre. CONCLUSIONS: System- and phantom-related GD remained ≤1 mm in central brain regions, suggesting that 7T MR images could be implemented in radiotherapy with clinically acceptable spatial accuracy and equally tolerated GD as in 3T MR/CT-based RTP. For peripheral regions, GD should be incorporated in safety margins for treatment uncertainties. Moreover, the effects of sequence-related factors on GD needs further investigation to obtain RTP-specific MR protocols. Elsevier 2019-01-03 /pmc/articles/PMC7807620/ /pubmed/33458423 http://dx.doi.org/10.1016/j.phro.2018.12.001 Text en © 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Research Article
Peerlings, Jurgen
Compter, Inge
Janssen, Fiere
Wiggins, Christopher J.
Postma, Alida A.
Mottaghy, Felix M.
Lambin, Philippe
Hoffmann, Aswin L.
Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours
title Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours
title_full Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours
title_fullStr Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours
title_full_unstemmed Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours
title_short Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours
title_sort characterizing geometrical accuracy in clinically optimised 7t and 3t magnetic resonance images for high-precision radiation treatment of brain tumours
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807620/
https://www.ncbi.nlm.nih.gov/pubmed/33458423
http://dx.doi.org/10.1016/j.phro.2018.12.001
work_keys_str_mv AT peerlingsjurgen characterizinggeometricalaccuracyinclinicallyoptimised7tand3tmagneticresonanceimagesforhighprecisionradiationtreatmentofbraintumours
AT compteringe characterizinggeometricalaccuracyinclinicallyoptimised7tand3tmagneticresonanceimagesforhighprecisionradiationtreatmentofbraintumours
AT janssenfiere characterizinggeometricalaccuracyinclinicallyoptimised7tand3tmagneticresonanceimagesforhighprecisionradiationtreatmentofbraintumours
AT wigginschristopherj characterizinggeometricalaccuracyinclinicallyoptimised7tand3tmagneticresonanceimagesforhighprecisionradiationtreatmentofbraintumours
AT postmaalidaa characterizinggeometricalaccuracyinclinicallyoptimised7tand3tmagneticresonanceimagesforhighprecisionradiationtreatmentofbraintumours
AT mottaghyfelixm characterizinggeometricalaccuracyinclinicallyoptimised7tand3tmagneticresonanceimagesforhighprecisionradiationtreatmentofbraintumours
AT lambinphilippe characterizinggeometricalaccuracyinclinicallyoptimised7tand3tmagneticresonanceimagesforhighprecisionradiationtreatmentofbraintumours
AT hoffmannaswinl characterizinggeometricalaccuracyinclinicallyoptimised7tand3tmagneticresonanceimagesforhighprecisionradiationtreatmentofbraintumours