Cargando…
Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans
BACKGROUND AND PURPOSE: Local implementation of plan-specific quality assurance (QA) methods for intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) treatment plans may vary because of dissimilarities in procedures, equipment and software. The purpose of this work is...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807668/ https://www.ncbi.nlm.nih.gov/pubmed/33458364 http://dx.doi.org/10.1016/j.phro.2018.01.002 |
_version_ | 1783636790991650816 |
---|---|
author | Seravalli, Enrica Houweling, Antonetta C. Van Battum, Leo Raaben, Thom A. Kuik, Marc de Pooter, Jacco A. Van Gellekom, Marion P.R. Kaas, Jochem de Vries, Wilfred Loeff, Erik A. Van de Kamer, Jeroen B. |
author_facet | Seravalli, Enrica Houweling, Antonetta C. Van Battum, Leo Raaben, Thom A. Kuik, Marc de Pooter, Jacco A. Van Gellekom, Marion P.R. Kaas, Jochem de Vries, Wilfred Loeff, Erik A. Van de Kamer, Jeroen B. |
author_sort | Seravalli, Enrica |
collection | PubMed |
description | BACKGROUND AND PURPOSE: Local implementation of plan-specific quality assurance (QA) methods for intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) treatment plans may vary because of dissimilarities in procedures, equipment and software. The purpose of this work is detecting possible differences between local QA findings and those of an audit, using the same set of treatment plans. METHODS: A pre-defined set of clinical plans was devised and imported in the participating institute’s treatment planning system for dose computation. The dose distribution was measured using an ionisation chamber, radiochromic film and an ionisation chamber array. The centres performed their own QA, which was compared to the audit findings. The agreement/disagreement between the audit and the institute QA results were assessed along with the differences between the dose distributions measured by the audit team and computed by the institute. RESULTS: For the majority of the cases the results of the audit were in agreement with the institute QA findings: ionisation chamber: 92%, array: 88%, film: 76% of the total measurements. In only a few of these cases the evaluated measurements failed for both: ionisation chamber: 2%, array: 4%, film: 0% of the total measurements. CONCLUSION: Using predefined treatment plans, we found that in approximately 80% of the evaluated measurements the results of local QA of IMRT and VMAT plans were in line with the findings of the audit. However, the percentage of agreement/disagreement depended on the characteristics of the measurement equipment used and on the analysis metric. |
format | Online Article Text |
id | pubmed-7807668 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-78076682021-01-14 Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans Seravalli, Enrica Houweling, Antonetta C. Van Battum, Leo Raaben, Thom A. Kuik, Marc de Pooter, Jacco A. Van Gellekom, Marion P.R. Kaas, Jochem de Vries, Wilfred Loeff, Erik A. Van de Kamer, Jeroen B. Phys Imaging Radiat Oncol Original Research Article BACKGROUND AND PURPOSE: Local implementation of plan-specific quality assurance (QA) methods for intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) treatment plans may vary because of dissimilarities in procedures, equipment and software. The purpose of this work is detecting possible differences between local QA findings and those of an audit, using the same set of treatment plans. METHODS: A pre-defined set of clinical plans was devised and imported in the participating institute’s treatment planning system for dose computation. The dose distribution was measured using an ionisation chamber, radiochromic film and an ionisation chamber array. The centres performed their own QA, which was compared to the audit findings. The agreement/disagreement between the audit and the institute QA results were assessed along with the differences between the dose distributions measured by the audit team and computed by the institute. RESULTS: For the majority of the cases the results of the audit were in agreement with the institute QA findings: ionisation chamber: 92%, array: 88%, film: 76% of the total measurements. In only a few of these cases the evaluated measurements failed for both: ionisation chamber: 2%, array: 4%, film: 0% of the total measurements. CONCLUSION: Using predefined treatment plans, we found that in approximately 80% of the evaluated measurements the results of local QA of IMRT and VMAT plans were in line with the findings of the audit. However, the percentage of agreement/disagreement depended on the characteristics of the measurement equipment used and on the analysis metric. Elsevier 2018-02-02 /pmc/articles/PMC7807668/ /pubmed/33458364 http://dx.doi.org/10.1016/j.phro.2018.01.002 Text en © 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Original Research Article Seravalli, Enrica Houweling, Antonetta C. Van Battum, Leo Raaben, Thom A. Kuik, Marc de Pooter, Jacco A. Van Gellekom, Marion P.R. Kaas, Jochem de Vries, Wilfred Loeff, Erik A. Van de Kamer, Jeroen B. Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans |
title | Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans |
title_full | Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans |
title_fullStr | Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans |
title_full_unstemmed | Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans |
title_short | Auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans |
title_sort | auditing local methods for quality assurance in radiotherapy using the same set of predefined treatment plans |
topic | Original Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807668/ https://www.ncbi.nlm.nih.gov/pubmed/33458364 http://dx.doi.org/10.1016/j.phro.2018.01.002 |
work_keys_str_mv | AT seravallienrica auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT houwelingantonettac auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT vanbattumleo auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT raabenthoma auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT kuikmarc auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT depooterjaccoa auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT vangellekommarionpr auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT kaasjochem auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT devrieswilfred auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT loefferika auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans AT vandekamerjeroenb auditinglocalmethodsforqualityassuranceinradiotherapyusingthesamesetofpredefinedtreatmentplans |