Cargando…

Evaluation of multi-institutional end-to-end testing for post-operative spine stereotactic body radiation therapy

BACKGROUND AND PURPOSE: Post-operative spine stereotactic body radiation therapy (SBRT) represents a significant challenge as there are many restrictions on beam geometry to avoid metal hardware as it surrounds the target volume. In this study, an international multi-institutional end-to-end test us...

Descripción completa

Detalles Bibliográficos
Autores principales: Furuya, Tomohisa, Lee, Young K., Archibald-Heeren, Ben R., Byrne, Mikel, Bosco, Bruno, Phua, Jun H., Shimizu, Hidetoshi, Hashimoto, Shimpei, Tanaka, Hiroshi, Sahgal, Arjun, Karasawa, Katsuyuki
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7807765/
https://www.ncbi.nlm.nih.gov/pubmed/33458345
http://dx.doi.org/10.1016/j.phro.2020.09.005
Descripción
Sumario:BACKGROUND AND PURPOSE: Post-operative spine stereotactic body radiation therapy (SBRT) represents a significant challenge as there are many restrictions on beam geometry to avoid metal hardware as it surrounds the target volume. In this study, an international multi-institutional end-to-end test using an in-house spine phantom was developed and executed. The aim was to evaluate the impact of titanium spine hardware on planned and delivered dose for post-operative spine SBRT. MATERIALS AND METHODS: Five centers performed simulation, planning and irradiation of the spine phantom, with/without titanium metal hardware (MB/B), following our pre-specified protocol. The doses were calculated using the centers’ treatment planning system (TPS) and measured with radiophotoluminescent glass dosimeters (RPLDs) embedded within each phantom. RESULTS: The dose differences between the RPLD measured and calculated doses in the target region were within ± 5% for both phantoms studied. Differences greater than 5% were observed for the spinal cord and the out-of-the target regions due to steeper dose gradient regions that are created in these plans. Dose measurements within ± 3% were observed between RPLDs that were embedded in MB and B inserts. For the spinal cord and the out-of-target regions surrounded by metal hardware, the dose measured using RPLDs was within 3% different near the titanium screws compared to the dose measured near only the metal rods. CONCLUSION: We have successfully performed the first multi-institutional end-to-end dose analysis using an in-house phantom built specifically for post-operative spine SBRT. The differences observed between the measured and planned doses in the presence of metal hardware were clinically insignificant.