A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study
Nowadays, given the technological advance in CT imaging and increasing heterogeneity in characteristics of CT scanners, a number of CT scanners with different manufacturers/technologies are often installed in a hospital centre and used by various departments. In this phantom study, a comprehensive a...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7808662/ https://www.ncbi.nlm.nih.gov/pubmed/33444367 http://dx.doi.org/10.1371/journal.pone.0245374 |
_version_ | 1783636948434288640 |
---|---|
author | Barca, Patrizio Paolicchi, Fabio Aringhieri, Giacomo Palmas, Federica Marfisi, Daniela Fantacci, Maria Evelina Caramella, Davide Giannelli, Marco |
author_facet | Barca, Patrizio Paolicchi, Fabio Aringhieri, Giacomo Palmas, Federica Marfisi, Daniela Fantacci, Maria Evelina Caramella, Davide Giannelli, Marco |
author_sort | Barca, Patrizio |
collection | PubMed |
description | Nowadays, given the technological advance in CT imaging and increasing heterogeneity in characteristics of CT scanners, a number of CT scanners with different manufacturers/technologies are often installed in a hospital centre and used by various departments. In this phantom study, a comprehensive assessment of image quality of 5 scanners (from 3 manufacturers and with different models) for head CT imaging, as clinically used at a single hospital centre, was hence carried out. Helical and/or sequential acquisitions of the Catphan-504 phantom were performed, using the scanning protocols (CTDI(vol) range: 54.7–57.5 mGy) employed by the staff of various Radiology/Neuroradiology departments of our institution for routine head examinations. CT image quality for each scanner/acquisition protocol was assessed through noise level, noise power spectrum (NPS), contrast-to-noise ratio (CNR), modulation transfer function (MTF), low contrast detectability (LCD) and non-uniformity index analyses. Noise values ranged from 3.5 HU to 5.7 HU across scanners/acquisition protocols. NPS curves differed in terms of peak position (range: 0.21–0.30 mm(-1)). A substantial variation of CNR values with scanner/acquisition protocol was observed for different contrast inserts. The coefficient of variation (standard deviation divided by mean value) of CNR values across scanners/acquisition protocols was 18.3%, 31.4%, 34.2%, 30.4% and 30% for teflon, delrin, LDPE, polystyrene and acrylic insert, respectively. An appreciable difference in MTF curves across scanners/acquisition protocols was revealed, with a coefficient of variation of f(50%)/f(10%) of MTF curves across scanners/acquisition protocols of 10.1%/7.4%. A relevant difference in LCD performance of different scanners/acquisition protocols was found. The range of contrast threshold for a typical object size of 3 mm was 3.7–5.8 HU. Moreover, appreciable differences in terms of NUI values (range: 4.1%-8.3%) were found. The analysis of several quality indices showed a non-negligible variability in head CT imaging capabilities across different scanners/acquisition protocols. This highlights the importance of a physical in-depth characterization of image quality for each CT scanner as clinically used, in order to optimize CT imaging procedures. |
format | Online Article Text |
id | pubmed-7808662 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-78086622021-02-02 A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study Barca, Patrizio Paolicchi, Fabio Aringhieri, Giacomo Palmas, Federica Marfisi, Daniela Fantacci, Maria Evelina Caramella, Davide Giannelli, Marco PLoS One Research Article Nowadays, given the technological advance in CT imaging and increasing heterogeneity in characteristics of CT scanners, a number of CT scanners with different manufacturers/technologies are often installed in a hospital centre and used by various departments. In this phantom study, a comprehensive assessment of image quality of 5 scanners (from 3 manufacturers and with different models) for head CT imaging, as clinically used at a single hospital centre, was hence carried out. Helical and/or sequential acquisitions of the Catphan-504 phantom were performed, using the scanning protocols (CTDI(vol) range: 54.7–57.5 mGy) employed by the staff of various Radiology/Neuroradiology departments of our institution for routine head examinations. CT image quality for each scanner/acquisition protocol was assessed through noise level, noise power spectrum (NPS), contrast-to-noise ratio (CNR), modulation transfer function (MTF), low contrast detectability (LCD) and non-uniformity index analyses. Noise values ranged from 3.5 HU to 5.7 HU across scanners/acquisition protocols. NPS curves differed in terms of peak position (range: 0.21–0.30 mm(-1)). A substantial variation of CNR values with scanner/acquisition protocol was observed for different contrast inserts. The coefficient of variation (standard deviation divided by mean value) of CNR values across scanners/acquisition protocols was 18.3%, 31.4%, 34.2%, 30.4% and 30% for teflon, delrin, LDPE, polystyrene and acrylic insert, respectively. An appreciable difference in MTF curves across scanners/acquisition protocols was revealed, with a coefficient of variation of f(50%)/f(10%) of MTF curves across scanners/acquisition protocols of 10.1%/7.4%. A relevant difference in LCD performance of different scanners/acquisition protocols was found. The range of contrast threshold for a typical object size of 3 mm was 3.7–5.8 HU. Moreover, appreciable differences in terms of NUI values (range: 4.1%-8.3%) were found. The analysis of several quality indices showed a non-negligible variability in head CT imaging capabilities across different scanners/acquisition protocols. This highlights the importance of a physical in-depth characterization of image quality for each CT scanner as clinically used, in order to optimize CT imaging procedures. Public Library of Science 2021-01-14 /pmc/articles/PMC7808662/ /pubmed/33444367 http://dx.doi.org/10.1371/journal.pone.0245374 Text en © 2021 Barca et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Barca, Patrizio Paolicchi, Fabio Aringhieri, Giacomo Palmas, Federica Marfisi, Daniela Fantacci, Maria Evelina Caramella, Davide Giannelli, Marco A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study |
title | A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study |
title_full | A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study |
title_fullStr | A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study |
title_full_unstemmed | A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study |
title_short | A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre—A phantom study |
title_sort | comprehensive assessment of physical image quality of five different scanners for head ct imaging as clinically used at a single hospital centre—a phantom study |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7808662/ https://www.ncbi.nlm.nih.gov/pubmed/33444367 http://dx.doi.org/10.1371/journal.pone.0245374 |
work_keys_str_mv | AT barcapatrizio acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT paolicchifabio acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT aringhierigiacomo acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT palmasfederica acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT marfisidaniela acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT fantaccimariaevelina acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT caramelladavide acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT giannellimarco acomprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT barcapatrizio comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT paolicchifabio comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT aringhierigiacomo comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT palmasfederica comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT marfisidaniela comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT fantaccimariaevelina comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT caramelladavide comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy AT giannellimarco comprehensiveassessmentofphysicalimagequalityoffivedifferentscannersforheadctimagingasclinicallyusedatasinglehospitalcentreaphantomstudy |