Cargando…

An ultrasound based platform for image-guided radiotherapy in canine bladder cancer patients

BACKGROUND AND PURPOSE: Ultrasound (US) is a non-invasive, non-radiographic imaging technique with high spatial and temporal resolution that can be used for localizing soft-tissue structures and tumors in real-time during radiotherapy (RT) (inter- and intra-fraction). A comprehensive approach incorp...

Descripción completa

Detalles Bibliográficos
Autores principales: Sick, Justin T., Rancilio, Nicholas J., Fulkerson, Caroline V., Plantenga, Jeannie M., Knapp, Deborah W., Stantz, Keith M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807639/
https://www.ncbi.nlm.nih.gov/pubmed/33458289
http://dx.doi.org/10.1016/j.phro.2019.10.003
Descripción
Sumario:BACKGROUND AND PURPOSE: Ultrasound (US) is a non-invasive, non-radiographic imaging technique with high spatial and temporal resolution that can be used for localizing soft-tissue structures and tumors in real-time during radiotherapy (RT) (inter- and intra-fraction). A comprehensive approach incorporating an in-house 3D-US system within RT is presented. This system is easier to adopt into existing treatment protocols than current US based systems, with the aim of providing millimeter intra-fraction alignment errors and sensitivity to track intra-fraction bladder movement. MATERIALS AND METHODS: An in-house integrated US manipulator and platform was designed to relate the computed tomographic (CT) scanner, 3D-US and linear accelerator coordinate systems. An agar-based phantom with measured speed of sound and densities consistent with tissues surrounding the bladder was rotated (0–45°) and translated (up to 55 mm) relative to the US and CT coordinate systems to validate this device. After acquiring and integrating CT and US images into the treatment planning system, US-to-US and US-to-CT images were co-registered to re-align the phantom relative to the linear accelerator. RESULTS: Statistical errors from US-to-US registrations for various patient orientations ranged from 0.1 to 1.7 mm for x, y, and z translation components, and 0.0–1.1° for rotational components. Statistical errors from US-to-CT registrations were 0.3–1.2 mm for the x, y and z translational components and 0.1–2.5° for the rotational components. CONCLUSIONS: An ultrasound-based platform was designed, constructed and tested on a CT/US tissue-equivalent phantom to track bladder displacement with a statistical uncertainty to correct and track inter- and intra-fractional displacements of the bladder during radiation treatments.