Clinical implementation of low-dose total body irradiation using topotherapy technique
BACKGROUND AND PURPOSE: The topotherapy technique was recently suggested as a robust alternative to helical radiation delivery for total body irradiation (TBI). It allows to deliver a discrete number of beams with fixed gantry. A Topotherapy-based low-dose TBI technique was optimized and clinically...
Autores principales: | , , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807637/ https://www.ncbi.nlm.nih.gov/pubmed/33458299 http://dx.doi.org/10.1016/j.phro.2019.11.009 |
Sumario: | BACKGROUND AND PURPOSE: The topotherapy technique was recently suggested as a robust alternative to helical radiation delivery for total body irradiation (TBI). It allows to deliver a discrete number of beams with fixed gantry. A Topotherapy-based low-dose TBI technique was optimized and clinically implemented. MATERIALS AND METHODS: TBI delivery was split in two parts: the first treating from the head to half thigh and the second the remaining legs. An in-silico investigation aimed to optimize plan parameters was first carried out on four patients. For the upper plan, field width and pitch were fixed to 5 cm and 0.5: the combined impact of five modulation factor (MF) values and different field configurations (6/8/12 fields) was investigated. For the lower plan, two anterior/posterior beams (field width: 5 cm; pitch: 0.5; MF:1.5) were used. After assessing the optimal technique, set-up/quality assurance/image-guidance procedures were defined and the technique clinically implemented: 23 patients were treated up to now. RESULTS: The best compromise between treatment time and planning target volume (PTV) coverage/homogeneity was found for MF = 1.5 and 8 fields. All clinical plans were automatically optimized using an “ad-hoc” plan template: excellent PTV coverage (PTV95%>98.5%) and homogeneity (median SD:4%) were found with a median beam-on time of 17/9 min for the upper/lower plan. All patients were successfully treated and transplanted. CONCLUSIONS: TBI delivered with the topotherapy approach robustly guarantees adequate coverage and dose homogeneity. Semi-automatic clinical plans can be quickly generated and efficiently delivered. |
---|