Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans

BACKGROUND AND PURPOSE: Local implementation of plan-specific quality assurance (QA) methods for intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) treatment plans may vary because of dissimilarities in procedures, equipment and software. The purpose of this work is...

Descripción completa

Detalles Bibliográficos
Autores principales: Seravalli, Enrica, Houweling, Antonetta C., Van Battum, Leo, Raaben, Thom A., Kuik, Marc, de Pooter, Jacco A., Van Gellekom, Marion P.R., Kaas, Jochem, de Vries, Wilfred, Loeff, Erik A., Van de Kamer, Jeroen B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807668/
https://www.ncbi.nlm.nih.gov/pubmed/33458364
http://dx.doi.org/10.1016/j.phro.2018.01.002
_version_ 1783636790991650816
author Seravalli, Enrica
Houweling, Antonetta C.
Van Battum, Leo
Raaben, Thom A.
Kuik, Marc
de Pooter, Jacco A.
Van Gellekom, Marion P.R.
Kaas, Jochem
de Vries, Wilfred
Loeff, Erik A.
Van de Kamer, Jeroen B.
author_facet Seravalli, Enrica
Houweling, Antonetta C.
Van Battum, Leo
Raaben, Thom A.
Kuik, Marc
de Pooter, Jacco A.
Van Gellekom, Marion P.R.
Kaas, Jochem
de Vries, Wilfred
Loeff, Erik A.
Van de Kamer, Jeroen B.
author_sort Seravalli, Enrica
collection PubMed
description BACKGROUND AND PURPOSE: Local implementation of plan-specific quality assurance (QA) methods for intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) treatment plans may vary because of dissimilarities in procedures, equipment and software. The purpose of this work is detecting possible differences between local QA findings and those of an audit, using the same set of treatment plans. METHODS: A pre-defined set of clinical plans was devised and imported in the participating institute’s treatment planning system for dose computation. The dose distribution was measured using an ionisation chamber, radiochromic film and an ionisation chamber array. The centres performed their own QA, which was compared to the audit findings. The agreement/disagreement between the audit and the institute QA results were assessed along with the differences between the dose distributions measured by the audit team and computed by the institute. RESULTS: For the majority of the cases the results of the audit were in agreement with the institute QA findings: ionisation chamber: 92%, array: 88%, film: 76% of the total measurements. In only a few of these cases the evaluated measurements failed for both: ionisation chamber: 2%, array: 4%, film: 0% of the total measurements. CONCLUSION: Using predefined treatment plans, we found that in approximately 80% of the evaluated measurements the results of local QA of IMRT and VMAT plans were in line with the findings of the audit. However, the percentage of agreement/disagreement depended on the characteristics of the measurement equipment used and on the analysis metric.
format Online
Article
Text
id pubmed-7807668
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-78076682021-01-14 Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans Seravalli, Enrica Houweling, Antonetta C. Van Battum, Leo Raaben, Thom A. Kuik, Marc de Pooter, Jacco A. Van Gellekom, Marion P.R. Kaas, Jochem de Vries, Wilfred Loeff, Erik A. Van de Kamer, Jeroen B. Phys Imaging Radiat Oncol Original Research Article BACKGROUND AND PURPOSE: Local implementation of plan-specific quality assurance (QA) methods for intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) treatment plans may vary because of dissimilarities in procedures, equipment and software. The purpose of this work is detecting possible differences between local QA findings and those of an audit, using the same set of treatment plans. METHODS: A pre-defined set of clinical plans was devised and imported in the participating institute’s treatment planning system for dose computation. The dose distribution was measured using an ionisation chamber, radiochromic film and an ionisation chamber array. The centres performed their own QA, which was compared to the audit findings. The agreement/disagreement between the audit and the institute QA results were assessed along with the differences between the dose distributions measured by the audit team and computed by the institute. RESULTS: For the majority of the cases the results of the audit were in agreement with the institute QA findings: ionisation chamber: 92%, array: 88%, film: 76% of the total measurements. In only a few of these cases the evaluated measurements failed for both: ionisation chamber: 2%, array: 4%, film: 0% of the total measurements. CONCLUSION: Using predefined treatment plans, we found that in approximately 80% of the evaluated measurements the results of local QA of IMRT and VMAT plans were in line with the findings of the audit. However, the percentage of agreement/disagreement depended on the characteristics of the measurement equipment used and on the analysis metric. Elsevier 2018-02-02 /pmc/articles/PMC7807668/ /pubmed/33458364 http://dx.doi.org/10.1016/j.phro.2018.01.002 Text en © 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Research Article
Seravalli, Enrica
Houweling, Antonetta C.
Van Battum, Leo
Raaben, Thom A.
Kuik, Marc
de Pooter, Jacco A.
Van Gellekom, Marion P.R.
Kaas, Jochem
de Vries, Wilfred
Loeff, Erik A.
Van de Kamer, Jeroen B.
Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans
title Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans
title_full Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans
title_fullStr Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans
title_full_unstemmed Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans
title_short Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans
title_sort auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807668/
https://www.ncbi.nlm.nih.gov/pubmed/33458364
http://dx.doi.org/10.1016/j.phro.2018.01.002
work_keys_str_mv AT seravallienrica auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT houwelingantonettac auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT vanbattumleo auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT raabenthoma auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT kuikmarc auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT depooterjaccoa auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT vangellekommarionpr auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT kaasjochem auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT devrieswilfred auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT loefferika auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans
AT vandekamerjeroenb auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans