A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study

Nowadays, given the technological advance in CT imaging and increasing heterogeneity in characteristics of CT scanners, a number of CT scanners with different manufacturers/technologies are often installed in a hospital centre and used by various departments. In this phantom study, a comprehensive a...

Descripción completa

Detalles Bibliográficos
Autores principales: Barca, Patrizio, Paolicchi, Fabio, Aringhieri, Giacomo, Palmas, Federica, Marfisi, Daniela, Fantacci, Maria Evelina, Caramella, Davide, Giannelli, Marco
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7808662/
https://www.ncbi.nlm.nih.gov/pubmed/33444367
http://dx.doi.org/10.1371/journal.pone.0245374
_version_ 1783636948434288640
author Barca, Patrizio
Paolicchi, Fabio
Aringhieri, Giacomo
Palmas, Federica
Marfisi, Daniela
Fantacci, Maria Evelina
Caramella, Davide
Giannelli, Marco
author_facet Barca, Patrizio
Paolicchi, Fabio
Aringhieri, Giacomo
Palmas, Federica
Marfisi, Daniela
Fantacci, Maria Evelina
Caramella, Davide
Giannelli, Marco
author_sort Barca, Patrizio
collection PubMed
description Nowadays, given the technological advance in CT imaging and increasing heterogeneity in characteristics of CT scanners, a number of CT scanners with different manufacturers/technologies are often installed in a hospital centre and used by various departments. In this phantom study, a comprehensive assessment of image quality of 5 scanners (from 3 manufacturers and with different models) for head CT imaging, as clinically used at a single hospital centre, was hence carried out. Helical and/or sequential acquisitions of the Catphan-504 phantom were performed, using the scanning protocols (CTDI(vol) range: 54.7–57.5 mGy) employed by the staff of various Radiology/Neuroradiology departments of our institution for routine head examinations. CT image quality for each scanner/acquisition protocol was assessed through noise level, noise power spectrum (NPS), contrast-to-noise ratio (CNR), modulation transfer function (MTF), low contrast detectability (LCD) and non-uniformity index analyses. Noise values ranged from 3.5 HU to 5.7 HU across scanners/acquisition protocols. NPS curves differed in terms of peak position (range: 0.21–0.30 mm(-1)). A substantial variation of CNR values with scanner/acquisition protocol was observed for different contrast inserts. The coefficient of variation (standard deviation divided by mean value) of CNR values across scanners/acquisition protocols was 18.3%, 31.4%, 34.2%, 30.4% and 30% for teflon, delrin, LDPE, polystyrene and acrylic insert, respectively. An appreciable difference in MTF curves across scanners/acquisition protocols was revealed, with a coefficient of variation of f(50%)/f(10%) of MTF curves across scanners/acquisition protocols of 10.1%/7.4%. A relevant difference in LCD performance of different scanners/acquisition protocols was found. The range of contrast threshold for a typical object size of 3 mm was 3.7–5.8 HU. Moreover, appreciable differences in terms of NUI values (range: 4.1%-8.3%) were found. The analysis of several quality indices showed a non-negligible variability in head CT imaging capabilities across different scanners/acquisition protocols. This highlights the importance of a physical in-depth characterization of image quality for each CT scanner as clinically used, in order to optimize CT imaging procedures.
format Online
Article
Text
id pubmed-7808662
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-78086622021-02-02 A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study Barca, Patrizio Paolicchi, Fabio Aringhieri, Giacomo Palmas, Federica Marfisi, Daniela Fantacci, Maria Evelina Caramella, Davide Giannelli, Marco PLoS One Research Article Nowadays, given the technological advance in CT imaging and increasing heterogeneity in characteristics of CT scanners, a number of CT scanners with different manufacturers/technologies are often installed in a hospital centre and used by various departments. In this phantom study, a comprehensive assessment of image quality of 5 scanners (from 3 manufacturers and with different models) for head CT imaging, as clinically used at a single hospital centre, was hence carried out. Helical and/or sequential acquisitions of the Catphan-504 phantom were performed, using the scanning protocols (CTDI(vol) range: 54.7–57.5 mGy) employed by the staff of various Radiology/Neuroradiology departments of our institution for routine head examinations. CT image quality for each scanner/acquisition protocol was assessed through noise level, noise power spectrum (NPS), contrast-to-noise ratio (CNR), modulation transfer function (MTF), low contrast detectability (LCD) and non-uniformity index analyses. Noise values ranged from 3.5 HU to 5.7 HU across scanners/acquisition protocols. NPS curves differed in terms of peak position (range: 0.21–0.30 mm(-1)). A substantial variation of CNR values with scanner/acquisition protocol was observed for different contrast inserts. The coefficient of variation (standard deviation divided by mean value) of CNR values across scanners/acquisition protocols was 18.3%, 31.4%, 34.2%, 30.4% and 30% for teflon, delrin, LDPE, polystyrene and acrylic insert, respectively. An appreciable difference in MTF curves across scanners/acquisition protocols was revealed, with a coefficient of variation of f(50%)/f(10%) of MTF curves across scanners/acquisition protocols of 10.1%/7.4%. A relevant difference in LCD performance of different scanners/acquisition protocols was found. The range of contrast threshold for a typical object size of 3 mm was 3.7–5.8 HU. Moreover, appreciable differences in terms of NUI values (range: 4.1%-8.3%) were found. The analysis of several quality indices showed a non-negligible variability in head CT imaging capabilities across different scanners/acquisition protocols. This highlights the importance of a physical in-depth characterization of image quality for each CT scanner as clinically used, in order to optimize CT imaging procedures. Public Library of Science 2021-01-14 /pmc/articles/PMC7808662/ /pubmed/33444367 http://dx.doi.org/10.1371/journal.pone.0245374 Text en © 2021 Barca et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Barca, Patrizio
Paolicchi, Fabio
Aringhieri, Giacomo
Palmas, Federica
Marfisi, Daniela
Fantacci, Maria Evelina
Caramella, Davide
Giannelli, Marco
A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study
title A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study
title_full A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study
title_fullStr A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study
title_full_unstemmed A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study
title_short A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study
title_sort comprehensive assessment of physical image quality of five different scanners for head ct imaging as clinically used at a single hospital centre—a phantom study
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7808662/
https://www.ncbi.nlm.nih.gov/pubmed/33444367
http://dx.doi.org/10.1371/journal.pone.0245374
work_keys_str_mv AT barcapatrizio acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT paolicchifabio acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT aringhierigiacomo acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT palmasfederica acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT marfisidaniela acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT fantaccimariaevelina acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT caramelladavide acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT giannellimarco acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT barcapatrizio comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT paolicchifabio comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT aringhierigiacomo comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT palmasfederica comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT marfisidaniela comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT fantaccimariaevelina comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT caramelladavide comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy
AT giannellimarco comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy